Seven-day artificial lung testing in an in-parallel configuration.

Hitoshi Sato1, Grant W Griffith, Candice M Hall, John M Toomasian, Ronald B Hirschl, Robert H Bartlett, Keith E Cook.   

Abstract

BACKGROUND: A thoracic artificial lung, the MC3 Biolung, is being developed as a bridge to lung transplantation or as a treatment for acute respiratory insufficiency.
METHODS: The thoracic artificial lung was tested in 10 sheep with the goal of 7 days of respiratory support. The sheep were recovered from surgery and monitored awake for 7 days. Hemodynamics, blood gases, blood cell counts, and organ function were recorded, and after 7 days, all sheep were euthanized and necropsied.
RESULTS: Seven sheep survived the full duration. Cardiac output and mean arterial blood pressure were unchanged, averaging 4.7 +/- 0.8 L/min and 98 +/- 10 mm Hg, respectively. Arterial oxygen tension and device oxygen transfer rate were also unchanged, averaging 110 +/- 26 mm Hg and 97.7 +/- 35 mL/min, respectively. Arterial carbon dioxide tension was within normal ranges during the entire experiment, averaging 37.4 +/- 3.8 mm Hg. Artificial lung blood flow decreased from 51% +/- 14% of cardiac output on day 1 to 30% +/- 16% by day 7 because of changes in natural and artificial lung resistance. White blood cell counts were significantly elevated on days 5 and 7, and lastly, kidney and liver function remained normal, although signs of kidney infarction or hemorrhage were noted.
CONCLUSIONS: The thoracic artificial lung is suitable for 7-day attachment, but improvements in blood biocompatibility are warranted.

Entities:  

Mesh:

Year:  2007        PMID: 17720415     DOI: 10.1016/j.athoracsur.2007.03.016

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  7 in total

1.  Fourteen Day In Vivo Testing of a Compliant Thoracic Artificial Lung.

Authors:  David J Skoog; Joshua R Pohlmann; David S Demos; Christopher N Scipione; Amit Iyengar; Rebecca E Schewe; Ahmed B Suhaib; Kelly L Koch; Keith E Cook
Journal:  ASAIO J       Date:  2017 Sep/Oct       Impact factor: 2.872

2.  Pulsatile flow past an oscillating cylinder.

Authors:  Adnan Qamar; Robinson Seda; Joseph L Bull
Journal:  Phys Fluids (1994)       Date:  2011-04-21       Impact factor: 3.521

3.  72-Hour in vivo evaluation of nitric oxide generating artificial lung gas exchange fibers in sheep.

Authors:  Angela Lai; Caitlin T Demarest; Chi Chi Do-Nguyen; Rei Ukita; David J Skoog; Neil M Carleton; Kagya A Amoako; Patrick J Montoya; Keith E Cook
Journal:  Acta Biomater       Date:  2019-04-03       Impact factor: 8.947

4.  Evaluation of a pumpless lung assist device in hypoxia-induced pulmonary hypertension in juvenile piglets.

Authors:  George T El-Ferzli; Joseph B Philips; Arlene Bulger; Namasivayam Ambalavanan
Journal:  Pediatr Res       Date:  2009-12       Impact factor: 3.756

5.  In Vivo 5 Day Animal Studies of a Compact, Wearable Pumping Artificial Lung.

Authors:  Shalv P Madhani; Brian J Frankowski; Sang-Ho Ye; Greg W Burgreen; William R Wagner; Robert Kormos; Jonathan D'Cunha; William J Federspiel
Journal:  ASAIO J       Date:  2019-01       Impact factor: 2.872

6.  In-parallel attachment of a low-resistance compliant thoracic artificial lung under rest and simulated exercise.

Authors:  Rebecca E Schewe; Christopher N Scipione; Kelly L Koch; Keith E Cook
Journal:  Ann Thorac Surg       Date:  2012-09-07       Impact factor: 4.330

7.  A Model of Pediatric End-Stage Lung Failure in Small Lambs <20 kg.

Authors:  Benjamin D Carr; Clinton J Poling; Pavel Hala; Matias Caceres Quinones; Aaron R Prater; Jennifer S McLeod; Robert H Bartlett; Alvaro Rojas-Pena; Ronald B Hirschl
Journal:  ASAIO J       Date:  2020-05       Impact factor: 3.826

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.