Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogs.

Yan Chen1, Hong-In Shin, Hee-Moon Kyung.   

Abstract

INTRODUCTION: The purpose of this study was to compare the influences of different implant modalities on orthodontic microimplants and surrounding tissues biomechanically and histologically.
METHODS: Fifty-six titanium alloy microimplants placed on the buccal side of the maxillae and the mandibles in 2 dogs were divided into 2 groups of 28; one group of microimplants was self-drilling, and the other was self-tapping. Approximately 200 g of continuous and constant forces were applied immediately between 2 microimplants by stretching closed nickel-titanium coil springs for 9 weeks. Peak insertion torque and removal torque were recorded immediately after the implants were placed and when the dogs were killed, respectively. Undecalcified sections of the microimplants and the surrounding tissues were studied with light microscope and fluorescent microscope.
RESULTS: Success rates were higher in the self-drilling group (93%) than in self-tapping group (86%). Higher peak insertion torque and peak removal torque values were seen in the self-drilling group in both the maxilla and the mandible. A tendency to fracture was found in self-drilling group. The percentage of bone-to-implant contact values was greater in the self-drilling group.
CONCLUSIONS: Self-drilling microimplants can provide better anchorage and can be recommended for use in the maxilla and in thin cortical bone areas of the mandible.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18174070     DOI: 10.1016/j.ajodo.2007.01.023

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  11 in total

1.  Miniscrews for orthodontic anchorage: nanoscale chemical surface analyses.

Authors:  Justin Silverstein; Osmar Barreto; Rodrigo França
Journal:  Eur J Orthod       Date:  2015-03-13       Impact factor: 3.075

2.  Patient expectations, acceptance and preferences in treatment with orthodontic mini-implants. A randomly controlled study. Part I: insertion techniques.

Authors:  Sandra Lehnen; Fraser McDonald; Christoph Bourauel; Martin Baxmann
Journal:  J Orofac Orthop       Date:  2011-03       Impact factor: 1.938

3.  Stability changes of miniscrew implants over time.

Authors:  Derid S Ure; Donald R Oliver; Ki Beom Kim; Ana Cláudia Melo; Peter H Buschang
Journal:  Angle Orthod       Date:  2011-05-25       Impact factor: 2.079

4.  A comparitive clinical study between self tapping and drill free screws as a source of rigid orthodontic anchorage.

Authors:  Nishant Gupta; S M Kotrashetti; Vijay Naik
Journal:  J Maxillofac Oral Surg       Date:  2011-05-28

5.  Mechanical properties of self-drilling orthodontic micro-implants with different diameters.

Authors:  Yan Chen; Hee Moon Kyung; Limin Gao; Won-Jae Yu; Eun-Ju Bae; Sang-Min Kim
Journal:  Angle Orthod       Date:  2010-09       Impact factor: 2.079

6.  Surface changes of anodic oxidized orthodontic titanium miniscrew.

Authors:  Sung-Hwan Choi; Jung-Yul Cha; Uk-Hyon Joo; Chung-Ju Hwang
Journal:  Angle Orthod       Date:  2011-10-19       Impact factor: 2.079

7.  Evaluation of primary stability of innovated orthodontic miniscrew system (STS): An ex-vivo study.

Authors:  Massoud Seifi; Negin-Sadat Matini
Journal:  J Clin Exp Dent       Date:  2016-07-01

8.  Bone cutting capacity and osseointegration of surface-treated orthodontic mini-implants.

Authors:  Ho-Young Kim; Sang-Cheol Kim
Journal:  Korean J Orthod       Date:  2016-11-14       Impact factor: 1.372

9.  Primary stability of self-drilling and self-tapping mini-implant in tibia of diabetes-induced rabbits.

Authors:  Jea-Beom Park; Eun-Young Kim; Janghyun Paek; Yoon-Ah Kook; Do-Min Jeong; Il-Sik Cho; Gerald Nelson
Journal:  Int J Dent       Date:  2014-05-11

10.  Enhanced compatibility and initial stability of Ti6Al4V alloy orthodontic miniscrews subjected to anodization, cyclic precalcification, and heat treatment.

Authors:  Eun-Ju Oh; Thuy-Duong T Nguyen; Seung-Youp Lee; Young-Mi Jeon; Tae-Sung Bae; Jong-Gee Kim
Journal:  Korean J Orthod       Date:  2014-09-25       Impact factor: 1.372

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.