Tissue response to surgical energy devices.

Courtney K Phillips1, Gregory W Hruby, Evren Durak, Daniel S Lehman, Peter A Humphrey, Mahesh M Mansukhani, Jaime Landman.   

Abstract

OBJECTIVES: There is little data available on the effects of energy-based surgical devices (ESD) on tissues other than arteries and veins. As such, we quantified the lateral thermal damage associated with contemporary ESD: the Harmonic ACE, a prototype bipolar device known as the Gyrus Trisector, the Harmonic LCS-C5, and the LigaSure V.
METHODS: We divided 24 domestic pigs into 4 groups, 1 group for each ESD tested. Segments of bladder, stomach, small bowel, colon, ureter, peritoneum, arteries, and veins were exposed to each ESD. The tissues were stained with hematoxylin and eosin and evaluated by an experienced pathologist to quantitate the lateral energy spread associated with each device. We measured blade temperatures of each device using the IR-Flex thermal camera.
RESULTS: The Trisector developed the lowest mean blade temperature (97.84 degrees F), whereas the LigaSure's was the second lowest (103.14 degrees F). The ACE and LCS-C5 created the highest blade temperatures, measuring 220.5 degrees F and 205.6 degrees F, respectively. The Trisector's mean full thickness and superficial lateral energy damage were 6.3 mm and 7.0 mm, respectively, whereas the Ligasure's was 4.5 mm and 5.9 mm, respectively. For the ACE, however, mean full thickness and superficial energy spread were 2.4 mm and 2.8 mm, respectively, whereas the LCS-C5's were 3.1 mm and 4.3 mm, respectively.
CONCLUSIONS: The Harmonic ACE and LCS-C5 produced the least thermal damage in the tissues tested. ESD-associated tissue energy damage is not directly related to blade temperature, but is likely the result of several factors including blade temperature, transection time, tissue properties, and the vascularity of each transected tissue.

Entities:  

Mesh:

Year:  2008        PMID: 18289646     DOI: 10.1016/j.urology.2007.11.035

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  25 in total

1.  Heat production, nerve function, and morphology following nerve close dissection with surgical instruments.

Authors:  Johan Carlander; C Koch; L Brudin; C Nordborg; O Gimm; K Johansson
Journal:  World J Surg       Date:  2012-06       Impact factor: 3.352

2.  BiClamp forceps significantly shorten the operation time for breast surgery.

Authors:  Hideki Takeuchi; Toshiaki Saeki; Takashi Shigekawa; Hiroshi Sano; Noriko Nakamiya; Kazuo Matsuura; Misono Misumi; Takao Takahashi; Nobuko Fujiuchi; Katsuhiko Okubo; Akihiko Osaki; Takaki Sakurai; Isamu Koyama
Journal:  Surg Today       Date:  2010-07-30       Impact factor: 2.549

3.  Thyroidectomy with LigaSure.

Authors:  Mile Ignjatović; Zoran Kostić
Journal:  Surg Today       Date:  2011-05-28       Impact factor: 2.549

4.  Histopathological Response and Adhesion Formation After Omentectomy with Ultrasonic Energy, Bipolar Sealing, and Suture Ligation.

Authors:  Gultekin Ozan Kucuk; Metin Ertem; Nuray Kepil
Journal:  Indian J Surg       Date:  2013-11-11       Impact factor: 0.656

5.  Optical coherence tomography for use in infrared laser sealing of blood vessels.

Authors:  Nicholas C Giglio; Thomas C Hutchens; Christopher M Cilip; Nathaniel M Fried
Journal:  IEEE Photonics Conf       Date:  2020-11-16

6.  Athermal versus ultrasonic nerve-sparing laparoscopic radical prostatectomy: a comparison of functional and oncological outcomes.

Authors:  Vincenzo Pagliarulo; Stefano Alba; Maria Filomena Gallone; Marcello Zingarelli; Alfonso Lorusso; Paolo Minafra; Giuseppe Maria Ludovico; Savino Di Stasi; Pasquale Ditonno
Journal:  World J Urol       Date:  2020-08-01       Impact factor: 4.226

7.  Ultrasonic versus monopolar energy-based surgical devices in terms of surgical smoke and lateral thermal damage (ULMOST): a randomized controlled trial.

Authors:  Chahien Choi; In-Gu Do; Taejong Song
Journal:  Surg Endosc       Date:  2018-04-09       Impact factor: 4.584

Review 8.  Large bowel injuries during gynecological laparoscopy.

Authors:  Kahraman Ulker; Turgut Anuk; Murat Bozkurt; Yetkin Karasu
Journal:  World J Clin Cases       Date:  2014-12-16       Impact factor: 1.337

Review 9.  Common uses and cited complications of energy in surgery.

Authors:  Ganesh Sankaranarayanan; Rajeswara R Resapu; Daniel B Jones; Steven Schwaitzberg; Suvranu De
Journal:  Surg Endosc       Date:  2013-04-23       Impact factor: 4.584

10.  Novel Optical Linear Beam Shaping Designs for use in Laparoscopic Laser Sealing of Vascular Tissues.

Authors:  Thomas C Hutchens; Nicholas C Giglio; Christopher M Cilip; Sarah G Rosenbury; Luke A Hardy; Duane E Kerr; William H Nau; Nathaniel M Fried
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2020-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.