Transcutaneous carbon dioxide monitoring in spontaneously breathing, nonintubated patients in the early postoperative period.

G Fanelli1, M Baciarello, G Squicciarini, G Malagutti, M Zasa, A Casati.   

Abstract

BACKGROUND: The authors investigated the accuracy of transcutaneous capnometry (TcPCO(2)) in estimating arterial blood carbon dioxide partial pressure (PaCO(2)) during spontaneous breathing in patients admitted to our surgical intensive care unit (ICU).
METHODS: Serial TcPCO(2) and PaCO(2) measurements were taken in stable patients undergoing postoperative monitoring after major abdominal, vascular, or thoracic surgery. Patients were enrolled 12 hours after extubation. Exclusion criteria were pulmonary dysfunction, hemodynamic instability, or anemia. Linear regression, mixed models, and Bland-Altman analyses were used to compare accuracy and correlation between the two variables. Data are presented as means (95% confidence intervals).
RESULTS: PaCO(2) values ranged between 26 mmHg and 52 mmHg. Mean values for TcPCO(2) and PaCO(2) were 35.3 (33.8-36.8) mmHg and 39.2 (37.6-40.7) (P<0.001). Bland-Altman analysis showed a bias of -3.8 (-5.1-2.5) mmHg with upper and lower limits of agreement of 5.6 (3.4-7.9) mmHg and -13.3 (-15.6 -11.1) mmHg, respectively. Thirty-one (55.4%) measurements disagreed by 3 mmHg or more. Linear mixed model analysis with adjustment for repeated measurements showed low correlation (r=0.63; P<0.0005). There were small but significant differences between patients in the correlation coefficient (P=0.04).
CONCLUSION: TcPCO(2) showed moderate bias with wide limits of agreement when compared to PaCO(2) in our patients. The regression model showed low correlation of the measurements when adjusting for repeated measurements and between-subject variances. TcPCO(2) may not provide an accurate quantitative estimation of PaCO(2) in stable, spontaneously breathing, nonintubated patients in the early postoperative period.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18612267

Source DB:  PubMed          Journal:  Minerva Anestesiol        ISSN: 0375-9393            Impact factor:   3.051


  4 in total

1.  Comparison of ear and chest probes in transcutaneous carbon dioxide pressure measurements during general anesthesia in adults.

Authors:  Tomoki Nishiyama; Yumiko Kohno; Keiko Koishi
Journal:  J Clin Monit Comput       Date:  2011-10-19       Impact factor: 2.502

2.  Transcutaneous continuous carbon dioxide tension monitoring reduced incidence, degree and duration of hypercapnia during combined regional anaesthesia and monitored anaesthesia care in shoulder surgery patients.

Authors:  Werner Baulig; Marija Keselj; Barbara Baulig; Sandra Guzzella; Alain Borgeat; José Aguirre
Journal:  J Clin Monit Comput       Date:  2014-10-14       Impact factor: 2.502

3.  Short term general anesthesia for retro-bulbar block in ophthalmic surgery generates no significant hypercapnia.

Authors:  Werner Baulig; Monica Weber; Beatrice Beck-Schimmer; Oliver M Theusinger; Peter Biro
Journal:  J Clin Monit Comput       Date:  2017-03-11       Impact factor: 2.502

4.  Cutaneous vesicles caused by transcutaneous gas-monitoring sensor.

Authors:  Murali Chakravarthy; Sandeep Narayan; Raghav Govindarajan; Subramnyam Rajeev; Vivek Jawali
Journal:  J Cutan Aesthet Surg       Date:  2009-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.