OBJECTIVE: The purpose of this study was to determine the frequency of discordance at MRI-guided vacuum-assisted biopsy and to assess the cancer rate in discordant lesions.
MATERIALS AND METHODS: With institutional review board approval, retrospective review was performed of a database of 342 lesions that had MRI-guided vacuum-assisted biopsy during a 39-month period. Biopsy samples were obtained in a 1.5-T magnet using a 9-gauge MRI-compatible vacuum-assisted biopsy device. Medical and pathology records were reviewed to determine the number of discordant lesions and surgical outcome. Statistical analysis was performed.
RESULTS: Among 342 lesions that had MRI-guided vacuum-assisted biopsy, results were discordant in 24 (7% [95% CI, 3-14%]) lesions. The discordance rate was significantly (p < 0.001) higher among MRI target lesions that were possibly missed rather than sampled. A trend was seen (p < 0.06) toward a higher discordance rate in MRI lesions that were sampled rather than excised at MRI-guided vacuum-assisted biopsy. Subsequent surgery in 20 discordant lesions yielded cancer in six (30% [12-54%]), including ductal carcinoma in situ (DCIS) in two and invasive carcinoma in four (three ductal and one lobular, all with DCIS). The cancer rate among discordant lesions was significantly higher in postmenopausal women than in premenopausal women (p <0.05).
CONCLUSION: Imaging-histologic discordance was found in 7% of lesions that had MRI-guided vacuum-assisted biopsy. Among discordant lesions, surgical excision revealed cancer in 30%. Imaging-histologic correlation is essential after MRI-guided vacuum-assisted biopsy to avoid delay in the diagnosis of breast cancer.

Entities:  

Mesh:

Year:  2007        PMID: 17885056     DOI: 10.2214/AJR.07.2531

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  10 in total

1.  Vacuum-assisted biopsies under MR guidance: results of 72 procedures.

Authors:  C Malhaire; C El Khoury; F Thibault; A Athanasiou; P Petrow; L Ollivier; A Tardivon
Journal:  Eur Radiol       Date:  2010-01-30       Impact factor: 5.315

2.  Outcomes of classic lobular neoplasia diagnosed on breast core needle biopsy: a retrospective multi-center study.

Authors:  Iskender Sinan Genco; Bugra Tugertimur; Qing Chang; Lauren Cassell; Sabina Hajiyeva
Journal:  Virchows Arch       Date:  2019-11-27       Impact factor: 4.064

Review 3.  Pathologic High-risk Lesions, Diagnosis and Management.

Authors:  Melissa Murray
Journal:  Clin Obstet Gynecol       Date:  2016-12       Impact factor: 2.190

4.  Results of Short-Term Follow-Up in BI-RADS 3 and 4a Breast Lesions with a Histological Diagnosis of Fibroadenoma at Percutaneous Needle Biopsy.

Authors:  Paola Clauser; Massimo Bazzocchi; Magda Marcon; Viviana Londero; Chiara Zuiani
Journal:  Breast Care (Basel)       Date:  2017-08-29       Impact factor: 2.860

5.  A novel MR-guided interventional device for 3D circumferential access to breast tissue.

Authors:  Matthew Smith; Xu Zhai; Ray Harter; Gale Sisney; Mai Elezabye; Sean Fain
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

6.  Short interval follow-up after a benign concordant MR-guided vacuum assisted breast biopsy--is it worthwhile?

Authors:  Sara D Shaylor; Samantha L Heller; Amy N Melsaether; Dipti Gupta; Avani Gupta; James Babb; Linda Moy
Journal:  Eur Radiol       Date:  2014-03-14       Impact factor: 5.315

Review 7.  Magnetic Resonance Imaging-Guided Breast Interventions: Role in Biopsy Targeting and Lumpectomies.

Authors:  Eva C Gombos; Jayender Jagadeesan; Danielle M Richman; Daniel F Kacher
Journal:  Magn Reson Imaging Clin N Am       Date:  2015-07-08       Impact factor: 2.266

8.  Diagnostic imaging strategy for MDCT- or MRI-detected breast lesions: use of targeted sonography.

Authors:  Satoko Nakano; Masahiko Ohtsuka; Akemi Mibu; Masato Karikomi; Hitomi Sakata; Masahiro Yamamoto
Journal:  BMC Med Imaging       Date:  2012-06-12       Impact factor: 1.930

9.  MR-guided vacuum-assisted breast biopsy of MRI-only lesions: a single center experience.

Authors:  Claudio Spick; Melanie Schernthaner; Katja Pinker; Panagiotis Kapetas; Maria Bernathova; Stephan H Polanec; Hubert Bickel; Georg J Wengert; Margaretha Rudas; Thomas H Helbich; Pascal A Baltzer
Journal:  Eur Radiol       Date:  2016-03-16       Impact factor: 5.315

10.  Challenges in MRI-Guided Breast Biopsy and Some Suggested Strategies: Case Based Review.

Authors:  Chotai Niketa; Kathleen Ann Pang; Joon Wei Lim
Journal:  Diagnostics (Basel)       Date:  2022-08-16
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.