In-parallel attachment of a low-resistance compliant thoracic artificial lung under rest and simulated exercise.

Rebecca E Schewe1, Christopher N Scipione, Kelly L Koch, Keith E Cook.   

Abstract

BACKGROUND: Previous thoracic artificial lungs (TALs) had blood flow impedance greater than that of the natural lungs, which could cause abnormal pulmonary hemodynamics. New compliant TALs (cTALs), however, have an impedance lower than that of the natural lung.
METHODS: In this study, a cTAL of new design was attached between the pulmonary artery (PA) and the left atrium (LA) in 5 sheep (60.2 ± 1.9 kg). A distal PA band was placed to control the percentage of cardiac output (CO) routed to the cTAL. Rest and exercise conditions were simulated using a continuous dobutamine infusion of 0 and 5 μg/kg/min, respectively. At each dose, a hemodynamic data set was acquired at baseline (no flow to the cTAL), and 60%, 75%, and 90% of CO was shunted to the cTAL.
RESULTS: Device resistance did not vary with blood flow rate, averaging 0.51 ± 0.03 mm Hg/(L/min). Under all conditions, CO was not significantly different from baseline. Pulmonary system impedance increased above baseline only with 5 μg/kg/min of dobutamine and 90% of CO diverted to the cTAL.
CONCLUSIONS: Results indicated minimal changes in pulmonary hemodynamics during PA-LA cTAL attachment for high device flows under rest and exercise conditions.
Copyright © 2012 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22959566      PMCID: PMC3606559          DOI: 10.1016/j.athoracsur.2012.07.034

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  14 in total

1.  Hemodynamic effects of attachment modes and device design of a thoracic artificial lung.

Authors:  F Boschetti; C E Perlman; K E Cook; L F Mockros
Journal:  ASAIO J       Date:  2000 Jan-Feb       Impact factor: 2.872

2.  Hemodynamic and gas transfer properties of a compliant thoracic artificial lung.

Authors:  Keith E Cook; Carrie E Perlman; Ralf Seipelt; Carl L Backer; Constantine Mavroudis; Lyle F Mockrost
Journal:  ASAIO J       Date:  2005 Jul-Aug       Impact factor: 2.872

3.  Effect of artificial lung compliance on in vivo pulmonary system hemodynamics.

Authors:  Hitoshi Sato; John W McGillicuddy; Grant W Griffith; Amy M Cosnowski; Sean D Chambers; Ronald B Hirschl; Robert H Bartlett; Keith E Cook
Journal:  ASAIO J       Date:  2006 May-Jun       Impact factor: 2.872

4.  Hemodynamic consequences of thoracic artificial lung attachment configuration: a computational model.

Authors:  Carrie E Perlman; Lyle F Mockros
Journal:  ASAIO J       Date:  2007 Jan-Feb       Impact factor: 2.872

5.  The relationship between pulmonary system impedance and right ventricular function in normal sheep.

Authors:  Alexander S Kuo; Hitoshi Sato; Junewai L Reoma; Keith E Cook
Journal:  Cardiovasc Eng       Date:  2009-12

6.  In-parallel artificial lung attachment at high flows in normal and pulmonary hypertension models.

Authors:  Begum Akay; Junewai L Reoma; Daniele Camboni; Joshua R Pohlmann; John M Albert; Ayushi Kawatra; Ayanna D Gouch; Robert H Bartlett; Keith E Cook
Journal:  Ann Thorac Surg       Date:  2010-07       Impact factor: 4.330

7.  Thoracic surgical procedures supported by a pumpless interventional lung assist.

Authors:  Karsten Wiebe; Jochen Poeling; Matthias Arlt; Alois Philipp; Daniele Camboni; Stefan Hofmann; Christof Schmid
Journal:  Ann Thorac Surg       Date:  2010-06       Impact factor: 4.330

8.  Hemodynamic and humoral conditions in stepwise reduction of pulmonary blood flow during venoarterial bypass in awake goats.

Authors:  Y Takewa; E Tatsumi; Y Taenaka; K Eya; T Nakatani; T Nishimura; Y S Sohn; T Masuzawa; Y Wakisaka; M Nakamura; S Endo; H Takano; S Kitamura
Journal:  ASAIO J       Date:  1997 Sep-Oct       Impact factor: 2.872

9.  Extracorporeal pumpless interventional lung assist in clinical practice: determinants of efficacy.

Authors:  T Müller; M Lubnow; A Philipp; T Bein; A Jeron; A Luchner; L Rupprecht; M Reng; J Langgartner; C E Wrede; M Zimmermann; D Birnbaum; C Schmid; G A J Riegger; M Pfeifer
Journal:  Eur Respir J       Date:  2008-11-14       Impact factor: 16.671

10.  Hemodynamic design requirements for in-series thoracic artificial lung attachment in a model of pulmonary hypertension.

Authors:  Begum Akay; Julie A Foucher; Daniele Camboni; Kelly L Koch; Ayushi Kawatra; Keith E Cook
Journal:  ASAIO J       Date:  2012 Jul-Aug       Impact factor: 2.872

View more
  2 in total

1.  Fourteen Day In Vivo Testing of a Compliant Thoracic Artificial Lung.

Authors:  David J Skoog; Joshua R Pohlmann; David S Demos; Christopher N Scipione; Amit Iyengar; Rebecca E Schewe; Ahmed B Suhaib; Kelly L Koch; Keith E Cook
Journal:  ASAIO J       Date:  2017 Sep/Oct       Impact factor: 2.872

2.  A Model of Pediatric End-Stage Lung Failure in Small Lambs <20 kg.

Authors:  Benjamin D Carr; Clinton J Poling; Pavel Hala; Matias Caceres Quinones; Aaron R Prater; Jennifer S McLeod; Robert H Bartlett; Alvaro Rojas-Pena; Ronald B Hirschl
Journal:  ASAIO J       Date:  2020-05       Impact factor: 3.826

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.