Literature DB >> 17639435

Treatment of periprosthetic femur fractures with the non-contact bridging plate: a new angular stable implant.

J B Erhardt1, K Grob, G Roderer, A Hoffmann, T N Forster, M S Kuster.   

Abstract

OBJECTIVES: We report the application of a new fixed angle plate (NCB DF, Zimmer inc. USA, Warsaw, IN) in the treatment of periprosthetic femur fractures. The NCB DF combines conventional plating technique with polyaxial screw placement and angular stability.
DESIGN: Prospective cohort study.
SETTING: A single level-1 trauma center. PATIENTS: From May 2003 to December 2005, a total of 24 patients with periprosthetic femur fractures were treated. The NCB DF femur plate was used in all cases. The average follow-up period was 12 months (3-31 months). Twelve patients had a periprosthetic fracture after total knee replacement (TKA) and 12 patients after total hip replacement (THA). The mean period from primary joint replacement to periprosthetic fracture was 8.2 years for the THA group and 7.2 years for the TKA group. INTERVENTION: A combined conventional/locking surgical technique was performed in all the cases. MAIN OUTCOME MEASURES: Union, non-union, mal-union, duration of surgery, range of motion, postoperative mobility, subjective patient satisfaction and complications.
RESULTS: The union rate was 90%, the mal-union rate 5% and the re-operation rate 15%. Postoperative mobility reached the preoperative level in all but for two patients. Three complications occurred relating to the implant or the procedure: one fatigue failure of the plate (non-union), one screw breakage, and one wound infection.
CONCLUSIONS: The NCB DF combines conventional plating technique with polyaxial screw placement and angular stability. This combination technique shows promising results regarding union and mal-union rates in periprosthetic fractures in elderly and osteoporotic patients.

Entities:  

Mesh:

Year:  2007        PMID: 17639435     DOI: 10.1007/s00402-007-0396-6

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  28 in total

1.  Polyaxial Locked Implants in the Treatment of Type Vancouver B1 Periprosthetic Fractures of the Femur: Retrospective Clinical Examination in 58 Cases with Review of the Literature.

Authors:  Jan Meiners; Maximilian Faschingbauer; Christine Voigt; Christian Jürgens; Arndt P Schulz
Journal:  Eur J Trauma Emerg Surg       Date:  2009-07-10       Impact factor: 3.693

2.  Plate failure following plate osteosynthesis in periprosthetic femoral fractures.

Authors:  Sandra Boesmueller; Sebastian F Baumbach; Marcus Hofbauer; Gerald E Wozasek
Journal:  Wien Klin Wochenschr       Date:  2015-06-18       Impact factor: 1.704

3.  Minimally invasive periprosthetic plate osteosynthesis using the locking attachment plate.

Authors:  C Kammerlander; S L Kates; M Wagner; T Roth; M Blauth
Journal:  Oper Orthop Traumatol       Date:  2013-08       Impact factor: 1.154

4.  Biomechanical comparison of two angular stable plate constructions for periprosthetic femur fracture fixation.

Authors:  Dirk Wähnert; Richard Schröder; Martin Schulze; Peter Westerhoff; Michael Raschke; Richard Stange
Journal:  Int Orthop       Date:  2013-10-11       Impact factor: 3.075

5.  Periprosthetic fracture around a stable femoral stem treated with locking plate osteosynthesis: distal femoral locking plate alone versus with cerclage cable.

Authors:  Young-Soo Shin; Seung-Beom Han
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-01-16

6.  Outcome of osteosynthesis for periprosthetic fractures after total knee arthroplasty: a retrospective study.

Authors:  Hasmukh Nagwadia; Prateek Joshi
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-01-03

7.  The value of a standardized and reproducible surgical technique in treatment of Vancouver B2 periprosthetic fractures: our experience.

Authors:  Stefano Biggi; Andrea Camera; Riccardo Tedino; Andrea Capuzzo; Stefano Tornago
Journal:  Eur J Trauma Emerg Surg       Date:  2018-06-19       Impact factor: 3.693

8.  Extracortical plate fixation with new plate inserts and cerclage wires for the treatment of periprosthetic hip fractures.

Authors:  Johannes D Bastian; Andre Butscher; Gianni Bigolin; Matthias A Zumstein; Hubert P Nötzli
Journal:  Int Orthop       Date:  2013-10-05       Impact factor: 3.075

9.  Radiographic Risk Factors For Interprosthetic Femur Fractures.

Authors:  Jason S Lipof; Ari D Amitai; Kyle T Judd; John T Gorczyca
Journal:  Iowa Orthop J       Date:  2017

10.  Retrograde intramedullary nailing for periprosthetic supracondylar fractures of the femur after total knee arthroplasty.

Authors:  Hyuk-Soo Han; Kyu-Won Oh; Seung-Baik Kang
Journal:  Clin Orthop Surg       Date:  2009-11-25