Literature DB >> 18567669

Minimally invasive off-pump valve-in-a-valve implantation: the atrial transcatheter approach for re-operative mitral valve replacement.

Jörg Kempfert1, Johannes M Blumenstein, Michael A Borger, Axel Linke, Sven Lehmann, Patrick Pritzwald-Stegmann, Michael W A Chu, Gerhard Schuler, Volkmar Falk, Friedrich Wilhelm Mohr, Thomas Walther.   

Abstract

AIMS: This study was designed to evaluate the feasibility and haemodynamic performance of transcatheter valve-in-a-valve (VinV) implantation for failed mitral xenografts using a minimally invasive, transatrial, off-pump approach. METHODS AND
RESULTS: Nine adult sheep (71.1 +/- 3.4 kg) underwent mitral valve replacement with a 25 mm bioprosthesis using standard conventional techniques. After weaning from cardiopulmonary bypass, a transcatheter 23 mm pericardial prosthesis mounted on a balloon-inflatable steel stent was deployed within the conventional xenograft. The VinV implantation (10.1 +/- 0.9 min) was performed off-pump using an antegrade transatrial approach under fluoroscopic guidance successfully in all nine sheep. Mean transvalvular gradient was 4.6 +/- 1.0 mmHg, and five of nine sheep had no paravalvular leak with two sheep each having mild and moderate one. All transcatheter prostheses were confirmed in good position on post-mortem analysis. In an in vitro model, the minimum force required to dislodge the valve was 22.5 +/- 3.0 N, which was well above the normal estimated forces generated by the left ventricle (LV). When this model was attached to a LV assist device, the VinV withstood pressure loops of 300 mmHg, without dislocation or embolization.
CONCLUSION: Transatrial, transcatheter mitral VinV implantation is feasible off pump. This is a truly minimally invasive concept to treat patients with failed xenografts using a right lateral minithoracotomy.

Entities:  

Mesh:

Year:  2008        PMID: 18567669     DOI: 10.1093/eurheartj/ehn285

Source DB:  PubMed          Journal:  Eur Heart J        ISSN: 0195-668X            Impact factor:   29.983


  3 in total

1.  Transcatheter valve used in a bailout technique during complicated open mitral valve surgery.

Authors:  Parla Astarci; David Glineur; Laurent De Kerchove; Gebrine El Khoury
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-06-28

2.  Use of a melody pulmonary valve in transcatheter valve-in-valve replacement for tricuspid valve bioprosthesis degeneration.

Authors:  David M Filsoof; David F Snipelisky; Brian P Shapiro
Journal:  Tex Heart Inst J       Date:  2014-10-01

Review 3.  Transcatheter valve-in-valve therapies: patient selection, prosthesis assessment and selection, results, and future directions.

Authors:  Manuel Wilbring; Konstantin Alexiou; Sems-Malte Tugtekin; Bjoern Sill; Gregor Simonis; Klaus Matschke; Utz Kappert
Journal:  Curr Cardiol Rep       Date:  2013-03       Impact factor: 2.931

  3 in total