Three-dimensional ultrasound-guided robotic needle placement: an experimental evaluation.

Emad M Boctor1, Michael A Choti, Everette C Burdette, Robert J Webster Iii.   

Abstract

BACKGROUND: Clinical use of image-guided needle placement robots has lagged behind laboratory-demonstrated robotic capability. Bridging this gap requires reliable and easy-to-use robotic systems.
METHODS: Our system for image-guided needle placement requires only simple, low-cost components and minimal, entirely off-line calibration. It rapidly aligns needles to planned entry paths using 3D ultrasound (US) reconstructed from freehand 2D scans. We compare system accuracy against clinical standard manual needle placement.
RESULTS: The US-guided robotic system is significantly more accurate than single manual insertions. When several manual withdrawals and reinsertions are allowed, accuracy becomes equivalent. In ex vivo experiments, robotic repeatability was 1.56 mm, compared to 3.19 and 4.63 mm for two sets of manual insertions. In an in vivo experiment with heartbeat and respiratory effects, robotic system accuracy was 5.5 mm.
CONCLUSIONS: A 3D US-guided robot can eliminate error bias and reduce invasiveness (the number of insertions required) compared to manual needle insertion. Remaining future challenges include target motion compensation. (c) 2008 John Wiley & Sons, Ltd.

Entities:  

Mesh:

Year:  2008        PMID: 18433079      PMCID: PMC4076171          DOI: 10.1002/rcs.184

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  18 in total

Review 1.  Stradx: real-time acquisition and visualization of freehand three-dimensional ultrasound.

Authors:  R W Prager; A Gee; L Berman
Journal:  Med Image Anal       Date:  1999-06       Impact factor: 8.545

2.  Focal hepatic ablation using interstitial photon radiation energy.

Authors:  L G Koniaris; D Y Chan; C Magee; S B Solomon; J H Anderson; D O Smith; T De Weese; L R Kavoussi; M A Choti
Journal:  J Am Coll Surg       Date:  2000-08       Impact factor: 6.113

3.  An integrated visualization system for surgical planning and guidance using image fusion and an open MR.

Authors:  D T Gering; A Nabavi; R Kikinis; N Hata; L J O'Donnell; W E Grimson; F A Jolesz; P M Black; W M Wells
Journal:  J Magn Reson Imaging       Date:  2001-06       Impact factor: 4.813

Review 4.  Three-dimensional ultrasound imaging.

Authors:  A Fenster; D B Downey; H N Cardinal
Journal:  Phys Med Biol       Date:  2001-05       Impact factor: 3.609

5.  Virtual remote center of motion control for needle placement robots.

Authors:  Emad M Boctor; Robert J Webster; Herve Mathieu; Allison M Okamura; Gabor Fichtinger
Journal:  Comput Aided Surg       Date:  2004

6.  Simulation and fully automatic multimodal registration of medical ultrasound.

Authors:  Wolfgang Wein; Ali Khamene; Dirk-André Clevert; Oliver Kutter; Nassir Navab
Journal:  Med Image Comput Comput Assist Interv       Date:  2007

7.  Local tumor recurrence following hepatic cryoablation: radiologic-histopathologic correlation in a rabbit model.

Authors:  B S Kuszyk; J K Boitnott; M A Choti; D A Bluemke; S Sheth; C A Magee; K M Horton; J Eng; E K Fishman
Journal:  Radiology       Date:  2000-11       Impact factor: 11.105

8.  Risk factors for distant recurrence of hepatocellular carcinoma in the liver after complete coagulation by microwave or radiofrequency ablation.

Authors:  N Izumi; Y Asahina; O Noguchi; M Uchihara; N Kanazawa; J Itakura; Y Himeno; S Miyake; T Sakai; N Enomoto
Journal:  Cancer       Date:  2001-03-01       Impact factor: 6.860

9.  Precision targeting of liver lesions using a novel electromagnetic navigation device in physiologic phantom and swine.

Authors:  Filip Banovac; Jonathan Tang; Sheng Xu; David Lindisch; Ho Young Chung; Elliot B Levy; Thomas Chang; Michael F McCullough; Ziv Yaniv; Bradford J Wood; Kevin Cleary
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

10.  Radiofrequency ablation of 231 unresectable hepatic tumors: indications, limitations, and complications.

Authors:  T F Wood; D M Rose; M Chung; D P Allegra; L J Foshag; A J Bilchik
Journal:  Ann Surg Oncol       Date:  2000-09       Impact factor: 5.344

View more
  15 in total

1.  Experimental evaluation of ultrasound-guided 3D needle steering in biological tissue.

Authors:  Momen Abayazid; Gustaaf J Vrooijink; Sachin Patil; Ron Alterovitz; Sarthak Misra
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-02-22       Impact factor: 2.924

2.  Sliding Mode Control of Steerable Needles.

Authors:  D Caleb Rucker; Jadav Das; Hunter B Gilbert; Philip J Swaney; Michael I Miga; Nilanjan Sarkar; Robert J Webster
Journal:  IEEE Trans Robot       Date:  2013-10       Impact factor: 5.567

3.  CASPER: computer-aided segmentation of imperceptible motion-a learning-based tracking of an invisible needle in ultrasound.

Authors:  Parmida Beigi; Robert Rohling; Septimiu E Salcudean; Gary C Ng
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-06-24       Impact factor: 2.924

4.  Three-dimensional ultrasound scanning.

Authors:  Aaron Fenster; Grace Parraga; Jeff Bax
Journal:  Interface Focus       Date:  2011-06-01       Impact factor: 3.906

5.  Spectral analysis of the tremor motion for needle detection in curvilinear ultrasound via spatiotemporal linear sampling.

Authors:  Parmida Beigi; Robert Rohling; Septimiu E Salcudean; Gary C Ng
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-04-08       Impact factor: 2.924

6.  Preliminary study on magnetic tracking-based planar shape sensing and navigation for flexible surgical robots in transoral surgery: methods and phantom experiments.

Authors:  Shuang Song; Changchun Zhang; Li Liu; Max Q-H Meng
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-10-05       Impact factor: 2.924

7.  Minimally invasive holographic surface scanning for soft-tissue image registration.

Authors:  Ray A Lathrop; Douglas M Hackworth; Robert J Webster
Journal:  IEEE Trans Biomed Eng       Date:  2010-06       Impact factor: 4.538

Review 8.  Catheter-based ultrasound technology for image-guided thermal therapy: current technology and applications.

Authors:  Vasant A Salgaonkar; Chris J Diederich
Journal:  Int J Hyperthermia       Date:  2015-03-23       Impact factor: 3.914

9.  Detection of Curved Robots using 3D Ultrasound.

Authors:  Hongliang Ren; Nikolay V Vasilyev; Pierre E Dupont
Journal:  Rep U S       Date:  2011-09-25

10.  Tubular Enhanced Geodesic Active Contours for Continuum Robot Detection using 3D Ultrasound.

Authors:  Hongliang Ren; Pierre E Dupont
Journal:  IEEE Int Conf Robot Autom       Date:  2012
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.