| Literature DB >> 18437199 |
Daisuke Takahashi1, Yang Xiao, Fei Hu, Michael Lewis.
Abstract
This paper surveys diabetes therapies from telemedicine viewpoint. In type 1 diabetes therapies, the exogenous insulin replacement is generally considered as a primary treatment. However, the complete replacement of exogenous insulin is still a challenging issue because of its complexity of modeling the dynamics, which is typically modeled nonlinearly. On the other hand, thanks to the progress of medical devices, currently the diabetes therapies are being automated. These medical devices include automated insulin pumps and blood glucose sensors. Insulin pumps are designed to create artificial insulin perfusion while they largely rely on the blood glucose profile measurements and these measurements are achieved by one or more blood glucose sensors. The blood glucose measurements are also important for the insulin-dependent diabetes therapies. An insulin pump along with sensors establishes a good feedback system providing the appropriate amount of the exogenous insulin on demand. Controlling the amount of exogenous insulin to suppress the blood glucose levels requires complicated computations. This paper mostly explains both type 1 and 2 diabetes and their mechanisms accompanied by descriptions of diabetes therapy and medical devices currently utilized in the therapy.Entities:
Year: 2008 PMID: 18437199 PMCID: PMC2271045 DOI: 10.1155/2008/639019
Source DB: PubMed Journal: Int J Telemed Appl ISSN: 1687-6415
Figure 1An example of insulin pump that integrates a blood glucose sensor [21]. Patient’s blood is sampled from the tip of his/her finger, and from the direct measurements of a blood glucose sensor, the amount of the short-acting insulin is adjusted and continuously delivered into the human body.
Figure 2An image of an insulin pump. This image also includes a blood glucose sensor. In the general use, an insulin pump continuously infuses a little amount of the short-acting exogenous insulin via an attached needle as a basal dosage. It is also able to add extra exogenous insulin for intensive care, for example, meal time. Insulin pumps have a lot of advantages compared with the insulin shots.
Figure 3Another example of insulin pumps of Medtronic. A processing module has a small display that shows a current rate of an exogenous insulin infusion. It also has several control buttons just below the display. Two of these buttons are used to manage a current insulin infusion rate.
Figure 4Piezoelectric pump and micro-needle device.
Specification comparison of commercial insulin pumps.
| Company | Animas | Deltec | Disetronic | MiniMed | Insulet |
|---|---|---|---|---|---|
| Model | IR-1250 [ | Cozmo [ | Spirit [ | Paradigm | OmniPod [ |
|
| |||||
| Dimensions | 79 × 51 × 19 | 80 × 47 × 24 | 80 × 56 × 20 | 522: 51 × 79 × 20 | Pod: 41 × 61 × 18 |
| 722: 51 × 79 × 20 | Pda: 66 × 110 × 26 | ||||
|
| |||||
| Screen size | 992 sq mm | 870 sq mm | Unavailable | 774 sq mm | 1,848 sq mm on PDA controller |
|
| |||||
| Basal delivery | Every 3 minutes | Every 3 minutes | Every 3 minutes | Varies | Information unavailable |
|
| |||||
| Basal temperature | Initially −90% to +200%, varies for every half an hour | Varies from 0% to 200% with an increment of 5% for every half an hour | Varies from 0% to 200% with an increment of 10% for every half an hour | +/−0.1 increment as single basal rate for 0.5 to 24 hours | Information unavailable |
|
| |||||
| Carb and correction factors | Manual entry and assist from EZ manager | Manual carbohydrate, BG from attached CoZ monitor | Manual carbohydrate, BG from Accu-check BG monitor | Manual carbohydrate, BG from BD meter or manual entry | Information unavailable |
|
| |||||
| Battery | AA lithium × 1 | AAA × 1 | AA × 1 Alkaline or Rechargeable | AAA | AAA × 2 |
|
| |||||
| Motor | DC | DC | DC | DC | Stepper |
|
| |||||
| Memory | Nonvolatile: 600 bolus, 270 basal, 120 daily totals, 30 alarms, and 60 primes | Nonvolatile: 90 days of basals, carbohydrates, boluses, correction boluses, and alarms | Nonvolatile: 90 days history recall of last 30 boluses, alerts, daily insulin totals, and temporary basal rate increase | 4000 events volatile: 24 boluses, and 7 days totals | 90 days of data |
|
| |||||
| Extra features | Clip-on covers, personalized carbohydrate, and correction factors, tracks residual bolus insulin | Carbohydrate and correction factors, tracks residual bolus insulin, detailed records of pump, and daily bolus total correction | Availability of different types of user menus, icon and menu driven programming, backlight display, and reversible display screen | Extended bolus, auto off | Integrated free style meter, 1000 common foods in PDA |
Figure 5An image shows a miniature size of in vivo blood glucose sensor that is being developed [15]. It is currently proven by FDA that in vivo blood glucose sensors are only safe and reliable in a short period of time (e.g., three days). Thus after the period, they are required to be replaced by a new one regularly to keep their accuracy.
