Literature DB >> 18310687

Radial head arthroplasty with a modular metal spacer to treat acute traumatic elbow instability. Surgical technique.

David Ring1, Graham King.   

Abstract

BACKGROUND: The use of a metal radial head prosthesis to help stabilize an elbow with traumatic instability is appealing because internal fixation of multifragment, displaced fractures of the radial head is susceptible to either early or late failure. The newer modular prostheses are easier to size and implant, but their effectiveness has not been investigated, to our knowledge.
METHODS: Twenty-seven patients in whom a radial head replacement with a modular metal spacer prosthesis had been performed to treat traumatic elbow instability were evaluated with use of the Mayo Elbow Performance Index (MEPI), the American Shoulder and Elbow Surgeons Elbow Evaluation Instrument (ASES), and the Disabilities of the Arm, Shoulder and Hand questionnaire (DASH). Radiographs were evaluated for arthrosis, periprosthetic radiolucency, and heterotopic ossification.
RESULTS: Seven patients underwent one or more subsequent operations to treat residual instability, heterotopic ossification and elbow contracture, ulnar neuropathy, or a misplaced screw. In two of these patients, the prosthesis was removed as part of an elbow contracture release or to treat infection. At an average of forty months postoperatively, elbow motion in the entire group of twenty-seven patients averaged 131 degrees of flexion with a 20 degrees flexion contracture, 73 degrees of pronation, and 57 degrees of supination. Stability was restored to all twenty-seven elbows, and twenty-two patients had a good or excellent result according to the MEPI. Seventeen patients had radiographic evidence of lucency around the neck of the prosthesis that was not associated with increased pain, thirteen patients had clinically inconsequential heterotopic ossification anterior to the radial neck, and nine patients had radiographic changes in the capitellum.
CONCLUSIONS: An intentionally loosely placed modular metal radial head prosthesis can help to restore stability in conjunction with repair of other fractures and reattachment of the lateral collateral ligament to the epicondyle in the setting of traumatic elbow instability with a comminuted fracture of the radial head. While a prosthesis that is too large can cause problems, lucencies around the stem of the intentionally loose prosthesis and most changes in the capitellum do not appear to cause problems, at least in the short term.

Entities:  

Mesh:

Year:  2008        PMID: 18310687     DOI: 10.2106/JBJS.G.01248

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  11 in total

1.  Is a silastic radial head still a reasonable option?

Authors:  Yariv Maghen; Andrew J Leo; Jennifer W Hsu; Michael R Hausman
Journal:  Clin Orthop Relat Res       Date:  2011-04       Impact factor: 4.176

2.  [Prosthesis after comminuted radial head fractures : midterm results].

Authors:  A Moghaddam; A Lennert; S Studier-Fischer; A Wentzensen; G Zimmermann
Journal:  Unfallchirurg       Date:  2008-12       Impact factor: 1.000

3.  Simple overlay device for determining radial head and neck height.

Authors:  Jun-Gyu Moon; Richard D Southgate; James S Fitzsimmons; Shawn W O'Driscoll
Journal:  Skeletal Radiol       Date:  2010-03-01       Impact factor: 2.199

4.  Do the radial head prosthesis components fit with the anatomical structures of the proximal radioulnar joint?

Authors:  Kilian Wegmann; Moritz K Hain; Christian Ries; Wolfram F Neiss; Lars P Müller; Klaus J Burkhart
Journal:  Surg Radiol Anat       Date:  2014-12-28       Impact factor: 1.246

Review 5.  Radial head fractures.

Authors:  R P van Riet; Mpj van den Bekerom; A Van Tongel; C Spross; R Barco; A C Watts
Journal:  Shoulder Elbow       Date:  2019-09-25

6.  Radial head replacement for acute complex fractures: what are the rate and risks factors for revision or removal?

Authors:  Andrew D Duckworth; Neil R Wickramasinghe; Nicholas D Clement; Charles M Court-Brown; Margaret M McQueen
Journal:  Clin Orthop Relat Res       Date:  2014-07       Impact factor: 4.176

7.  Single-staged treatment using a standardized protocol results in functional motion in the majority of patients with a terrible triad elbow injury.

Authors:  Akash Gupta; David Barei; Ansab Khwaja; Daphne Beingessner
Journal:  Clin Orthop Relat Res       Date:  2014-07       Impact factor: 4.176

8.  Radial head arthroplasty with an uncemented modular metallic radial head prosthesis: short- and mid-term results.

Authors:  M Muhm; R de Castro; H Winkler
Journal:  Eur J Trauma Emerg Surg       Date:  2010-09-23       Impact factor: 3.693

9.  Is the distal radioulnar variance useful for identification of radial head prosthesis overlengthening? A cadaver study.

Authors:  Kilian Wegmann; Wei Zhang; Volker Strauss; Christian Ries; Lars P Müller; Klaus J Burkhart
Journal:  Eur J Med Res       Date:  2016-03-21       Impact factor: 2.175

10.  Outcome of Radial Head Arthroplasty in Comminuted Radial Head Fractures: Short and Midterm Results.

Authors:  Arash Moghaddam; Tim Friedrich Raven; Eike Dremel; Stefan Studier-Fischer; Paul Alfred Grutzner; Bahram Biglari
Journal:  Trauma Mon       Date:  2016-02-06
View more