| Literature DB >> 17921573 |
R T Flynn1, D L Barbee, T R Mackie, R Jeraj.
Abstract
Selective subvolume boosting can theoretically improve tumour control probability while maintaining normal tissue complication probabilities similar to those of uniform dose distributions. In this work the abilities of intensity-modulated x-ray therapy (IMXT) and intensity-modulated proton therapy (IMPT) to deliver boosts to multiple subvolumes of varying size and proximities are compared in a thorough phantom study. IMXT plans were created using the step-and-shoot (IMXT-SAS) and helical tomotherapy (IMXT-HT) methods. IMPT plans were created with the spot scanning (IMPT-SS) and distal gradient tracking (IMPT-DGT) methods. IMPT-DGT is a generalization of the distal edge tracking method designed to reduce the number of proton beam spots required to deliver non-uniform dose distributions relative to IMPT-SS. The IMPT methods were delivered over both 180 degrees and 360 degrees arcs. The IMXT-SAS and IMPT-SS methods optimally satisfied the non-uniform dose prescriptions the least and the most, respectively. The IMPT delivery methods reduced the normal tissue integral dose by a factor of about 2 relative to the IMXT delivery methods, regardless of the delivery arc. The IMPT-DGT method reduced the number of proton beam spots by a factor of about 3 relative to the IMPT-SS method.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17921573 PMCID: PMC2712448 DOI: 10.1088/0031-9155/52/20/001
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609