Accuracy in tangential breast treatment set-up: a portal imaging study.

G van Tienhoven1, J H Lanson, D Crabeels, S Heukelom, B J Mijnheer.   

Abstract

To test the accuracy and reproducibility of the tangential breast treatment set-up used in The Netherlands Cancer Institute, a portal imaging study was performed in 12 patients treated for early stage breast cancer. With an on-line electronic portal imaging device (EPID) images were obtained of each patient in several fractions and compared with simulator films and with each other. In five patients, multiple images (on the average 7) per fraction were obtained to evaluate set-up variations due to respiratory movement. The central lung distance (CLD) and other set-up parameters varied within one fraction about 1 mm (1 SD). The average variation of these parameters between various fractions was about 2 mm (1 SD). The differences between simulator and treatment set-up over all patients and all fractions was on the average 2-3 mm for the central beam edge to skin distance and the central lung distance. It can be concluded that the tangential breast treatment set-up is very stable and reproducible and that respiration does not have a significant influence on treatment volume. The EPID appears to be an adequate tool for studies of treatment set-up accuracy like this.

Entities:  

Mesh:

Year:  1991        PMID: 1792329     DOI: 10.1016/0167-8140(91)90171-c

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  6 in total

1.  Assessment of setup accuracy in patients receiving postmastectomy radiotherapy using electronic portal imaging.

Authors:  Funda Gul Koseoglu; Nina Tuncel; Adem Unal Kizildag; Melahat Garipagaoglu; Mustafa Adli; Cagatay Andic
Journal:  Radiat Med       Date:  2007-02-27

2.  Correction of systematic set-up error in breast and head and neck irradiation through a no-action level (NAL) protocol.

Authors:  Eva M Lozano; Luis A Pérez; Javier Torres; Carmen Carrascosa; Miguel Sanz; Fermín Mendicote; Antonio Gil
Journal:  Clin Transl Oncol       Date:  2011-01       Impact factor: 3.405

3.  Dosimetric and isocentric variations due to patient setup errors in CT-based treatment planning for breast cancer by electronic portal imaging.

Authors:  Hari Mukundan; Deboleena Mukherjee; Kirti Tyagi; Sachin Taneja; Subhash Ranjan; Samaresh Sahu
Journal:  Med J Armed Forces India       Date:  2018-10-15

4.  Guide to clinical use of electronic portal imaging.

Authors:  M G Herman; J J Kruse; C R Hagness
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

5.  Surface imaging-based analysis of intrafraction motion for breast radiotherapy patients.

Authors:  David B Wiant; Stacy Wentworth; Jacqueline M Maurer; Caroline L Vanderstraeten; Jonathon A Terrell; Benjamin J Sintay
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

6.  Real-time intra-fraction motion management in breast cancer radiotherapy: analysis of 2028 treatment sessions.

Authors:  D Reitz; G Carl; S Schönecker; M Pazos; P Freislederer; M Niyazi; U Ganswindt; F Alongi; M Reiner; C Belka; S Corradini
Journal:  Radiat Oncol       Date:  2018-07-16       Impact factor: 3.481

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.