Literature DB >> 17879783

Setup verification and in vivo dosimetry during intraoperative radiation therapy (IORT) for prostate cancer.

Antonella Soriani1, Valeria Landoni, Simona Marzi, Giuseppe Iaccarino, Biancamaria Saracino, Giorgio Arcangeli, Marcello Benassi.   

Abstract

The purpose of this study was to check the setup and dose delivered to the patients during intraoperative electron beam radiation therapy (IORT) for prostate cancer. Twenty eight patients underwent IORT after radical prostatectomy for prostate cancer by means of a dedicated mobile accelerator, Novac7 (by Hitesys, SpA, Italy). A 9 MeV electron beam at high dose per pulse was used. Eighteen patients received IORT at escalating doses of 16, 18, and 20 Gy at 85% isodose, six patients for each dose level. Further, ten patients received 20 Gy at 85% isodose. The electron applicator position was checked in all cases by means of two orthogonal images obtained with brilliance intensifier. Target and organ at risk doses were measured in vivo by a MOSFETs dosimetry system. MOSFETs and microMOSFET dosimeters were inserted into sterile catheters and directly positioned into the rectal lumen, for ten patients, and into the bladder to urethra anastomosis, in the last 14 cases. Verification at 0 degree led to very few adjustments of setup while verifications at 90 degrees often suggested to bring the applicator closer to the target. In vivo dosimetry showed an absorbed dose into the rectum wall < or =1% of the total dose. The average dose value inside the anastomosis, for the 12 patients analyzed, was 23.7 Gy with a standard deviation of +/-7.6%, when the prescription was 20 Gy at 85% isodose. Using a C-arm mobile image intensifier, it is possible to assess if the positioning is correct and safe. Radio-opaque clips and liquid were necessary to obtain good visible images. In vivo MOSFETs dosimetry is feasible and reliable. A satisfactory agreement between measured and expected doses was found.

Entities:  

Mesh:

Year:  2007        PMID: 17879783     DOI: 10.1118/1.2750965

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Practical issues regarding angular and energy response in in vivo intraoperative electron radiotherapy dosimetry.

Authors:  Juan López-Tarjuelo; Ana Bouché-Babiloni; Virginia Morillo-Macías; Agustín Santos-Serra; Carlos Ferrer-Albiach
Journal:  Rep Pract Oncol Radiother       Date:  2016-10-19

2.  In vivo dosimetry in intraoperative electron radiotherapy: microMOSFETs, radiochromic films and a general-purpose linac.

Authors:  Juan López-Tarjuelo; Ana Bouché-Babiloni; Virginia Morillo-Macías; Noelia de Marco-Blancas; Agustín Santos-Serra; Juan David Quirós-Higueras; Carlos Ferrer-Albiach
Journal:  Strahlenther Onkol       Date:  2014-06-26       Impact factor: 3.621

3.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

4.  In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT).

Authors:  Filipa Costa; Sandra Sarmento; Dora Gomes; Helena Magalhães; Rosário Arrais; Graciete Moreira; Maria Fátima Cruz; José Pedro Silva; Lúcio Santos; Olga Sousa
Journal:  Br J Radiol       Date:  2016-05-18       Impact factor: 3.039

5.  Intraoperative radiation therapy as adjuvant treatment in locally advanced stage tumours involving the middle ear: a hypothesis-generating retrospective study.

Authors:  G Cristalli; G Mercante; L Marucci; A Soriani; S Telera; G Spriano
Journal:  Acta Otorhinolaryngol Ital       Date:  2016-04-29       Impact factor: 2.124

6.  Radical prostatectomy and intraoperative radiation therapy in high-risk prostate cancer.

Authors:  Giansilvio Marchioro; Alessandro Volpe; Roberto Tarabuzzi; Giuseppina Apicella; Marco Krengli; Carlo Terrone
Journal:  Adv Urol       Date:  2012-01-30

7.  Implementation of an intraoperative electron radiotherapy in vivo dosimetry program.

Authors:  Juan López-Tarjuelo; Virginia Morillo-Macías; Ana Bouché-Babiloni; Enrique Boldó-Roda; Rafael Lozoya-Albacar; Carlos Ferrer-Albiach
Journal:  Radiat Oncol       Date:  2016-03-15       Impact factor: 3.481

Review 8.  Present state and issues in IORT Physics.

Authors:  Frank W Hensley
Journal:  Radiat Oncol       Date:  2017-01-27       Impact factor: 3.481

Review 9.  In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Igor Olaciregui-Ruiz; Sam Beddar; Peter Greer; Nuria Jornet; Boyd McCurdy; Gabriel Paiva-Fonseca; Ben Mijnheer; Frank Verhaegen
Journal:  Phys Imaging Radiat Oncol       Date:  2020-08-29

10.  Dose evaluation for skin and organ in hepatocellular carcinoma during angiographic procedure.

Authors:  Daniela D'Alessio; Claudia Giliberti; Antonella Soriani; Livio Carpanese; Giuseppe Pizzi; Giulio Vallati; Lidia Strigari
Journal:  J Exp Clin Cancer Res       Date:  2013-10-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.