Cardiac output measurement in children: comparison of Aesculon cardiac output monitor and thermodilution.

M Tomaske1, W Knirsch, O Kretschmar, K Woitzek, C Balmer, A Schmitz, U Bauersfeld, M Weiss.   

Abstract

BACKGROUND: We compared cardiac output (CO) measurements by the non-invasive electrical velocimetry (Aesculon) monitor with the pulmonary artery catheter (PAC) thermodilution method in children.
METHODS: CO values using the Aesculon monitor and PAC thermodilution were simultaneously recorded during cardiac catheterization in children. Measurements were performed under general anaesthesia. To compare, three consecutive measurements for each patient within 3 min were obtained. The means of the three values were compared using simple regression and Bland-Altman analysis. Data were presented as mean (sd). A mean percentage of <30% was defined to indicate clinical useful reliability of the Aesculon monitor.
RESULTS: A total of 50 patients with a median (range) age of 7.5 (0.5-16.5) yr were enrolled in the study. Mean CO values were 3.7 (1.5) litre min(-1) (PAC thermodilution) and 3.1 (1.7) litre min(-1) (Aesculon) monitor). Analysis for CO measurement showed a good correlation between the two methods (r=0.894; P<0.0001). The bias between the two methods was 0.66 litre min(-1) with a precision of 1.49 litre min(-1). The mean percentage error for CO measurements was 48.9% for the Aesculon monitor when compared with PAC thermodilution.
CONCLUSIONS: Electrical velocimetry using the Aesculon monitor did not provide reliable CO values when compared with PAC thermodilution. Whether the Aesculon monitor can be used as a CO trend monitor has to be assessed by further investigations in patients with changing haemodynamics.

Entities:  

Mesh:

Year:  2008        PMID: 18305081     DOI: 10.1093/bja/aen024

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  21 in total

1.  Changes in cardiac output and stroke volume as measured by non-invasive CO monitoring in infants with RSV bronchiolitis.

Authors:  Julie Caplow; Sarah C McBride; Garry M Steil; Jackson Wong
Journal:  J Clin Monit Comput       Date:  2012-04-24       Impact factor: 2.502

Review 2.  Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis.

Authors:  Koichi Suehiro; Alexandre Joosten; Linda Suk-Ling Murphy; Olivier Desebbe; Brenton Alexander; Sang-Hyun Kim; Maxime Cannesson
Journal:  J Clin Monit Comput       Date:  2015-08-29       Impact factor: 2.502

3.  A comparative evaluation of electrical velocimetry and inert gas rebreathing for the non-invasive assessment of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Ursula Hoffmann; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  Clin Res Cardiol       Date:  2011-07-01       Impact factor: 5.460

4.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

5.  Propofol effect on cerebral oxygenation in children with congenital heart disease.

Authors:  Thilo Fleck; Stephan Schubert; Peter Ewert; Brigitte Stiller; Nicole Nagdyman; Felix Berger
Journal:  Pediatr Cardiol       Date:  2014-10-14       Impact factor: 1.655

6.  Hemodynamic alterations recorded by electrical cardiometry during ligation of ductus arteriosus in preterm infants.

Authors:  Reyin Lien; Kai-Hsiang Hsu; Jaw-Ji Chu; Yu-Sheng Chang
Journal:  Eur J Pediatr       Date:  2014-10-15       Impact factor: 3.183

7.  Comparison of electrical velocimetry and cardiac magnetic resonance imaging for the non-invasive determination of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Christina Doesch; Theano Papavassiliu; Stefan O Schoenberg; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  J Clin Monit Comput       Date:  2015-06-27       Impact factor: 2.502

8.  Influence of patent ductus arteriosus and ventilators on electrical velocimetry for measuring cardiac output in very-low/low birth weight infants.

Authors:  T Torigoe; S Sato; Y Nagayama; T Sato; H Yamazaki
Journal:  J Perinatol       Date:  2015-01-29       Impact factor: 2.521

9.  Anesthetic Management of a Patient With Emanuel Syndrome.

Authors:  Masanori Tsukamoto; Takashi Hitosugi; Kanako Esaki; Takeshi Yokoyama
Journal:  Anesth Prog       Date:  2016

10.  Effect of patent ductus arteriosus and patent foramen ovale on left ventricular stroke volume measurement by electrical velocimetry in comparison to transthoracic echocardiography in neonates.

Authors:  Martin Ernst Blohm; Jana Hartwich; Denise Obrecht; Jan Felix Kersten; Dominique Singer
Journal:  J Clin Monit Comput       Date:  2016-04-12       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.