Literature DB >> 18213461

Pre-drilling force and insertion torques during orthodontic mini-implant insertion in relation to root contact.

Benedict Wilmes1, Yu-Yu Su, Lilian Sadigh, Dieter Drescher.   

Abstract

BACKGROUND AND AIM: The use of mini-implants for skeletal anchorage has greatly broadened the therapeutic spectrum in orthodontics over the last few years. The alveolar ridge is the most frequent insertion site, which however is associated with tooth injury, a risk not to be underestimated. The objective of this study was to examine the quantitative parameters of pre-drilling and implant insertion in association with the degree of a root contact.
MATERIAL AND METHODS: Eleven lower jaw bones of adult pigs were prepared and embedded in resin. At 320 sites in the toothbearing alveolar ridge a 1.3 mm pre-drilling was carried out up to the complete implant length. The vertical force exerted against the pre-drilling upon penetration of the different bone layers and at a root contact was measured at a drift-speed of 0.5 mm/s. Dual Top screws (1.6 x 8 mm) were then inserted into the prepared implant sites, the insertion torques were measured, and recorded as a function of the rotation angle. After explantation, we prepared histological slides from the level of the implant's maximum diameter. The implant's contact with cortical and cancellous bone and to the roots was measured and correlated to vertical pre-drilling forces and insertion torques.
RESULTS: Vertical pre-drilling forces and insertion torques of orthodontic mini-implants varied in relation to the type of tissue penetrated and the degree of root contact. The insertion torques ranged from 32 to 345 Nmm and pre-drilling forces up to 6 N overall.
CONCLUSION: Root contact can be recognized during pre-drilling by a distinct increase in resistance, and during mini-implant insertion by higher torques.

Entities:  

Mesh:

Year:  2008        PMID: 18213461     DOI: 10.1007/s00056-008-0726-5

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  8 in total

1.  Evaluating miniscrew position using orthopantomograms compared to cone-beam computed tomography.

Authors:  Ruth Bennemann; Martin Baxmann; Ludger Keilig; Susanne Reimann; Bert Braumann; Christoph Bourauel
Journal:  J Orofac Orthop       Date:  2012-05-12       Impact factor: 1.938

2.  Biomechanical properties of orthodontic miniscrews. An in-vitro study.

Authors:  Britta Florvaag; Peter Kneuertz; Frank Lazar; Jürgen Koebke; Joachim E Zöller; Bert Braumann; Robert A Mischkowski
Journal:  J Orofac Orthop       Date:  2010-02-05       Impact factor: 1.938

3.  Influence of geometric design characteristics on primary stability of orthodontic miniscrews.

Authors:  Eman Saad Radwan; Mona A Montasser; Ahmed Maher
Journal:  J Orofac Orthop       Date:  2018-04-10       Impact factor: 1.938

4.  Patient expectations, acceptance and preferences in treatment with orthodontic mini-implants. A randomly controlled study. Part I: insertion techniques.

Authors:  Sandra Lehnen; Fraser McDonald; Christoph Bourauel; Martin Baxmann
Journal:  J Orofac Orthop       Date:  2011-03       Impact factor: 1.938

5.  Insertion torque recordings for the diagnosis of contact between orthodontic mini-implants and dental roots: protocol for a systematic review.

Authors:  Reint Meursinge Reynders; Luisa Ladu; Laura Ronchi; Nicola Di Girolamo; Jan de Lange; Nia Roberts; Annette Plüddemann
Journal:  Syst Rev       Date:  2015-04-02

6.  Mechanical strength of stainless steel and titanium alloy mini-implants with different diameters: an experimental laboratory study.

Authors:  Sérgio Estelita Barros; Viviane Vanz; Kelly Chiqueto; Guilherme Janson; Eduardo Ferreira
Journal:  Prog Orthod       Date:  2021-03-22       Impact factor: 2.750

7.  Effects of insertion angle and implant thread type on the fracture properties of orthodontic mini-implants during insertion.

Authors:  Il-Sik Cho; Tae-Woo Kim; Sug-Joon Ahn; Il-Hyung Yang; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2012-12-10       Impact factor: 2.079

Review 8.  Insertion torque recordings for the diagnosis of contact between orthodontic mini-implants and dental roots: a systematic review.

Authors:  Reint Meursinge Reynders; Luisa Ladu; Laura Ronchi; Nicola Di Girolamo; Jan de Lange; Nia Roberts; Annette Plüddemann
Journal:  Syst Rev       Date:  2016-03-31
  8 in total