The sound intensity and characteristics of variable-pitch pulse oximeters.

Hiroo Yamanaka1, Junichi Haruna, Takashi Mashimo, Takeshi Akita, Keiko Kinouchi.   

Abstract

OBJECTIVE: Various studies worldwide have found that sound levels in hospitals significantly exceed the World Health Organization (WHO) guidelines, and that this noise is associated with audible signals from various medical devices. The pulse oximeter is now widely used in health care; however the health effects associated with the noise from this equipment remain largely unclarified. Here, we analyzed the sounds of variable-pitch pulse oximeters, and discussed the possible associated risk of sleep disturbance, annoyance, and hearing loss.
METHODS: The Nellcor N 595 and Masimo SET Radical pulse oximeters were measured for equivalent continuous A-weighted sound pressure levels (L(Aeq)), loudness levels, and loudness. Pulse beep pitches were also identified using Fast Fourier Transform (FFT) analysis and compared with musical pitches as controls.
RESULTS: Almost all alarm sounds and pulse beeps from the instruments tested exceeded 30 dBA, a level that may induce sleep disturbance and annoyance. Several alarm sounds emitted by the pulse oximeters exceeded 70 dBA, which is known to induce hearing loss. The loudness of the alarm sound of each pulse oximeter did not change in proportion to the sound volume level. The pitch of each pulse beep did not correspond to musical pitch levels.
CONCLUSIONS: The results indicate that sounds from pulse oximeters pose a potential risk of not only sleep disturbance and annoyance but also hearing loss, and that these sounds are unnatural for human auditory perception.

Entities:  

Mesh:

Year:  2008        PMID: 18483870     DOI: 10.1007/s10877-008-9125-0

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  19 in total

1.  Excessive noise levels in the neonatal ICU: potential effects on auditory system development.

Authors:  William D T Kent; Andre K W Tan; Maxine C Clarke; Trevor Bardell
Journal:  J Otolaryngol       Date:  2002-12

2.  On the psychophysical law.

Authors:  S S STEVENS
Journal:  Psychol Rev       Date:  1957-05       Impact factor: 8.934

3.  The sounds of saturation.

Authors:  David C Santamore; Thomas G Cleaver
Journal:  J Clin Monit Comput       Date:  2004-04       Impact factor: 2.502

4.  Noise-induced stress as a predictor of burnout in critical care nurses.

Authors:  M Topf; E Dillon
Journal:  Heart Lung       Date:  1988-09       Impact factor: 2.210

5.  Noise in the operating room.

Authors:  R A Shapiro; T Berland
Journal:  N Engl J Med       Date:  1972-12-14       Impact factor: 91.245

6.  A study of noise and its relationship to patient discomfort in the recovery room.

Authors:  B B Minckley
Journal:  Nurs Res       Date:  1968 May-Jun       Impact factor: 2.381

Review 7.  Environmental impact of the NICU on developmental outcomes.

Authors:  S Blackburn
Journal:  J Pediatr Nurs       Date:  1998-10       Impact factor: 2.145

8.  Crying wolf: false alarms in a pediatric intensive care unit.

Authors:  S T Lawless
Journal:  Crit Care Med       Date:  1994-06       Impact factor: 7.598

9.  Hearing acuity of anesthesiologists and alarm detection.

Authors:  M S Wallace; M N Ashman; M J Matjasko
Journal:  Anesthesiology       Date:  1994-07       Impact factor: 7.892

10.  Assessment of patients' experience of discomforts during respirator therapy.

Authors:  I Bergbom-Engberg; H Haljamäe
Journal:  Crit Care Med       Date:  1989-10       Impact factor: 7.598

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.