Virginie Callot1, Guillaume Duhamel, Patrick J Cozzone.
Abstract
OBJECT: To evaluate the feasibility of mouse spinal cord MR imaging using echo-planar imaging (EPI).Entities:
Mesh:
Year: 2007 PMID: 17661098 PMCID: PMC4271416 DOI: 10.1007/s10334-007-0079-x
Source DB: PubMed Journal: MAGMA ISSN: 0968-5243 Impact factor: 2.310
Fig. 1From left to right: 1-, 2-, 4- and 8-shot spin-echo EPI (SE-EPI), conventional spin-echo (c-SE) and gradient-echo (GE) images, acquired at the level of C1/C2, C3 and C7. In-plane resolution is identical for each sequence (133 × 133 µm2). Scan times for the acquisition of the seven slice series (C1–C7) are, respectively, equal to: 0.42, 0.83, 1.66, 3.33, 10.67 and 4.48 min
Fig. 24-Shot SE-EPI images (TE 14 ms, NEX 10, slice thickness 0.75 mm, total acquisition time 3.33 min). Resolutions (voxel volumes) are: a 133 × 133 µm2 (13.3 nl) and b 74 × 94 µm2 (5.2 nl)
Fig. 3From left to right: a T2-weighted images for different TE and the resultant T2 map; b T1-weighted images for different TI and the resultant T1 map