Figure 6An image shows a fingerstick meter reading. In the use of the sensor, first off, cut the tip of a finger or earlobe bleeding. Then put the blood on a blood glucose sensor, which is a protuberance at the bottom of the module. The result is shown on the display in the middle-top of the module.
Figure 7An image shows a list-watch type noninvasive blood glucose measurement [1]. The sensor can avoid damaging the tip of a finger or earlobe to sample blood in a noninvasive way. Therefore, it basically utilizes a quite low electric current to draw blood glucose penetrable through the skin. This sampled glucose is aggregated into a sensing part of the device integrating an electrode that is used for measuring blood glucose levels.
Figure 8A capillary and interstitial compartmental model [43, 44]. glucose uptake, and diffusions between capillary and interstitial tissues are represented as arrows.
Figure 9Microneedle honeycomb formation.
Figure 10A control flow of closed-loop control models [15]. Entire loop of the control is closed from outside by utilizing an insulin pump and in vivo blood glucose sensors. Both insulin injections and glucose measurements are carried out subcutaneously.
Figure 11A control flow of partially closed-loop insulin therapy models from [15]. In general, in this case, insulin dosages are supported by an expert system that generates an optimal insulin dosage from a comparison between a model case and a prediction of a future blood glucose transition.
Specification comparison of commercial blood glucose sensors.
| Features | Abbott freestyle navigator [ | MiniMed paradigm real-time system [ | MiniMed guardian real-time system [ | DexCom [ |
|---|---|---|---|---|
| Accuracy | Varies | Consensus error grid: 98.9% | Consensus error grid: 98.9% | Consensus error grid: 95.4% |
|
| ||||
| Startup initiation time | 10 hours | 2 hours | 2 hours | 2 hours |
|
| ||||
| Sensor life | 5-day wear indication | Above 72 hours | Above 72 hours | Above 72 hours |
|
| ||||
| Calibration method | Requires calibration at 10, 12, 24, and 72 hours after the insertion of the sensor | Alarms when calibration value is not entered on time. First and second calibration should be done for 2 and 6 hours after insertion | Alarms when calibration value is not entered on time. First and second calibration should be done for 2 and 6 hours after insertion | First calibration after 30 minutes and then for every 12 hours. Manual calibration is not possible |
|
| ||||
| Frequency of display | Every 1 minute | Every 5 minutes | Every 5 minutes | Every 5 minutes |
|
| ||||
| Transmitter memory | — | Yes, the transmitter stores missed data for up to 40 minutes | Yes, the transmitter stores missed data for up to 40 minutes | No, transmission lost is data lost |
|
| ||||
| Range of monitor to transmitter | 10 feet | 6 feet | 6 feet | 5 feet |
|
| ||||
| Monitor batteries | Uses 2 AAA batteries with replacement for every three months | No separate monitor required. Uses insulin pump | Uses 2 AAA Batteries. Indication is set for chance of battery | Uses rechargeable batteries |
|
| ||||
| Monitor size | 3“ × 2.5” | Separate monitor is not available. Uses insulin pump for display | 3“ × 2.7” | 3“ × 2.5” |
|
| ||||
| Alarms on user-set low and high thresholds | Applicable | Applicable | Applicable | Applicable |