Máximo Rivero-Ayerza1, Emil Jessurun, Steve Ramcharitar, Yves van Belle, Patrick W Serruys, Luc Jordaens.
Abstract
AIMS: Left ventricular (LV) lead implantation is feasible using remote magnetic navigation of a guidewire (Stereotaxis, St Louis, MO, USA). A novel software that performs a three-dimensional (3D) reconstruction of vessels based on two or more angiographic views has been developed recently (CardiOp-B system, Paeion Inc., Haifa, Israel). The objective of this paper is to evaluate: (i) the performance of the 3D reconstruction software which reproduce the anatomy of the coronary sinus (CS) and (ii) the efficacy of remotely navigating a magnetic guidewire within the CS based on this reconstruction. METHODS ANDEntities:
Mesh:
Year: 2008 PMID: 18587136 PMCID: PMC2517161 DOI: 10.1093/europace/eun173
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Procedure characteristics and left ventricular lead position
| Patient number | Procedure time (min) | Fluoroscopy time (min) | LVL fluoroscopy (min) | Target SB accessed w/guidewire | Final LVL location | Pacing threshold (mV) |
|---|---|---|---|---|---|---|
| 1 | 150 | 18.4 | 0.2 | Yes | PL | 1 |
| 2 | 85 | 36.7 | 3 | Yes | AL | 0.4 |
| 3 | 75 | 23.3 | 2.36 | Yes | PL | 0.5 |
| 4 | – | – | – | Yes | PL | 1.1 |
| 5 | 95 | 23.2 | 1.64 | Yes | PL | 0.7 |
| 6 | 140 | 46 | 0.13 | Yes | Ant | 1.6 |
| 7 | 96 | 14.48 | 4.5 | Yes | L | 1 |
| 8 | 101 | 19.2 | 0.33 | Yes | L | 1.2 |
| 9 | 75 | 18.4 | 0.26 | Yes | PL | 0.7 |
| 10 | 70 | 7.5 | 2.26 | Yes | L | 0.8 |
| 11 | 85 | 27.35 | 1.05 | Yes | L | 0.9 |
| 12 | 110 | 9.41 | 0.18 | Yes | Ant | 4.8 |
| 13 | 98 | 16.21 | 2 | Yes | PL | 2 |
| 14 | 60 | 8.7 | 3.05 | Yes | L | 1.5 |
| 15 | 110 | 21.42 | 2.4 | Yes | PL | 1.1 |
| 16 | 130 | 58 | 2.2 | Yes | L | 0.4 |
| Mean (SD) | 98.6 (25) | 23.2 (14) | 1.7 (1.3) | 1.2 (1) |
AL, antero-lateral; Ant, anterior; LVL, left ventricular lead; L, lateral; PL, postero-lateral side; SB, coronary sinus side branch; SD, standard deviation.
Quality of reconstruction and navigation
| Patient number | Quality of 3D reconstruction | Edge adjustment needed | Type of navigation |
|---|---|---|---|
| 1 | Good | Yes | Automatic |
| 2 | Good | Yes | Automatic |
| 3 | Good | Yes | Automatic |
| 4 | Not obtained | – | – |
| 5 | Good | Yes | Automatic |
| 6 | Good | Yes | Automatic |
| 7 | Good | Yes | Automatic |
| 8 | Good | No | Automatic |
| 9 | Good | Yes | Automatic + manual |
| 10 | Good | No | Automatic |
| 11 | Poor | Yes | Automatic |
| 12 | Good | No | Automatic |
| 13 | Poor | Yes | Automatic + manual |
| 14 | Good | No | Automatic |
| 15 | Good | Yes | Automatic |
| 16 | Good | No | Automatic |