Literature DB >> 17339542

Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation.

Johannes Nordmeyer1, Sachin Khambadkone, Louise Coats, Silvia Schievano, Philipp Lurz, Giovanni Parenzan, Andrew M Taylor, James E Lock, Philipp Bonhoeffer.   

Abstract

BACKGROUND: We analyzed the incidence, risk factors and treatment options for stent fracture after percutaneous pulmonary valve (PPV) implantation (PPVI). METHODS AND
RESULTS: After PPVI, 123 patients had chest x-ray in anteroposterior and lateral projection, echocardiography, and clinical evaluation during structured follow-up. Of these 123 patients, 26 (21.1%) developed stent fracture 0 to 843 days after PPVI (stent fracture-free survival at 1 year, 85.1%; at 2 years, 74.5%; and at 3 years, 69.2%). Stent fracture was classified as type I: no loss of stent integrity (n=17); type II: loss of integrity with restenosis on echocardiography (n=8); and type III: separation of fragments or embolization (n=1). In a multivariate Cox regression, we analyzed various factors, of which 3 were associated with a higher risk of stent fracture: implantation into "native" right ventricular outflow tract (P=0.04), no calcification along the right ventricular outflow tract (judged with fluoroscopy, P=0.02), recoil of PPV (qualitatively, PPV diameter in frontal or lateral plane with fully inflated balloon > diameter after balloon deflation, P=0.03). Substernal PPV location, high-pressure post-PPVI dilatation of PPV, pre-PPVI right ventricular outflow tract gradients, and other indicators of PPV compression or asymmetry did not pose increased risk. Patients with type I fracture remain under follow-up. Patients with type II fracture had 2nd PPVI or are awaiting such procedure, and 1 patient with type III fracture required surgical explantation.
CONCLUSIONS: Stent fracture after PPVI can be managed effectively by risk stratification, systematic classification, and anticipatory management strategies. Serial x-ray and echocardiography are recommended for surveillance.

Entities:  

Mesh:

Year:  2007        PMID: 17339542     DOI: 10.1161/CIRCULATIONAHA.106.674259

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  33 in total

Review 1.  Computer modeling to tailor therapy for congenital heart disease.

Authors:  Michael A Quail; Andrew M Taylor
Journal:  Curr Cardiol Rep       Date:  2013-09       Impact factor: 2.931

2.  Short- and medium-term outcomes after transcatheter pulmonary valve placement in the expanded multicenter US melody valve trial.

Authors:  Doff B McElhinney; William E Hellenbrand; Evan M Zahn; Thomas K Jones; John P Cheatham; James E Lock; Julie A Vincent
Journal:  Circulation       Date:  2010-07-19       Impact factor: 29.690

3.  Clinical issues and outcomes in adults following repair of tetralogy of fallot.

Authors:  Bejal Pandya; Michael A Quail; Seamus Cullen
Journal:  Curr Treat Options Cardiovasc Med       Date:  2013-10

4.  Percutaneous pulmonic valve implantation.

Authors:  Arun Kumar; Clifford Kavinsky; Zahid Amin; Ziyad M Hijazi
Journal:  Curr Treat Options Cardiovasc Med       Date:  2009-12

5.  Prolonged ischaemia impairs muscle blood flow and oxygen uptake dynamics during subsequent heavy exercise.

Authors:  Azmy Faisal; Kenneth S Dyson; Richard L Hughson
Journal:  J Physiol       Date:  2010-10-01       Impact factor: 5.182

Review 6.  Compression, distortion and dislodgement of large caliber stents in congenital heart defects caused by cardiopulmonary resuscitation: a case series and review of the literature.

Authors:  Nikolaus A Haas; Christoph M Happel; Smita Jategaonkar; Axel Moysich; Andreas Hanslik; Deniz Kececioglu; Eugen Sandica; Kai Thorsten Laser
Journal:  Clin Res Cardiol       Date:  2014-04-04       Impact factor: 5.460

7.  Four-dimensional computed tomography: a method of assessing right ventricular outflow tract and pulmonary artery deformations throughout the cardiac cycle.

Authors:  Silvia Schievano; Claudio Capelli; Carol Young; Philipp Lurz; Johannes Nordmeyer; Catherine Owens; Philipp Bonhoeffer; Andrew M Taylor
Journal:  Eur Radiol       Date:  2010-08-01       Impact factor: 5.315

Review 8.  Transcatheter Pulmonary Valve Replacement: Current State of Art.

Authors:  Wail Alkashkari; Amani Alsubei; Ziyad M Hijazi
Journal:  Curr Cardiol Rep       Date:  2018-03-15       Impact factor: 2.931

9.  Percutaneous transjugular implantation of MELODY® valve into tricuspid bioprosthesis.

Authors:  Castigliano M Bhamidipati; D Scott Lim; Michael Ragosta; Gorav Ailawadi
Journal:  J Card Surg       Date:  2013-06-17       Impact factor: 1.620

10.  Effective transcatheter valve implantation after pulmonary homograft failure: a new perspective on the Ross operation.

Authors:  Johannes Nordmeyer; Philipp Lurz; Victor T Tsang; Louise Coats; Fiona Walker; Andrew M Taylor; Sachin Khambadkone; Marc R de Leval; Philipp Bonhoeffer
Journal:  J Thorac Cardiovasc Surg       Date:  2009-04-25       Impact factor: 5.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.