Literature DB >> 17944642

Evaluation of the Airway Scope, a new video laryngoscope, in tracheal intubation by naive operators: a manikin study.

T Miki1, G Inagawa, T Kikuchi, Y Koyama, T Goto.   

Abstract

BACKGROUND AND AIM: The Airway Scope (AWS) is a new video laryngoscope. The purpose of this study was to compare the AWS and Macintosh laryngoscopes with regard to their usefulness for beginners in tracheal intubation.
METHODS: Thirty-one nurses with no previous experience of tracheal intubation used each device 10 times to intubate the trachea of a manikin (Laerdal Airway Managemant Trainer, Laerdal, Stavanger, Norway). Intubation correctly completed within 30 s was regarded as successful. The intubation time, success rate, number of esophageal intubations and dental clicks, and subjective evaluation of difficulty of intubation [rated from 1 (extremely easy) to 5 (extremely difficult)] were recorded.
RESULTS: The average intubation time was significantly shorter for AWS than for the Macintosh laryngoscope (16.7 +/- 11.0 s vs. 23.2 +/- 24.9 s; mean +/- standard deviation; P = 0.0297), and the success rate with AWS was significantly greater (91.3% vs. 79.4%; P < 0.001). In the total of 310 intubation attempts for each device, the number of esophageal intubations (0/310 with AWS vs. 20/310 with the Macintosh laryngoscope) and dental clicks (0/310 with AWS vs. 40/310 with the Macintosh laryngoscope) differed significantly (P < 0.001). AWS was rated as easier to use than the Macintosh laryngoscope (P < 0.001).
CONCLUSION: AWS is more useful for beginners as it provides quicker and easier tracheal intubation.

Entities:  

Mesh:

Year:  2007        PMID: 17944642     DOI: 10.1111/j.1399-6576.2007.01450.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  9 in total

1.  Comparison of intubation performance between the King Vision and Macintosh laryngoscopes in novice personnel: a randomized, crossover manikin study.

Authors:  Yuki Akihisa; Koichi Maruyama; Yukihide Koyama; Rieko Yamada; Akira Ogura; Tomio Andoh
Journal:  J Anesth       Date:  2013-06-30       Impact factor: 2.078

Review 2.  [Indirect laryngoscopy : Alternatives to securing the airway].

Authors:  R R Noppens; C Werner; T Piepho
Journal:  Anaesthesist       Date:  2010-02       Impact factor: 1.041

3.  Influence of clinical experience of the Macintosh laryngoscope on performance with the Pentax-AWS Airway Scope(®), a rigid video-laryngoscope, by paramedics in Japan.

Authors:  Kohei Ota; Takuma Sadamori; Shinji Kusunoki; Tadatsugu Otani; Tomoko Tamura; Kazunobu Une; Yoshiko Kida; Junji Itai; Yasumasa Iwasaki; Nobuyuki Hirohashi; Masakazu Nakao; Koichi Tanigawa
Journal:  J Anesth       Date:  2014-12-20       Impact factor: 2.078

4.  Novel diagnostic device for oral and pharyngeal examinations of children: folding-scope for the oral and pharyngeal cavities.

Authors:  Koichi Tsunoda; Sotaro Sekimoto; Atsunobu Tsunoda
Journal:  BMJ Case Rep       Date:  2010-12-02

5.  Comparison of Parker-tipped and Murphy-tipped tracheal tubes in Airway Scope-assisted intubation in a manikin.

Authors:  Takuro Sanuki; Shingo Sugioka; Motoko Hirokane; Naotaka Kishimoto; Yoshiko Matsuda; Junichiro Kotani
Journal:  J Anesth       Date:  2009-11-18       Impact factor: 2.078

6.  Three-day tracheal intubation manikin training for novice doctors using Macintosh laryngoscope, McGRATH MAC videolaryngoscope and Pentax AirwayScope.

Authors:  Yuryo Murakami; Ryusuke Ueki; Miyuki Niki; Munetaka Hirose; Noriko Shimode
Journal:  Medicine (Baltimore)       Date:  2021-01-29       Impact factor: 1.817

7.  Comparison of the Glidescope and Pentax AWS laryngoscopes to the Macintosh laryngoscope for use by advanced paramedics in easy and simulated difficult intubation.

Authors:  Sajid Nasim; Chrisen H Maharaj; Muhammad A Malik; John O' Donnell; Brendan D Higgins; John G Laffey
Journal:  BMC Emerg Med       Date:  2009-05-17

8.  Usefulness of video-laryngoscopy with the Airway Scope for intubation performance and learning: an experimental manikin controlled study.

Authors:  Pierre-Louis Declercq; Michael Bubenheim; Stéphanie Gelinotte; Kévin Guernon; Jean-Baptiste Michot; Vincent Royon; Dorothée Carpentier; Gaëtan Béduneau; Fabienne Tamion; Christophe Girault
Journal:  Ann Intensive Care       Date:  2016-08-26       Impact factor: 6.925

9.  Impact of a custom-made 3D printed ergonomic grip for direct laryngoscopy on novice intubation performance in a simulated easy and difficult airway scenario-A manikin study.

Authors:  Sung-Hoon Kim; Jeonghoon Kwon; Youn-Jung Kim; Hyung-Joo Lee; Hyo-Chang Seo; Seung Bok Lim; Segyeong Joo; Dong-Woo Seo; Won-Young Kim; Sang-Bum Hong
Journal:  PLoS One       Date:  2018-11-20       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.