Fatigue fracture of a proximally modular, distally tapered fluted implant with diaphyseal fixation.

Martín A Buttaro1, Michael B Mayor, Douglas Van Citters, Francisco Piccaluga.   

Abstract

We report and analyze the causes of a fracture in a proximally modular, distally tapered fluted MP stem in a 48-year-old woman (168 cm, 67 kg) with severe proximal bone deficiency. Evidence of fatigue failure with striations initiated laterally was observed in the laser etching of the tensile aspect of the prosthesis. However, metallurgical analysis suggested that laser engraving did not alter the microstructure of the stem. Stress due to the absence of proximal femoral bone support may have been sufficiently high to put this particular stem at risk for fatigue fracture. This important complication should be addressed when choosing this therapeutic option in cases with substantial proximal femoral bone loss. Strut allograft support should be recommended in such cases.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17689793     DOI: 10.1016/j.arth.2006.07.007

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  12 in total

1.  Distal non-traumatic fracture of the cementless MP stem: A case report and review of previous reported cases.

Authors:  Hussein Abdelaziz; Christian Ansorge; Thorsten Gehrke; Mustafa Citak
Journal:  J Clin Orthop Trauma       Date:  2019-07-30

2.  Proximal femoral reconstructions with bone impaction grafting and metal mesh.

Authors:  Martín A Buttaro; Fernando Comba; Francisco Piccaluga
Journal:  Clin Orthop Relat Res       Date:  2009-03-18       Impact factor: 4.176

3.  Fatigue fracture of tibial arthroplasty implant masked by contralateral knee arthritis.

Authors:  Barry J O'Neill; May Cleary; John P McElwain
Journal:  Int J Surg Case Rep       Date:  2013-03-14

4.  Influence of prosthesis design and implantation technique on implant stresses after cementless revision THR.

Authors:  Markus O Heller; Manav Mehta; William R Taylor; Dong-Yeong Kim; Andrew Speirs; Georg N Duda; Carsten Perka
Journal:  J Orthop Surg Res       Date:  2011-05-13       Impact factor: 2.359

5.  Novel concept of a modular hip implant could contribute to less implant failure in THA: a hypothesis.

Authors:  Ronny Grunert; Stefan Schleifenbaum; Robert Möbius; Michael Kopper; Christian Rotsch; Welf-Guntram Drossel; Niels Hammer; Torsten Prietzel
Journal:  Patient Saf Surg       Date:  2018-01-08

6.  Fatigue fracture of a cemented Omnifit CoCr femoral stem: implant and failure analysis.

Authors:  Noah Bonnheim; Hannah Gramling; Michael Ries; Sanjai Shukla; Beatrice Iliescu; Lisa Pruitt
Journal:  Arthroplast Today       Date:  2017-08-02

7.  Fracture of the femoral component after a lightning strike injury: A case report.

Authors:  Xavier Lizano-Díez; Eduard Alentorn-Geli; Alfonso León-García; Fernando Marqués-López
Journal:  Acta Orthop Traumatol Turc       Date:  2016-12-28       Impact factor: 1.511

8.  Early outcomes after revision total hip arthroplasty with a modern modular femoral revision stem in 65 consecutive cases.

Authors:  Christopher E Pelt; Marissa L Stagg; Christin Van Dine; Mike B Anderson; Christopher L Peters; Jeremy M Gililland
Journal:  Arthroplast Today       Date:  2018-11-17

9.  Anterolateral Bone Window for Revision Broken Cemented Stem of Unipolar Hemiarthroplasty.

Authors:  Mohamed Mosa Mohamed Mahmoud; Bahaaeldin Ibrahim; Amr Abdelhalem Amr; Maysara Abdelhalem Bayoumy
Journal:  Adv Orthop       Date:  2021-02-23

10.  A simple new technique for the removal of fractured femoral stems: a case report.

Authors:  Goran Bicanic; Kresimir Crnogaca; Domagoj Delimar
Journal:  J Med Case Rep       Date:  2014-05-15
View more