Literature DB >> 18372456

Radiation dose reduction strategy for CT protocols: successful implementation in neuroradiology section.

Alice B Smith1, William P Dillon, Benison C Lau, Robert Gould, Francis R Verdun, Edward B Lopez, Max Wintermark.   

Abstract

PURPOSE: To retrospectively quantify the effect of systematic use of tube current modulation for neuroradiology computed tomographic (CT) protocols on patient dose and image quality.
MATERIALS AND METHODS: This HIPAA-compliant study had institutional review board approval, with waiver of informed consent. The authors evaluated the effect of dose modulation on four types of neuroradiologic CT studies: brain CT performed without contrast material (unenhanced CT) in adult patients, unenhanced brain CT in pediatric patients, adult cervical spine CT, and adult cervical and intracranial CT angiography. For each type of CT study, three series of 100 consecutive studies were reviewed: 100 studies performed without dose modulation, 100 studies performed with z-axis dose modulation, and 100 studies performed with x-y-z-axis dose modulation. For each examination, the weighted volume CT dose index (CTDI(vol)) and dose-length product (DLP) were recorded and noise was measured. Each study was also reviewed for image quality. Continuous variables (CTDI(vol), DLP, noise) were compared by using t tests, and categorical variables (image quality) were compared by using Wilcoxon rank-sum tests.
RESULTS: For unenhanced CT of adult brains, the CTDI(vol) and DLP, respectively, were reduced by 60.9% and 60.3%, respectively, by using z-axis dose modulation and by 50.4% and 22.4% by using x-y-z-axis dose modulation. Significant dose reductions (P < .001) were also observed for pediatric unenhanced brain CT, cervical spine CT, and adult cervical and intracranial CT angiography performed with each dose modulation technique. Image quality and noise were unaffected by the use of either dose modulation technique (P > .05).
CONCLUSION: Use of dose-modulation techniques for neuroradiology CT examinations affords significant dose reduction while image quality is maintained. (c) RSNA, 2008.

Entities:  

Mesh:

Year:  2008        PMID: 18372456     DOI: 10.1148/radiol.2472071054

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  29 in total

1.  The use of adaptive statistical iterative reconstruction in pediatric head CT: a feasibility study.

Authors:  G A Vorona; G Zuccoli; T Sutcavage; B L Clayton; R C Ceschin; A Panigrahy
Journal:  AJNR Am J Neuroradiol       Date:  2012-05-24       Impact factor: 3.825

2.  Short- and long-term effects of clinical audits on compliance with procedures in CT scanning.

Authors:  Antonio Oliveri; Nigel Howarth; Pierre Alain Gevenois; Denis Tack
Journal:  Eur Radiol       Date:  2015-11-14       Impact factor: 5.315

3.  Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.

Authors:  Han Chen; Cheng Xu; Mats Persson; Mats Danielsson
Journal:  J Med Imaging (Bellingham)       Date:  2015-11-06

Review 4.  Imaging of accidental paediatric head trauma.

Authors:  Phua Hwee Tang; Choie Cheio Tchoyoson Lim
Journal:  Pediatr Radiol       Date:  2009-01-06

5.  Neuroradiology back to the future: spine imaging.

Authors:  E G Hoeffner; S K Mukherji; A Srinivasan; D J Quint
Journal:  AJNR Am J Neuroradiol       Date:  2012-05-10       Impact factor: 3.825

6.  Interruption of the aortic arch: diagnosis with multidetector computed tomography.

Authors:  Shuhei Sato; Noriaki Akagi; Mayu Uka; Katsuya Kato; Yoshihiro Okumura; Susumu Kanazawa
Journal:  Jpn J Radiol       Date:  2011-01-26       Impact factor: 2.374

7.  Radiation dose reduction in CT-guided spine biopsies does not reduce diagnostic yield.

Authors:  K A Shpilberg; B N Delman; L N Tanenbaum; S J Esses; R Subramaniam; A H Doshi
Journal:  AJNR Am J Neuroradiol       Date:  2014-07-17       Impact factor: 3.825

8.  Standardization and optimization of CT protocols to achieve low dose.

Authors:  Sigal Trattner; Gregory D N Pearson; Cynthia Chin; Dianna D Cody; Rajiv Gupta; Christopher P Hess; Mannudeep K Kalra; James M Kofler; Mayil S Krishnam; Andrew J Einstein
Journal:  J Am Coll Radiol       Date:  2014-03       Impact factor: 5.532

Review 9.  Multidetector CT in children: current concepts and dose reduction strategies.

Authors:  Rutger A J Nievelstein; Ingrid M van Dam; Aart J van der Molen
Journal:  Pediatr Radiol       Date:  2010-06-10

10.  Adenosine-stress dynamic myocardial perfusion imaging using 128-slice dual-source CT: optimization of the CT protocol to reduce the radiation dose.

Authors:  Sung Mok Kim; Yoo Na Kim; Yeon Hyeon Choe
Journal:  Int J Cardiovasc Imaging       Date:  2012-10-18       Impact factor: 2.357

View more