Literature DB >> 18024954

Electrical velocimetry for measuring cardiac output in children with congenital heart disease.

K Norozi1, C Beck, W A Osthaus, I Wille, A Wessel, H Bertram.   

Abstract

BACKGROUND: The purpose of this study was to evaluate the agreement of cardiac output measurements obtained by electrical velocimetry (CO(EV)) and those that derived from the direct Fick-oxygen principle (CO(F)) in infants and children with congenital heart defects.
METHODS: Simultaneous measurements of CO(EV) and CO(F) were compared in 32 paediatric patients, aged 11 days to 17.8 yr, undergoing diagnostic right and left heart catheterization. For non-invasive measurements of cardiac output by electrical velocimetry, which is a variation of impedance cardiography, standard surface electrodes were applied to the left side of the neck and the left side of the thorax at the level of the xiphoid process. Cardiac output determined using direct Fick-oxygen principle was calculated by direct measurement of oxygen consumption (VO2) and invasive determination of the arterio-venous oxygen content difference.
RESULTS: An excellent correlation (r=0.97) was found between CO(EV) and CO(F) (P<0.001). The slope of the regression equation [0.96 (SD 0.04)] was not significantly different from the line of identity. The bias between the two methods (CO(EV)-CO(F)) was 0.01 litre min(-1) and the limits of agreement, defined as the bias (2 SD), were -0.47 and +0.45 litre min(-1).
CONCLUSIONS: CO(EV) demonstrates acceptable agreement with data derived from CO(F) in infants and children with congenital heart disease. The new technique is simple, completely non-invasive, and provides beat-to-beat estimation of CO.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18024954     DOI: 10.1093/bja/aem320

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  40 in total

1.  Changes in cardiac output and stroke volume as measured by non-invasive CO monitoring in infants with RSV bronchiolitis.

Authors:  Julie Caplow; Sarah C McBride; Garry M Steil; Jackson Wong
Journal:  J Clin Monit Comput       Date:  2012-04-24       Impact factor: 2.502

2.  Comparison of transpulmonary thermodilution and ultrasound dilution technique: novel insights into volumetric parameters from an animal model.

Authors:  Martin Boehne; Florian Schmidt; Lars Witt; Harald Köditz; Michael Sasse; Robert Sümpelmann; Harald Bertram; Armin Wessel; Wilhelm Alexander Osthaus
Journal:  Pediatr Cardiol       Date:  2012-02-14       Impact factor: 1.655

Review 3.  Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis.

Authors:  Koichi Suehiro; Alexandre Joosten; Linda Suk-Ling Murphy; Olivier Desebbe; Brenton Alexander; Sang-Hyun Kim; Maxime Cannesson
Journal:  J Clin Monit Comput       Date:  2015-08-29       Impact factor: 2.502

4.  Does obesity affect the non-invasive measurement of cardiac output performed by electrical cardiometry in children and adolescents?

Authors:  Luis Altamirano-Diaz; Eva Welisch; Ralf Rauch; Michael Miller; Teresa Sohee Park; Kambiz Norozi
Journal:  J Clin Monit Comput       Date:  2017-02-17       Impact factor: 2.502

5.  A comparative evaluation of electrical velocimetry and inert gas rebreathing for the non-invasive assessment of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Ursula Hoffmann; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  Clin Res Cardiol       Date:  2011-07-01       Impact factor: 5.460

6.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

7.  Propofol effect on cerebral oxygenation in children with congenital heart disease.

Authors:  Thilo Fleck; Stephan Schubert; Peter Ewert; Brigitte Stiller; Nicole Nagdyman; Felix Berger
Journal:  Pediatr Cardiol       Date:  2014-10-14       Impact factor: 1.655

8.  Hemodynamic alterations recorded by electrical cardiometry during ligation of ductus arteriosus in preterm infants.

Authors:  Reyin Lien; Kai-Hsiang Hsu; Jaw-Ji Chu; Yu-Sheng Chang
Journal:  Eur J Pediatr       Date:  2014-10-15       Impact factor: 3.183

9.  Comparison of electrical velocimetry and cardiac magnetic resonance imaging for the non-invasive determination of cardiac output.

Authors:  Frederik Trinkmann; Manuel Berger; Christina Doesch; Theano Papavassiliu; Stefan O Schoenberg; Martin Borggrefe; Jens J Kaden; Joachim Saur
Journal:  J Clin Monit Comput       Date:  2015-06-27       Impact factor: 2.502

10.  Influence of patent ductus arteriosus and ventilators on electrical velocimetry for measuring cardiac output in very-low/low birth weight infants.

Authors:  T Torigoe; S Sato; Y Nagayama; T Sato; H Yamazaki
Journal:  J Perinatol       Date:  2015-01-29       Impact factor: 2.521