Alison Beavis1, Michael Dawson, Philip Doble, Richard A Scolyer, Roger Bourne, Ling-Xi L Li, Rajmohan Murali, Jonathan R Stretch, Cynthia L Lean, Roger F Uren, John F Thompson.
Abstract
BACKGROUND: The sentinel lymph node (SLN) biopsy technique is a reliable means of determining the tumor-harboring status of regional lymph nodes in melanoma patients. When technetium 99 m-labeled antimony trisulfide colloid (99 mTc-Sb2S3) particles are used to perform preoperative lymphoscintigraphy for SLN identification, they are retained in the SLN but are absent or present in only tiny amounts in non-SLNs. The present study investigated the potential for a novel means of assessing the accuracy of surgical identification of SLNs. This involved the use of inductively coupled plasma-mass spectrometry (ICP-MS) to analyze antimony concentrations in fine-needle biopsy (FNB) samples from surgically procured lymph nodes.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18172734 PMCID: PMC2234448 DOI: 10.1245/s10434-007-9693-0
Source DB: PubMed Journal: Ann Surg Oncol ISSN: 1068-9265 Impact factor: 5.344
Operating parameters for Agilent 7500ce Inductively Coupled Plasma–Mass Spectrometry
| Parameter | Condition |
|---|---|
| Radiofrequency power | 1500 W |
| Sample depth | 8 mm |
| Plasma gas flow | 15.0 L min−1 |
| Carrier gas flow | 1.05 L min−1 |
| Peristaltic pump | .15 rps |
| Sample read delay | 60 s |
| Rinsing time | 45 s |
| Dwell time | 100 ms |
| Replicates | 6 |
rps, revolution per second.
Concentration of antimony (ppb) in fine-needle biopsy sample digests collected from sentinel lymph nodes
| Sample No. | Antimony concentration (ppb) |
|---|---|
| 1 | <LOD |
| 2 | <LOD |
| 3 | <LOD |
| 4 | <LOD |
| 5 | .0666 |
| 6 | .0979 |
| 7 | .1738 |
| 8 | .2985 |
| 9 | .3020 |
| 10 | .3085 |
| 11 | .3257 |
| 12 | .3807 |
| 13 | .3812 |
| 14 | .3864 |
| 15 | .3903 |
| 16 | .4045 |
| 17 | .4971 |
| 18 | .6079 |
| 19 | .6407 |
| 20 | .7243 |
| 21 | .8582 |
| 22 | 1.091 |
| 23 | 1.438 |
| 24 | 1.444 |
| 25 | 1.724 |
| 26 | 1.760 |
| 27 | 1.847 |
| 28 | 2.013 |
| 29 | 2.150 |
| 30 | 2.298 |
| 31 | 2.756 |
| 32 | 3.248 |
LOD, limit of detection (.048 ppb).
Concentration of antimony (ppb) in fine-needle biopsy sample digests collected from nonsentinel lymph nodes
| Sample No. | Antimony concentration (ppb) |
|---|---|
| 1 | <LOD |
| 2 | <LOD |
| 3 | <LOD |
| 4 | <LOD |
| 5 | <LOD |
| 6 | <LOD |
| 7 | <LOD |
| 8 | <LOD |
| 9 | <LOD |
| 10 | <LOD |
| 11 | <LOD |
| 12 | <LOD |
| 13 | <LOD |
| 14 | <LOD |
| 15 | .1159 |
LOD, limit of detection (.048 ppb).
Fig. 1.Antimony concentration in digested fine-needle biopsy samples from sentinel lymph nodes and nonsentinel lymph nodes.