Thirty-day in-parallel artificial lung testing in sheep.

Hitoshi Sato1, Candice M Hall, Nathan G Lafayette, Joshua R Pohlmann, Niru Padiyar, John M Toomasian, Jonathan W Haft, Keith E Cook.   

Abstract

BACKGROUND: Thirty-day testing of the MC3 Biolung (MC3 Inc, Ann Arbor, MI) total artificial lung (TAL) was performed to prepare for future clinical testing.
METHODS: TAL inlet and outlet grafts were sewn to the pulmonary artery and left atrium of 8 sheep (35.6 +/- 1.6 kg), and the TAL was attached the next day. Hemodynamic and sheep blood gas data were measured every 1 to 4 hours. TAL blood gases were measured twice daily, and organ function was assessed three times per week. The TAL was replaced if its resistance increased 300% or if the oxygen content difference across the TAL decreased 25% versus baseline. After 30 days, the sheep were euthanized and necropsied.
RESULTS: Five sheep survived 30 days. Three sheep were euthanized before 30 days due to bleeding, mechanical graft failure, or gastric distress. Survivors had normal, stable hemodynamics and blood gases. Average device use was 9.5 +/- 2.1 days. TAL oxygen transfer was 108 +/- 9.2 mL/min with 51% +/- 6.3% of cardiac output flowing to the TAL. TAL resistance and flow were 1.3 +/- 0.3 mm Hg x min/L and 2.4 +/- 0.2 L/min at baseline versus 2.6 +/- 0.9 mm Hg x min/L and 2.0 +/- 0.2 L/min for the remaining 30 days. Platelet and white blood cell counts increased 88% and 84% from baseline, respectively, after 10 days and were stable thereafter. Ischemic lesions in the kidney were seen in most sheep at necropsy, but organ function was normal.
CONCLUSIONS: Thirty-day respiratory support was feasible with the Biolung, but improvements in biocompatibility and anticoagulation regimen are warranted to reduce the thrombogenicity of the device.

Entities:  

Mesh:

Year:  2007        PMID: 17888959     DOI: 10.1016/j.athoracsur.2007.05.051

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  19 in total

1.  Fourteen Day In Vivo Testing of a Compliant Thoracic Artificial Lung.

Authors:  David J Skoog; Joshua R Pohlmann; David S Demos; Christopher N Scipione; Amit Iyengar; Rebecca E Schewe; Ahmed B Suhaib; Kelly L Koch; Keith E Cook
Journal:  ASAIO J       Date:  2017 Sep/Oct       Impact factor: 2.872

2.  Early in vivo experience with the pediatric continuous-flow total artificial heart.

Authors:  Jamshid H Karimov; David J Horvath; Nicole Byram; Gengo Sunagawa; Barry D Kuban; Shengqiang Gao; Raymond Dessoffy; Kiyotaka Fukamachi
Journal:  J Heart Lung Transplant       Date:  2018-03-30       Impact factor: 10.247

3.  Long-term animal model of venovenous extracorporeal membrane oxygenation with atrial septal defect as a bridge to lung transplantation.

Authors:  Daniele Camboni; Alvaro Rojas; Peter Sassalos; David Spurlock; Kelly L Koch; Sarah Menchak; Jennifer Singleton; Erika Boothman; Jonathan W Haft; Robert H Bartlett; Keith E Cook
Journal:  ASAIO J       Date:  2013 Nov-Dec       Impact factor: 2.872

Review 4.  Artificial lung basics: fundamental challenges, alternative designs and future innovations.

Authors:  Heather Nolan; Dongfang Wang; Joseph B Zwischenberger
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

5.  Long-term support with an ambulatory percutaneous paracorporeal artificial lung.

Authors:  Xiaoqin Zhou; Dongfang Wang; Ryan Sumpter; Gary Pattison; Cherry Ballard-Croft; Joseph B Zwischenberger
Journal:  J Heart Lung Transplant       Date:  2012-03-22       Impact factor: 10.247

6.  Large Animal Model of Pumpless Arteriovenous Extracorporeal CO₂ Removal Using Room Air via Subclavian Vessels.

Authors:  Lucas J Witer; Ryan A Howard; John M Trahanas; Benjamin S Bryner; Fares Alghanem; Hayley R Hoffman; Marie S Cornell; Robert H Bartlett; Alvaro Rojas-Peña
Journal:  ASAIO J       Date:  2016 Jan-Feb       Impact factor: 2.872

7.  Thirty-day in-vivo performance of a wearable artificial pump-lung for ambulatory respiratory support.

Authors:  Zhongjun J Wu; Tao Zhang; Giacomo Bianchi; Xufeng Wei; Ho-Sung Son; Kang Zhou; Pablo G Sanchez; Jose Garcia; Bartley P Griffith
Journal:  Ann Thorac Surg       Date:  2011-11-25       Impact factor: 4.330

8.  Biocompatibility assessment of a long-term wearable artificial pump-lung in sheep.

Authors:  Kang Zhou; Shuqiong Niu; Giacomo Bianchi; Xufeng Wei; Narayana Garimella; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2013-03-03       Impact factor: 3.094

9.  A biohybrid artificial lung prototype with active mixing of endothelialized microporous hollow fibers.

Authors:  Alexa A Polk; Timothy M Maul; Daniel T McKeel; Trevor A Snyder; Craig A Lehocky; Bruce Pitt; Donna Beer Stolz; William J Federspiel; William R Wagner
Journal:  Biotechnol Bioeng       Date:  2010-06-15       Impact factor: 4.530

10.  Use of a low-resistance compliant thoracic artificial lung in the pulmonary artery to pulmonary artery configuration.

Authors:  Christopher N Scipione; Rebecca E Schewe; Kelly L Koch; Andrew W Shaffer; Amit Iyengar; Keith E Cook
Journal:  J Thorac Cardiovasc Surg       Date:  2013-02-10       Impact factor: 5.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.