Ronald R Hugate1, Ian D Dickey, Qingshan Chen, Christina M Wood, Franklin H Sim, Michael G Rock.
Abstract
Secure fixation of acetabular components in total hip arthroplasty can be challenging. The purpose of this study was to perform biomechanical analysis of cup fixation strength using fixed-angle vs standard screw fixation. Multihole, porous-backed acetabular prostheses were implanted in both acetabuli of 8 cadaveric pelves using standard press-fit techniques. Fixed-angle screws were used on the left side, and standard cancellous screws were used in the right. The use of fixed-angle screws enhanced acetabular fixation substantially under subfailure cyclic loading conditions and load-to-failure. The triradiate screw configuration increases the bending moment required to fail the specimens as well. Fixed-angle screws may be useful for achieving rigid fixation of acetabular prostheses in challenging clinical scenarios.Entities:
Mesh:
Year: 2008 PMID: 18639433 DOI: 10.1016/j.arth.2008.04.019