Literature DB >> 18218364

Correction of distortion in endoscope images.

W E Smith1, N Vakil, S A Maislin.   

Abstract

Images formed with endoscopes suffer from a spatial distortion due to the wide-angle nature of the endoscope's objective lens. This change in the size of objects with position precludes quantitative measurement of the area of the objects, which is important in endoscopy for accurately measuring ulcer and lesion sizes over time. A method for correcting the distortion characteristic of endoscope images is presented. A polynomial correction formula was developed for the endoscope lens and validated by comparing quantitative test areas before and after the distortion correction. The distortion correction has been incorporated into a computer program that could readily be applied to electronic images obtained at endoscopy using a desk-top computer. The research presented here is a key step towards the quantitative determination of the area of regions of interest in endoscopy.

Entities:  

Year:  1992        PMID: 18218364     DOI: 10.1109/42.126918

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  6 in total

1.  Technique of distortion correction in endoscopic images using a polynomial expansion.

Authors:  K V Asari; S Kumar; D Radhakrishnan
Journal:  Med Biol Eng Comput       Date:  1999-01       Impact factor: 2.602

2.  Foveated endoscopic lens.

Authors:  Nathan Hagen; Tomasz S Tkaczyk
Journal:  J Biomed Opt       Date:  2012-02       Impact factor: 3.170

3.  Methods for removing glare in digital endoscope images.

Authors:  Eric Abel; Wei Xi; Paul White
Journal:  Surg Endosc       Date:  2011-09-30       Impact factor: 4.584

4.  Clinical study using novel endoscopic system for measuring size of gastrointestinal lesion.

Authors:  Kiyoshi Oka; Takeshi Seki; Tomohiro Akatsu; Takao Wakabayashi; Kazuo Inui; Junji Yoshino
Journal:  World J Gastroenterol       Date:  2014-04-14       Impact factor: 5.742

5.  A clinically translatable hyperspectral endoscopy (HySE) system for imaging the gastrointestinal tract.

Authors:  Jonghee Yoon; James Joseph; Dale J Waterhouse; A Siri Luthman; George S D Gordon; Massimiliano di Pietro; Wladyslaw Januszewicz; Rebecca C Fitzgerald; Sarah E Bohndiek
Journal:  Nat Commun       Date:  2019-04-23       Impact factor: 14.919

6.  A Novel Method for Space Circular Target Detection in Machine Vision.

Authors:  Wenxue Hu; Jiannan Chi; Jiahui Liu; Zuoyun Yang
Journal:  Sensors (Basel)       Date:  2022-01-20       Impact factor: 3.576

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.