Literature DB >> 18309173

Mechanical thromboembolectomy for acute ischemic stroke: comparison of the catch thrombectomy device and the Merci Retriever in vivo.

Caspar Brekenfeld1, Gerhard Schroth, Marwan El-Koussy, Krassen Nedeltchev, Michael Reinert, Johannes Slotboom, Jan Gralla.   

Abstract

BACKGROUND AND
PURPOSE: The purpose of the study was to compare efficacy and potential complications of 2 commercially available devices for mechanical thromboembolectomy.
METHODS: Devices were tested in an established animal model allowing the use of routine angiography catheters and thrombectomy devices. Radio-opaque thrombi were used for visualization of thrombus-device interaction during angiography. The Merci Retrieval System and the Catch Thromboembolectomy System were assessed each in 10 vessel occlusions. For every occluded vessel up to 5 retrieval attempts were performed.
RESULTS: Sufficient recanalization was achieved with the Merci Retriever in 90% of occlusions, and with the Catch device recanalization was achieved in 70% of occlusions. Recanalization at the first attempt occurred significantly more often with the Merci Retriever compared to the Catch device (OR, 21; 95% CI, 1.78-248.11). Consequently, significantly more attempts (P=0.02) had to be performed with the Catch device; therefore, time to recanalization was longer. Thrombus fragmentations during retrieval were caused more often by the Catch device compared to the Merci Retriever (OR, 15.6; 95% CI, 1.73-140.84), resulting in a higher distal embolization rate. During retrieval both devices lost thrombotic material at the tip of the guide catheter, which was then aspirated in most cases.
CONCLUSIONS: Both distal devices are effective for thromboembolectomy. To avoid loss of thrombotic material and distal embolization, the use of large luminal balloon guide catheters and aspiration during retrieval seems to be mandatory. The design of the Merci Retriever appears to be more efficient during thrombus mobilization and retrieval with less fragmentation compared to the Catch Thromboembolectomy System.

Entities:  

Mesh:

Year:  2008        PMID: 18309173     DOI: 10.1161/STROKEAHA.107.495614

Source DB:  PubMed          Journal:  Stroke        ISSN: 0039-2499            Impact factor:   7.914


  26 in total

1.  The impact of thromboemboli histology on the performance of a mechanical thrombectomy device.

Authors:  I Yuki; I Kan; H V Vinters; R H Kim; A Golshan; F A Vinuela; J W Sayre; Y Murayama; F Vinuela
Journal:  AJNR Am J Neuroradiol       Date:  2011-12-29       Impact factor: 3.825

2.  Solitaire flow-restoration device for treatment of acute ischemic stroke: safety and recanalization efficacy study in a swine vessel occlusion model.

Authors:  R Jahan
Journal:  AJNR Am J Neuroradiol       Date:  2010-07-15       Impact factor: 3.825

3.  In vivo evaluation of the first dedicated combined flow-restoration and mechanical thrombectomy device in a swine model of acute vessel occlusion.

Authors:  P Mordasini; N Frabetti; J Gralla; G Schroth; U Fischer; M Arnold; C Brekenfeld
Journal:  AJNR Am J Neuroradiol       Date:  2010-10-21       Impact factor: 3.825

4.  In vivo evaluation of the Phenox CRC mechanical thrombectomy device in a swine model of acute vessel occlusion.

Authors:  P Mordasini; M Hiller; C Brekenfeld; G Schroth; U Fischer; J Slotboom; M Arnold; J Gralla
Journal:  AJNR Am J Neuroradiol       Date:  2009-12-17       Impact factor: 3.825

5.  Experimental evaluation of immediate recanalization effect and recanalization efficacy of a new thrombus retriever for acute stroke treatment in vivo.

Authors:  P Mordasini; C Brekenfeld; J V Byrne; U Fischer; M Arnold; S Jung; G Schroth; J Gralla
Journal:  AJNR Am J Neuroradiol       Date:  2012-07-26       Impact factor: 3.825

6.  Histopathologic evaluation of arterial wall response to 5 neurovascular mechanical thrombectomy devices in a swine model.

Authors:  B Gory; D Bresson; I Kessler; M L Perrin; A Guillaudeau; K Durand; S Ponsonnard; C Couquet; C Yardin; C Mounayer
Journal:  AJNR Am J Neuroradiol       Date:  2013-03-28       Impact factor: 3.825

Review 7.  New stent retriever devices.

Authors:  N C Beadell; Helmi Lutsep
Journal:  Curr Atheroscler Rep       Date:  2013-06       Impact factor: 5.113

8.  Impact of retrievable stents on acute ischemic stroke treatment.

Authors:  C Brekenfeld; G Schroth; P Mordasini; U Fischer; M-L Mono; A Weck; M Arnold; M El-Koussy; J Gralla
Journal:  AJNR Am J Neuroradiol       Date:  2011-05-12       Impact factor: 3.825

9.  Impact of Merci device fracture on clinical outcome of acute ischemic stroke after mechanical thrombectomy.

Authors:  Zhong-Song Shi; Gary R Duckwiler; Yince Loh; David S Liebeskind; Nestor R Gonzalez; Satoshi Tateshima; Reza Jahan; Jeffrey L Saver; Fernando Viñuela
Journal:  CNS Neurosci Ther       Date:  2012-08-20       Impact factor: 5.243

10.  Frequency and relevance of anterior cerebral artery embolism caused by mechanical thrombectomy of middle cerebral artery occlusion.

Authors:  W Kurre; K Vorlaender; M Aguilar-Pérez; E Schmid; H Bäzner; H Henkes
Journal:  AJNR Am J Neuroradiol       Date:  2013-03-07       Impact factor: 3.825