S G Foster1, P M Embree, W R O'Brien.
Abstract
An ultrasonic human-blood-flow velocity profile measurement method using time-domain correlation of consecutive echo pairs has been developed. The time shift between a pair of range gated echoes is estimated by searching for the shift that results in the maximum correlation. The time shift indicates the distance a group of scatterers has moved, from which flow velocity is estimated. The basis for the computer simulations and error analyses of the scheme includes a band-passed white Gaussian noise signal model for an echo from a scattering medium, the estimate of flow velocity from both a single scatterer and multiple scatterers, and a derived precision estimation. The error analysis via computer simulation includes an evaluation of errors associated with the correlation method. For a uniform flow velocity profile, beamwidth modulation represents the greatest error source. However, for a nonuniform flow velocity profile, the jitter caused by a small flow velocity gradient can exceed the other error sources. A detailed computer simulation evaluated the interdependencies of window length, beam width, vessel diameter, and viewing angle on the estimation of flow velocity.Entities:
Year: 1990 PMID: 18285029 DOI: 10.1109/58.55306
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725