Ambulatory oxygenator right ventricular assist device for total right heart and respiratory support.

Dongfang Wang1, Scott D Lick, Xiaoqin Zhou, Xiaojun Liu, Robert J Benkowski, Joseph B Zwischenberger.   

Abstract

BACKGROUND: Our goal is ambulatory total respiratory and right heart assistance allowing a bridge to lung transplant. To that end, we have coupled a compact paracorporeal gas exchange device with a right ventricular assist device (RVAD) to create an "OxyRVAD."
METHODS: Through a limited left thoracotomy, the main pulmonary artery (PA) and right atrium (RA) were exposed in 5 anesthetized sheep. After a systemic heparin bolus, a 12-mm outer diameter crimped graft glued to tubing was anastomosed (end to side) to the main PA and a VAD atrial cannula was placed through the RA appendage. The chest was drained and closed, then the PA graft flowed at 1 to 2 L/min as a shunt to the RA overnight. The next day, the animal was reanticoagulated, and the shunt cannulae clamped and divided. The OxyRVAD unit, consisting of commercially available components including an axial flow pump and low-resistance cardiopulmonary bypass gas exchange device, was interposed. Pumping from RA to PA was maintained at 3 L/min.
RESULTS: Five consecutive sheep survived the implant, and stood and ate normally after initiation of the OxyRVAD. Three survived the full 2-week study, and a fourth was sacrificed on day 13 owing to a storm-related power failure. For these 4 sheep, pump flow was stable at 3 L/min. Carbon dioxide removal was constant and total during the experiment at 200 +/- 19 mL/min. Oxygen transfer was 144 +/- 44 mL/min. One sheep had progressive thrombocytopenia and was sacrificed on day 5 after implant.
CONCLUSIONS: Our ambulatory OxyRVAD can provide total assistance for the right heart and lungs in normal awake sheep for 14 days.

Entities:  

Mesh:

Year:  2007        PMID: 17954089     DOI: 10.1016/j.athoracsur.2007.03.068

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  5 in total

1.  Pediatric Artificial Lung: A Low-Resistance Pumpless Artificial Lung Alleviates an Acute Lamb Model of Increased Right Ventricle Afterload.

Authors:  Fares Alghanem; Benjamin S Bryner; Emilia M Jahangir; Uditha P Fernando; John M Trahanas; Hayley R Hoffman; Robert H Bartlett; Alvaro Rojas-Peña; Ronald B Hirschl
Journal:  ASAIO J       Date:  2017 Mar/Apr       Impact factor: 2.872

2.  Biocompatibility assessment of a long-term wearable artificial pump-lung in sheep.

Authors:  Kang Zhou; Shuqiong Niu; Giacomo Bianchi; Xufeng Wei; Narayana Garimella; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2013-03-03       Impact factor: 3.094

3.  Fabrication and in vivo thrombogenicity testing of nitric oxide generating artificial lungs.

Authors:  Kagya A Amoako; Patrick J Montoya; Terry C Major; Ahmed B Suhaib; Hitesh Handa; David O Brant; Mark E Meyerhoff; Robert H Bartlett; Keith E Cook
Journal:  J Biomed Mater Res A       Date:  2013-04-24       Impact factor: 4.396

4.  Right ventricular unloading and respiratory support with a wearable artificial pump-lung in an ovine model.

Authors:  Yang Liu; Pablo G Sanchez; Xufeng Wei; Tieluo Li; Amelia C Watkins; Shu-ying Li; Bartley P Griffith; Zhongjun J Wu
Journal:  J Heart Lung Transplant       Date:  2014-02-26       Impact factor: 10.247

5.  Feasibility and efficacy of bypassing the right ventricle and pulmonary circulation to treat right ventricular failure: an experimental study.

Authors:  Jan Spillner; Christian Stoppe; Nima Hatam; Andrea Amerini; Ares Menon; Christoph Nix; Ulrich Steinseifer; Yousef Abusabha; Hanna Giessen; Rüdiger Autschbach; Marcus Haushofer
Journal:  J Cardiothorac Surg       Date:  2012-02-06       Impact factor: 1.637

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.