Literature DB >> 17821985

Technical note: a physical phantom for assessment of accuracy of deformable alignment algorithms.

Rojano Kashani1, Martina Hub, Marc L Kessler, James M Balter.   

Abstract

The purpose of this study was to investigate the feasibility of a simple deformable phantom as a QA tool for testing and validation of deformable image registration algorithms. A diagnostic thoracic imaging phantom with a deformable foam insert was used in this study. Small plastic markers were distributed through the foam to create a lattice with a measurable deformation as the ground truth data for all comparisons. The foam was compressed in the superior-inferior direction using a one-dimensional drive stage pushing a flat "diaphragm" to create deformations similar to those from inhale and exhale states. Images were acquired at different compressions of the foam and the location of every marker was manually identified on each image volume to establish a known deformation field with a known accuracy. The markers were removed digitally from corresponding images prior to registration. Different image registration algorithms were tested using this method. Repeat measurement of marker positions showed an accuracy of better than 1 mm in identification of the reference marks. Testing the method on several image registration algorithms showed that the system is capable of evaluating errors quantitatively. This phantom is able to quantitatively assess the accuracy of deformable image registration, using a measure of accuracy that is independent of the signals that drive the deformation parameters.

Entities:  

Mesh:

Year:  2007        PMID: 17821985     DOI: 10.1118/1.2739812

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Verification of MLC based real-time tumor tracking using an electronic portal imaging device.

Authors:  Sarah Han-Oh; Byong Yong Yi; Fritz Lerma; Barry L Berman; Minzhi Gui; Cedric Yu
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Analysis of deformable image registration accuracy using computational modeling.

Authors:  Hualiang Zhong; Jinkoo Kim; Indrin J Chetty
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

Review 3.  Accurate accumulation of dose for improved understanding of radiation effects in normal tissue.

Authors:  David A Jaffray; Patricia E Lindsay; Kristy K Brock; Joseph O Deasy; W A Tomé
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-01       Impact factor: 7.038

4.  Objective assessment of deformable image registration in radiotherapy: a multi-institution study.

Authors:  Rojano Kashani; Martina Hub; James M Balter; Marc L Kessler; Lei Dong; Lifei Zhang; Lei Xing; Yaoqin Xie; David Hawkes; Julia A Schnabel; Jamie McClelland; Sarang Joshi; Quan Chen; Weiguo Lu
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

5.  The potential of helical tomotherapy in the treatment of head and neck cancer.

Authors:  Dirk Van Gestel; Dirk Verellen; Lien Van De Voorde; Bie de Ost; Geert De Kerf; Olivier Vanderveken; Carl Van Laer; Danielle Van den Weyngaert; Jan B Vermorken; Vincent Gregoire
Journal:  Oncologist       Date:  2013-05-30

6.  Estimation of three-dimensional intrinsic dosimetric uncertainties resulting from using deformable image registration for dose mapping.

Authors:  Francisco J Salguero; Nahla K Saleh-Sayah; Chenyu Yan; Jeffrey V Siebers
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

7.  Multiple anatomy optimization of accumulated dose.

Authors:  W Tyler Watkins; Joseph A Moore; James Gordon; Geoffrey D Hugo; Jeffrey V Siebers
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

8.  Results of a multi-institutional benchmark test for cranial CT/MR image registration.

Authors:  Kenneth Ulin; Marcia M Urie; Joel M Cherlow
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-08       Impact factor: 7.038

9.  The utilization of consistency metrics for error analysis in deformable image registration.

Authors:  Edward T Bender; Wolfgang A Tomé
Journal:  Phys Med Biol       Date:  2009-08-28       Impact factor: 3.609

10.  Validation of deformable registration in head and neck cancer using analysis of variance.

Authors:  A Mencarelli; S van Beek; S van Kranen; C Rasch; M van Herk; J-J Sonke
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

View more