Literature DB >> 18361986

Correlation between ciliary sulcus diameter measured by 35 MHz ultrasound biomicroscopy and other ocular measurements.

Kyun-Hyung Kim1, Hyung-Ho Shin, Hyo-Myung Kim, Jong-Suk Song.   

Abstract

PURPOSE: To measure horizontal ciliary sulcus diameters by 35 MHz ultrasound biomicroscopy (UBM) and evaluate correlations between these and other ocular measurements.
SETTING: Department of Ophthalmology, Korea University College of Medicine, Seoul, Republic of Korea.
METHODS: A pilot group comprising 20 eyes of 10 patients was enrolled in a prospective trial. Horizontal sulcus diameter was measured using a 35 MHz UBM unit; other ocular measurements (eg, corneal curvature, refractive error, white-to-white [WTW] distance, anterior chamber depth [ACD]) were also taken in all patients. Linear regression analysis was used to evaluate the correlation between horizontal sulcus diameter and the other measurements. A regression equation was determined as a result of the analysis and applied to a second validation group (14 eyes).
RESULTS: The mean horizontal diameter of the ciliary sulcus in the pilot group was 11.56 mm +/- 0.64 (SD). The mean corneal curvature was significantly and negatively correlated with the horizontal sulcus diameter (P<.001). Refractive error, ACD, and horizontal WTW diameter were not correlated with the horizontal sulcus diameter by multiple regression analysis. In the validation group, sulcus diameters estimated using the regression equation were significantly correlated with actual values, and these 2 sets of values were not statistically different.
CONCLUSIONS: Mean corneal curvature was significantly and negatively correlated with horizontal sulcus diameter. The derived regression formula based on mean corneal curvature measurements may help predict sulcus diameters for proper sizing of posterior chamber phakic intraocular lenses.

Entities:  

Mesh:

Year:  2008        PMID: 18361986     DOI: 10.1016/j.jcrs.2007.11.034

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Outcomes and possible risk factors associated with axis alignment and rotational stability after implantation of the Toric implantable collamer lens for high myopic astigmatism.

Authors:  Xun-Lun Sheng; Wei-Ning Rong; Qin Jia; Ya-Ni Liu; Wen-Juan Zhuang; Qing Gu; Yan Sun; Bo Pan; De-Jun Zhu
Journal:  Int J Ophthalmol       Date:  2012-08-18       Impact factor: 1.779

2.  Angle-to-angle and spur-to-spur distance analysis with high-resolution optical coherence tomography.

Authors:  Robert Montés-Micó; Pedro Tañá-Rivero; Salvador Aguilar-Córcoles; María Ruiz-Santos; María Dolores Rodríguez-Carrillo; Ramón Ruiz-Mesa
Journal:  Eye Vis (Lond)       Date:  2020-08-16

3.  Haptic breakage after transscleral fixation of a single-piece acrylic intraocular lens.

Authors:  Hiroki Tanaka; Kaoru Fujinami; Ken Watanabe; Toru Noda; Kunihiko Akiyama
Journal:  Case Rep Ophthalmol       Date:  2014-07-12

4.  Comparative analysis of the visual performance after implantation of the toric implantable collamer lens in stable keratoconus: a 4-year follow-up after sequential procedure (CXL+TICL implantation).

Authors:  Farideh Doroodgar; Feazollah Niazi; Azad Sanginabadi; Sana Niazi; Alireza Baradaran-Rafii; Cyrus Alinia; Eznollah Azargashb; Mohammad Ghoreishi
Journal:  BMJ Open Ophthalmol       Date:  2017-09-28

5.  Effects of optical diameter of intraocular lenses with intrascleral fixation on higher-order aberrations.

Authors:  Daisuke Kunita; Makoto Inoue; Yuji Itoh; Naoko Matsuki; Toshiyuki Nagamoto; Akito Hirakata
Journal:  BMC Ophthalmol       Date:  2017-06-02       Impact factor: 2.209

6.  Adjustment formulae to improve the correlation of white-to-white measurement with direct measurement of the ciliary sulcus diameter by ultrasound biomicroscopy.

Authors:  Seyed Javad Hashemian; Majid Mohebbi; Mehdi Yaseri; Mohammad Ebrahim Jafari; Shadrokh Nabili; Seyed Mahyar Hashemian; Mahsa Sadat Hashemian
Journal:  J Curr Ophthalmol       Date:  2017-12-16

7.  Visual Outcomes, Footplate Position and Vault Achieved with the Visian Implantable Collamer Lens for Myopic Astigmatism.

Authors:  Dan Z Reinstein; Ryan S Vida; Timothy J Archer
Journal:  Clin Ophthalmol       Date:  2021-11-20

8.  Analysis of intraocular positions of posterior implantable collamer lens by full-scale ultrasound biomicroscopy.

Authors:  Xi Zhang; Xun Chen; Xiaoying Wang; Fei Yuan; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2018-05-09       Impact factor: 2.209

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.