Comparison of the MTI Photoscreener and the Welch-Allyn SureSight autorefractor in a tertiary care center.

David L Rogers1, Daniel E Neely, Janice B Chapman, David A Plager, Derek T Sprunger, Naval Sondhi, Gavin J Roberts, Susan Ofner.   

Abstract

INTRODUCTION: The MTI photoscreener (MTI) and the Welch-Allyn SureSight autorefractor are commonly used for preschool vision screening. We compared both of these methods on 100 consecutive patients in a prospective, randomized, masked, clinical trial conducted at a tertiary care center.
METHODS: One hundred patients between 1 and 6 years of age were included in the study. All participants underwent a comprehensive eye examination with cycloplegic refraction. Examination failure analysis was done on the SureSight data using the manufacturer's referral criteria, the Vision in Preschoolers study (VIP) 90% specificity criteria, the VIP 94% specificity criteria, and the referral criteria proposed by Rowatt and colleagues.
RESULTS: Data were successfully obtained on 76% of children using the SureSight and 96% with the MTI. The sensitivity and specificity of the SureSight to detect clinically significant amblyogenic factors using the manufacturer's criteria was 96.6 and 38.1%, using the VIP 90% criteria was 79.3 and 64.3%, using the VIP 94% criteria was 67.2 and 69.0%, and using criteria proposed by Rowatt and colleagues was 62.1 and 73.8%. The sensitivity and specificity of the MTI photoscreener was 94.8 and 88.1%, respectively.
CONCLUSIONS: Using the manufacturer's referral criteria, the SureSight had a sensitivity equal to the MTI photoscreener; however, the specificity was low and over-referrals were anticipated. As specificity levels were increased, a substantial number of children with amblyogenic risk factors were not appropriately identified within our study population..

Entities:  

Mesh:

Year:  2008        PMID: 18314072     DOI: 10.1016/j.jaapos.2007.09.004

Source DB:  PubMed          Journal:  J AAPOS        ISSN: 1091-8531            Impact factor:   1.220


  6 in total

1.  Receding and disparity cues aid relaxation of accommodation.

Authors:  Anna M Horwood; Patricia M Riddell
Journal:  Optom Vis Sci       Date:  2009-11       Impact factor: 1.973

2.  Accuracy of the Welch Allyn SureSight for measurement of magnitude of astigmatism in 3- to 7-year-old children.

Authors:  Erin M Harvey; Velma Dobson; Joseph M Miller; Candice E Clifford-Donaldson; Tina K Green; Dawn H Messer; Katherine A Garvey
Journal:  J AAPOS       Date:  2009-10       Impact factor: 1.220

3.  Beyond screening for risk factors: objective detection of strabismus and amblyopia.

Authors:  Reed M Jost; Susan E Yanni; Cynthia L Beauchamp; David R Stager; David Stager; Lori Dao; Eileen E Birch
Journal:  JAMA Ophthalmol       Date:  2014-07       Impact factor: 7.389

4.  Impact of confidence number on accuracy of the SureSight Vision Screener.

Authors: 
Journal:  Optom Vis Sci       Date:  2010-02       Impact factor: 1.973

Review 5.  Advantages, limitations, and diagnostic accuracy of photoscreeners in early detection of amblyopia: a review.

Authors:  Irene Sanchez; Sara Ortiz-Toquero; Raul Martin; Victoria de Juan
Journal:  Clin Ophthalmol       Date:  2016-07-22

6.  Smartphone photography for screening amblyogenic conditions in children.

Authors:  Richa Gupta; Siddharth Agrawal; Rajat M Srivastava; Vinita Singh; Vishal Katiyar
Journal:  Indian J Ophthalmol       Date:  2019-10       Impact factor: 1.848

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.