Literature DB >> 1844129

Release and excretion of metal in patients who have a total hip-replacement component made of titanium-base alloy.

J J Jacobs1, A K Skipor, J Black, R m Urban, J O Galante.   

Abstract

Serum concentration and urinary excretion of titanium, aluminum, and vanadium were measured for patients who had a well functioning cementless primary total hip replacement of one of two different designs, for patients who had a loose total hip replacement that was to be revised, and for control subjects who had no implant. Serum concentrations of titanium were elevated approximately twofold in the patients who had a loose implant, compared with the values for the control subjects. No major differences in terms of urine concentration of titanium, serum concentration of aluminum, or urine concentration of aluminum were observed among any of the groups that were studied. Concentrations of vanadium were uniformly low in all groups.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1844129

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  31 in total

1.  Modulation of the phenotypic and functional properties of phagocytic macrophages by wear particles from orthopaedic implants.

Authors:  N Al-Saffar; P A Revell; A Kobayashi
Journal:  J Mater Sci Mater Med       Date:  1997-11       Impact factor: 3.896

Review 2.  Surface treatments and roughness properties of Ti-based biomaterials.

Authors:  Andrea Bagno; Carlo Di Bello
Journal:  J Mater Sci Mater Med       Date:  2004-09       Impact factor: 3.896

Review 3.  Titanium dioxide nanoparticles: a review of current toxicological data.

Authors:  Hongbo Shi; Ruth Magaye; Vincent Castranova; Jinshun Zhao
Journal:  Part Fibre Toxicol       Date:  2013-04-15       Impact factor: 9.400

4.  Aluminum levels and stores in patients with total hip endoprostheses from TiAIV or TiAINb alloys.

Authors:  D D Dittert; G Warnecke; H G Willert
Journal:  Arch Orthop Trauma Surg       Date:  1995       Impact factor: 3.067

5.  Biodistribution of titanium dioxide from biologic compartments.

Authors:  Daniel G Olmedo; Deborah R Tasat; María Beatriz Guglielmotti; Rómulo Luis Cabrini
Journal:  J Mater Sci Mater Med       Date:  2008-04-04       Impact factor: 3.896

6.  Serum and urine metal levels in patients with metal-on-metal surface arthroplasty.

Authors:  Anastasia K Skipor; Patricia A Campbell; Leslie M Patterson; Harlan C Anstutz; Thomas P Schmalzried; Joshua J Jacobs
Journal:  J Mater Sci Mater Med       Date:  2002-12       Impact factor: 3.896

7.  The study of metal ion release and cytotoxicity in Co-Cr-Mo and Ti-Al-V alloy in total knee prosthesis - scanning electron microscopic observation.

Authors:  Shizuko Ichinose; Takeshi Muneta; Ichiro Sekiya; Soichiro Itoh; Hideki Aoki; Motoki Tagami
Journal:  J Mater Sci Mater Med       Date:  2003-01       Impact factor: 3.896

8.  Titanium transport through the blood stream. An experimental study on rats.

Authors:  Daniel G Olmedo; Débora Tasat; María B Guglielmotti; Rómulo L Cabrini
Journal:  J Mater Sci Mater Med       Date:  2003-12       Impact factor: 3.896

9.  PIXE micro-beam mapping of metals in human peri-implant tissues.

Authors:  P Passi; A Zadro; S Galassini; P Rossi; G Moschini
Journal:  J Mater Sci Mater Med       Date:  2002-11       Impact factor: 3.896

10.  An experimental study of the dissemination of Titanium and Zirconium in the body.

Authors:  Daniel Olmedo; María Beatriz Guglielmotti; Rómulo Luis Cabrini
Journal:  J Mater Sci Mater Med       Date:  2002-08       Impact factor: 3.896