Blood pressure changes associated with tilting in normotensive subjects: differences in response pattern as measured by oscillometry and auscultation.

M B Sidery1, I A Macdonald.   

Abstract

Three non-invasive instruments were used to measure blood pressure in the supine position and on tilting--a conventional and a random-zero sphygmomanometer, and an oscillometric device (Accutorr 1A). Twenty normotensives volunteered for the study. There was no statistically significant difference in systolic blood pressure and diastolic blood pressure measured by the conventional and random-zero sphygmomanometers in the supine position. There was a difference between these recordings and those of the Accutorr, with the Accutorr giving higher readings of systolic blood pressure (p less than 0.001), analysis of variance, 95% confidence interval of the difference between the Accutorr and the random-zero was 5.1-15.7 mmHg) and lower readings of diastolic blood pressure (p less than 0.0001, analysis of variance, 95% confidence interval of the difference between the Accutorr and the random-zero was -12.2- -2.2 mmHg). On tilting, the Accutorr showed an increase in systolic blood pressure while the other two machines did not (p less than 0.01, analysis of variance). By contrast, the Accutorr detected a smaller rise in diastolic blood pressure than with the other two instruments (p less than 0.05, analysis of variance). The difference between blood pressure measurements made in the supine position by the two different techniques, auscultation and oscillometry, might be expected. However, the two different techniques do not detect the same blood pressure responses to a change in posture.

Mesh:

Year:  1991        PMID: 1822764     DOI: 10.1007/BF01826214

Source DB:  PubMed          Journal:  Clin Auton Res        ISSN: 0959-9851            Impact factor:   4.435


  21 in total

1.  Comparison of indirect and direct methods of measuring arterial blood pressure.

Authors:  F H VAN BERGEN; D S WEATHERHEAD; A E TRELOAR; A B DOBKIN; J J BUCKLEY
Journal:  Circulation       Date:  1954-10       Impact factor: 29.690

2.  Continuous finger arterial pressure: utility in the cardiovascular laboratory.

Authors:  B P Imholz; W Wieling; G J Langewouters; G A van Montfrans
Journal:  Clin Auton Res       Date:  1991-03       Impact factor: 4.435

3.  Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exercise.

Authors:  W B White; P Lund-Johansen; P Omvik
Journal:  Am J Cardiol       Date:  1990-01-01       Impact factor: 2.778

4.  The diastolic dilemma.

Authors:  D Short
Journal:  Br Med J       Date:  1976-09-18

Review 5.  The end of the p value?

Authors:  S J Evans; P Mills; J Dawson
Journal:  Br Heart J       Date:  1988-09

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  The variability of measurements of casual blood pressure. II. Survey experience.

Authors:  P Armitage; W Fox; G A Rose; C M Tinker
Journal:  Clin Sci       Date:  1966-04       Impact factor: 6.124

8.  Effect of different degrees of tilt on cardiac output, heart rate, and blood pressure in normal man.

Authors:  J Tuckman; J Shillingford
Journal:  Br Heart J       Date:  1966-01

9.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems.

Authors:  E O'Brien; J Petrie; W Littler; M de Swiet; P L Padfield; K O'Malley; M Jamieson; D Altman; M Bland; N Atkins
Journal:  J Hypertens       Date:  1990-07       Impact factor: 4.844

10.  Chronic effect of ketanserin in mild to moderate essential hypertension.

Authors:  A J Woittiez; G J Wenting; A H van den Meiracker; H J Ritsema van Eck; A J Man in't Veld; F A Zantvoort; M A Schalekamp
Journal:  Hypertension       Date:  1986-02       Impact factor: 10.190

View more
  1 in total

1.  Yohimbine in neurally mediated syncope. Pathophysiological implications.

Authors:  R Mosqueda-Garcia; R Fernandez-Violante; J Tank; M Snell; G Cunningham; R Furlan
Journal:  J Clin Invest       Date:  1998-11-15       Impact factor: 14.808

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.