Literature DB >> 18581341

MRA of intracranial aneurysms embolized with platinum coils: a vascular phantom study at 1.5T and 3T.

Shingo Kakeda1, Yukunori Korogi, Yasuhiro Hiai, Toru Sato, Norihiro Ohnari, Junji Moriya, Koji Kamada.   

Abstract

PURPOSE: To analyze the influence of matrix and echo time (TE) of three-dimensional time-of-flight (3D TOF) magnetic resonance angiography (MRA) on the depiction of residual flow in aneurysms embolized with platinum coils at 1.5T and 3T.
MATERIALS AND METHODS: A simulated intracranial aneurysm of the vascular phantom was loosely packed to maintain the patency of some residual aneurysmal lumen with platinum coils and connected to an electromagnetic flow pump with pulsatile flow. MRAs were obtained altering the matrix and TE of 3D TOF sequences at 1.5T and 3T.
RESULTS: The increased spatial resolution and the shorter TE offered better image quality at 3T. For the depiction of an aneurysm remnant, the high-spatial-resolution 3T MRA (matrix size of 384 x 224 and 512 x 256) with a short TE of < or =3.3 msec were superior to the 1.5T MRA obtained with any sequences.
CONCLUSION: 3T MRA is superior to 1.5T MRA for the assessment of aneurysms embolized with platinum coils; the combination of the 512 x 256 matrix and short TE (3.3 msec or less) seems feasible at 3T. (c) 2008 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18581341     DOI: 10.1002/jmri.21421

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  Design and construction of a multipath vessel phantom for interventional training.

Authors:  Y Thakur; H N Nikolov; I B Gulka; D W Holdsworth; M Drangova
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

2.  Follow-up assessment of coiled intracranial aneurysms using zTE MRA as compared with TOF MRA: a preliminary image quality study.

Authors:  Song'an Shang; Jing Ye; Xianfu Luo; Jianxun Qu; Yong Zhen; Jingtao Wu
Journal:  Eur Radiol       Date:  2017-04-05       Impact factor: 5.315

3.  Normal flow signal of the pterygoid plexus on 3T MRA in patients without DAVF of the cavernous sinus.

Authors:  K Watanabe; S Kakeda; R Watanabe; N Ohnari; Y Korogi
Journal:  AJNR Am J Neuroradiol       Date:  2012-12-28       Impact factor: 3.825

4.  Temporal evolution of susceptibility artifacts from coiled aneurysms on MR angiography: an in vivo canine study.

Authors:  G Spilberg; S L Carniato; R M King; I M J van der Bom; M Mehra; R P Walvick; A K Wakhloo; M J Gounis
Journal:  AJNR Am J Neuroradiol       Date:  2011-12-22       Impact factor: 3.825

  4 in total