Literature DB >> 18283567

Catastrophic failure of 4 cobalt-alloy Omnifit hip arthroplasty femoral components.

Li-On Lam1, Karl Stoffel, Alan Kop, Eric Swarts.   

Abstract

BACKGROUND: Femoral component neck fracture is an uncommon type of failure in total hip arthroplasty. We present a report on 4 retrieved cobalt-chrome femoral components that fractured at the neck, where we investigated the mechanisms of failure.
METHODS: The 4 retrieved implants were analyzed with regard to their macro- and microstructures and the fracture surfaces were examined using electron microscopy. The medical record of each patient was also examined for any history of complications prior to failure of the implant.
RESULTS: These fractures occurred immediately adjacent to the base of the modular head. Skirted modular heads were used in 3 of the 4 failed components. This constructs promotes corrosion. Cyclic fatigue-loading in combination with the material factors of course grain microstructure and extensive carbide precipitation along the grain boundaries were also identified as the cause of implant failure.
INTERPRETATION: Our findings suggest that a solution annealing step could be introduced into the manufacturing process to improve the microstructure of the cobalt chrome alloy. We also advise caution in using a skirted modular head in combination with a device of known suboptimum microstructure, for a greater margin of safety.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18283567     DOI: 10.1080/17453670710014707

Source DB:  PubMed          Journal:  Acta Orthop        ISSN: 1745-3674            Impact factor:   3.717


  8 in total

Review 1.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

2.  Severe metal-induced osteolysis many years after unipolar hip endoprosthesis.

Authors:  Matthew A Mann; Dylan Tanzer; Michael Tanzer
Journal:  Clin Orthop Relat Res       Date:  2013-01-05       Impact factor: 4.176

3.  Treatment of femoral neck fractures in elderly patients over 60 years of age - which is the ideal modality of primary joint replacement?

Authors:  Christian Ossendorf; Max J Scheyerer; Guido A Wanner; Hans-Peter Simmen; Clément Ml Werner
Journal:  Patient Saf Surg       Date:  2010-10-20

4.  Hip and knee arthroplasty implants contraindicated in obesity.

Authors:  J D Craik; M D Bircher; M Rickman
Journal:  Ann R Coll Surg Engl       Date:  2016-03-29       Impact factor: 1.891

5.  Comparative study of the corrosion behavior of peripheral stents in an accelerated corrosion model: experimental in vitro study of 28 metallic vascular endoprostheses.

Authors:  Karolin J Paprottka; Philipp M Paprottka; Maximilian F Reiser; Tobias Waggershauser
Journal:  Diagn Interv Radiol       Date:  2015 Sep-Oct       Impact factor: 2.630

6.  Catastrophic failure and metallosis of the acetabular component in total hip arthroplasty.

Authors:  Fırat Ozan; Murat Kahraman; Ali Baktır; Kürşat Gençer
Journal:  J Orthop Surg Res       Date:  2021-05-29       Impact factor: 2.359

7.  Fatigue fracture of a cemented Omnifit CoCr femoral stem: implant and failure analysis.

Authors:  Noah Bonnheim; Hannah Gramling; Michael Ries; Sanjai Shukla; Beatrice Iliescu; Lisa Pruitt
Journal:  Arthroplast Today       Date:  2017-08-02

8.  Exactech Opteon Femoral Component Fracture 12 Years after Arthroplasty.

Authors:  Shaun P Patel; Valentin Antoci; John J Kadzielski; Mark S Vrahas
Journal:  Case Rep Orthop       Date:  2016-02-03
  8 in total