Optically stimulated luminescence (OSL) of carbon-doped aluminum oxide (Al2O3:C) for film dosimetry in radiotherapy.

V Schembri1, B J M Heijmen.   

Abstract

INTRODUCTION AND
PURPOSE: Conventional x-ray films and radiochromic films have inherent challenges for high precision radiotherapy dosimetry. Here we have investigated basic characteristics of optically stimulated luminescence (OSL) of irradiated films containing carbon-doped aluminum oxide (Al2O3:C) for dosimetry in therapeutic photon and electron beams.
MATERIALS AND METHODS: The OSL films consist of a polystyrene sheet, with a top layer of a mixture of single crystals of Al2O3:C, ground into a powder, and a polyester base. The total thickness of the films is 0.3 mm. Measurements have been performed in a water equivalent phantom, using 4, 6, 10, and 18 MV photon beams, and 6-22 MeV electron beams. The studies include assessment of the film response (acquired OSL signal/delivered dose) on delivered dose (linearity), dose rate (1-6 Gy/min), beam quality, field size and depth (6 MV, ranges 4 x 4-30 x 30 cm2, dmax-35 cm). Doses have been derived from ionization chamber measurements. OSL films have also been compared with conventional x-ray and GafChromic films for dosimetry outside the high dose area, with a high proportion of low dose scattered photons. In total, 787 OSL films have been irradiated.
RESULTS: Overall, the OSL response for electron beams was 3.6% lower than for photon beams. Differences between the various electron beam energies were not significant. The 6 and 18 MV photon beams differed in response by 4%. No response dependencies on dose rate were observed. For the 6 MV beam, the field size and depth dependencies of the OSL response were within +/-2.5%. The observed inter-film response variation for films irradiated with the same dose varied from 1% to 3.2% (1 SD), depending on the measurement day. At a depth of 20 cm, 5 cm outside the 20 x 20 cm2 6 and 18 MV beams, an over response of 17% was observed. In contrast to GafChromic and conventional x-ray films, the response of the Al2O3:C films is linear in the clinically relevant dose range 0-200 cGy.
CONCLUSIONS: Measurement of the OSL signal of irradiated films containing Al2O3:C is a promising technique for film dosimetry in radiotherapy with no or small response variations with dose rate, beam quality, field size and depth, and a linear response from 0 to 200 cGy.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17654914     DOI: 10.1118/1.2737160

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Theoretical and empirical investigations of KCl : Eu2+ for nearly water-equivalent radiotherapy dosimetry.

Authors:  Yuanshui Zheng; Zhaohui Han; Joseph P Driewer; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

2.  Angular dependence of the nanoDot OSL dosimeter.

Authors:  James R Kerns; Stephen F Kry; Narayan Sahoo; David S Followill; Geoffrey S Ibbott
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

3.  The effective dose of different scanning protocols using the Sirona GALILEOS(®) comfort CBCT scanner.

Authors:  D Chambers; R Bohay; L Kaci; R Barnett; J Battista
Journal:  Dentomaxillofac Radiol       Date:  2014-10-31       Impact factor: 2.419

4.  Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.

Authors:  H Harold Li; Zhiyan Xiao; Rachael Hansel; Nels Knutson; Deshan Yang
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

5.  On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy.

Authors:  Fasihah Hanum Yusof; Ngie Min Ung; Jeannie Hsiu Ding Wong; Wei Loong Jong; Vannyat Ath; Vincent Chee Ee Phua; Siew Ping Heng; Kwan Hoong Ng
Journal:  PLoS One       Date:  2015-06-08       Impact factor: 3.240

6.  Response of Nanodot Optically Stimulated Luminescence Dosimeters to Therapeutic Electron Beams.

Authors:  Y Retna Ponmalar; Ravikumar Manickam; S Sathiyan; K M Ganesh; R Arun; Henry Finlay Godson
Journal:  J Med Phys       Date:  2017 Jan-Mar

7.  Evaluation of NanoDot Optically Stimulated Luminescence Dosimeter for Cone-shaped Small-field Dosimetry of Cyberknife Stereotactic Radiosurgery Unit: A Monte Carlo Simulation and Dosimetric Verification Study.

Authors:  Fadil Akyol; Neslihan Sarigul; Mete Yeginer; Yagiz Yedekci; Haluk Utku
Journal:  J Med Phys       Date:  2019 Jan-Mar

8.  Recent developments of optically stimulated luminescence materials and techniques for radiation dosimetry and clinical applications.

Authors:  A S Pradhan; J I Lee; J L Kim
Journal:  J Med Phys       Date:  2008-07

9.  Optically stimulated luminescent dosimetry for high dose rate brachytherapy.

Authors:  Christopher Jason Tien; Robert Ebeling; Jessica R Hiatt; Bruce Curran; Edward Sternick
Journal:  Front Oncol       Date:  2012-08-06       Impact factor: 6.244

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.