Radiation dose to patients and staff during angiography of the lower limbs. Derivation of local dose reference levels.

T Topaltzikis1, C Rountas, R Moisidou, I Fezoulidis, C Kappas, K Theodorou.   

Abstract

BACKGROUND: The Euratom directive 97/43 recommends the use of patient dose surveys in diagnostic radiology and the establishment of reference dose levels (DRLs).
PURPOSE: To perform measurements of the dose delivered during diagnostic angiography of the lower limbs using thermoluminescence dosimeters (TLDs), extraction of DRLs and estimation of the effective dose and radiation risk for this particular examination.
METHODS: Dose measurement was performed on 30 patients by using TLD sachets attached in 5 different positions not only on the patient, but also to the radiologist. All the appropriate factors were recorded. Measurement of the ESD was performed after each examination.
RESULTS: The mean entrance skin dose (ESD) was calculated to be 70.8, 67.7, 24.3, 18.4, 9.7 mGy at the level of aorta bifurcation, pelvis, femur, knees, and at feet, respectively. The average effective dose is 9.8 mSv with the radiation risks for fatal cancer to be 5.4 x 10(-4). The effective dose of the radiologist was calculated to be 0.023 mSv per procedure.
CONCLUSION: Radiation dose variation depends on the physical characteristics of the patient, on the procedure preferences by radiologists and the difficulties in conducting procedures. The main reason for the increased patient dose, compared to other studies, is the number of frames rather than the duration of fluoroscopy. For DSA of the lower limbs, the DRL was chosen to be an entrance skin dose of 96.4 mGy in the pelvic region. The dose to the radiologist is negligible.

Entities:  

Mesh:

Year:  2008        PMID: 18396433     DOI: 10.1016/j.ejmp.2008.02.005

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  Effective-dose estimation in interventional radiological procedures.

Authors:  Maria D Falco; Salvatore Masala; Matteo Stefanini; Paolo Bagalà; Daniele Morosetti; Eros Calabria; Alessia Tonnetti; Gianluca Verona-Rinati
Journal:  Radiol Phys Technol       Date:  2018-03-08

2.  Accuracy of multislice CT angiography for the assessment of in-stent restenoses in the iliac arteries at reduced dose: a phantom study.

Authors:  K Perisinakis; E Manousaki; K Zourari; D Tsetis; A Tzedakis; A Papadakis; A Karantanas; J Damilakis
Journal:  Br J Radiol       Date:  2011-03       Impact factor: 3.039

3.  Ionizing radiation-induced cataract in interventional cardiology staff.

Authors:  Ahmad Bitarafan Rajabi; Feridoun Noohi; Hassan Hashemi; Majid Haghjoo; Mohammad Miraftab; Nahid Yaghoobi; Fereydon Rastgou; Hadi Malek; Hoshang Faghihi; Hassan Firouzabadi; Soheila Asgari; Farhad Rezvan; Hamidreza Khosravi; Sara Soroush; Mehdi Khabazkhoob
Journal:  Res Cardiovasc Med       Date:  2015-01-22

4.  Patient skin dose measurements using a cable free system MOSFETs based in fluoroscopically guided percutaneous vertebroplasty, percutaneous disc decompression, radiofrequency medial branch neurolysis, and endovascular critical limb ischemia.

Authors:  Maria D Falco; Salvatore Masala; Matteo Stefanini; Roberto Fiori; Roberto Gandini; Paolo Bagalà; Daniele Morosetti; Eros Calabria; Alessia Tonnetti; Gianluca Verona-Rinati; Riccardo Santoni; Giovanni Simonetti
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.