B Fan1, E M Lewiecki, M Sherman, Y Lu, P D Miller, H K Genant, J A Shepherd.
Abstract
UNLABELLED: The precision of Hologic Apex v2.0 analysis software is significantly improved from Hologic Delphi v11.2 software and is comparable to GE Lunar Prodigy v7.5 software. Apex and Delphi precisions were, respectively, 1.0% vs. 1.2% (L1-L4 spine), 1.l % vs. 1.3% (total femur), 1.6% vs. 1.9% (femoral neck), and 0.7% vs. 0.9% (dual total femur).Entities:
Mesh:
Year: 2008 PMID: 18373054 PMCID: PMC5059410 DOI: 10.1007/s00198-008-0594-2
Source DB: PubMed Journal: Osteoporos Int ISSN: 0937-941X Impact factor: 4.507
Subject characteristics
| Mean ± SD | ||||
|---|---|---|---|---|
| Facility 1 | Facility 2 | Facility 3 | Pooled | |
|
| 28 | 29 | 30 | 87 |
| Age | 63.4 ± 9.2 | 64.1 ± 9.4 | 62.3 ± 9.3 | 63.0 ± 9.1 |
| Height | 160.9 ± 7.2 | 160.5 ± 7.5 | 159.6 ± 8.3 | 160.3 ± 7.5 |
| Weight | 64.0 ± 10.6 | 65.0 ± 16.1 | 68.0 ± 18.5 | 64.0 ± 15.3 |
Facility 1: New Mexico Clinical Research & Osteoporosis Center, Facility 2: Colorado Center for Bone Research, Facility 3: University of California at San Francisco BMD: bone mineral density
Comparison of hologic apex and delphi software results
| Slope*** (Delphi/Apex) | SEE*(g/cm2) | R2 | Mean difference, g/cm2 (%) | |
|---|---|---|---|---|
| Lumbar spine BMD | 1.007 | 0.014 | 0.99 | −0.006 (1.1)** |
| Left total hip BMD | 1.001 | 0.016 | 0.98 | −0.000(1.9) |
| Right total hip BMD | 0.999 | 0.016 | 0.98 | −0.000 (1.8) |
| Left neck BMD | 1.020 | 0.024 | 0.92 | −0.015(3.4)** |
| Right neck BMD | 1.014 | 0.024 | 0.92 | −0.012 (3.7)** |
| Lumbar spine BMC | 0.998 | 0.700 | 0.99 | 0.059(1.2)** |
| Left total hip BMC | 1.000 | 0.593 | 0.98 | 0.027(2.2) |
| Right total hip BMC | 0.998 | 0.635 | 0.97 | 0.062(2.2) |
| Left neck BMC | 1.026 | 0.144 | 0.90 | −0.094(4.1)** |
| Right neck BMC | 1.012 | 0.143 | 0.91 | −0.05(4.0)** |
*SEE: standard error of the estimate
** Significantly different than 0 (p-value < 0.005)
*** No intercepts were statistically different that 0. Thus, the slope is with an intercept = 0
Fig. 1Bland-Altman plots of lumbar spine BMD Hologic Apex and Delphi software. The solid line is the best fit line with the 95% confidence limits shown as dotted lines
Fig. 2Bland-Altman plots of left and right femoral neck BMD Hologic Apex and Delphi software. The solid line is the best fit line with the 95% confidence limits shown as dotted lines
Fig. 3Bland-Altman plots of left and right total hip BMD Hologic Apex and Delphi software. The solid line is the best fit line with the 95% confidence limits shown as dotted lines
Fig. 4Comparing precision of Hologic Apex and Delphi software for lumbar spine and left and right hip regions
RMS standard deviation (RMS SD) and coefficient of variation (RMS CV) of hologic apex and GE healthcare lunar prodigy software
| RMS SD (CV) in sBMD units, mg/cm2 | ||||
|---|---|---|---|---|
| Facility 1 | Facility 2 | Facility 3 | Pooled | |
| Hologic Apex | ||||
| Lumbar spine BMD | 9.9 (1.1%) | 8.7 (0.8%) | 11.4 (1.1%) | 10.0 (1.0%) |
| Left femoral neck BMD | 10.8 (1.3%) | 14.2 (1.7%) | 14.3 (1.7%) | 13.4 (1.6%) |
| Right femoral neck BMD | 18.3 (2.4%) | 17.0 (2.1%) | 19.5 (2.3%) | 18.4 (2.3%) |
| Left total hip BMD | 5.5 (0.8%) | 7.9 (0.9%) | 11.4 (1.4%) | 8.6 (1.1%) |
| Right total hip BMD | 8.1 (1.0%) | 7.0 (0.8%) | 7.9 (0.9%) | 7.6 (0.9%) |
| Dual total hip BMD | 5.2 (0.6%) | 5.3 (0.6%) | 7.5 (0.9%) | 6.1 (0.7%) |
| GE Lunar Prodigy | ||||
| Lumbar spine BMD | 10.2 (1.0%) | 10.7 (1.0%) | 9.1 (0.9%) | 10.0 (1.0%) |
| Left femoral neck BMD | 10.8 (1.5%) | 13.8 (1.7%) | 10.9 (1.3%) | 11.9 (1.5%) |
| Right femoral neck BMD | 15.3 (2.1%) | 12.5 (1.5%) | 14.1 (1.7%) | 14.0 (1.8%) |
| Left total hip BMD | 6.9 (1.0%) | 9.1 (1.0%) | 6.5 (0.8%) | 7.6 (0.9%) |
| Right total hip BMD | 8.0 (1.0%) | 7.2 (0.8%) | 7.6 (0.9%) | 7.6 (0.9%) |
| Dual total hip BMD | 5.0 (0.6%) | 6.6 (0.8%) | 5.4 (0.6%) | 5.7 (0.6%) |