Paul F Lachiewicz1, Elizabeth S Soileau, Paul Bryant.
Abstract
It is unclear whether resurfacing or fully porous-coated or proximally coated stems are the best femoral components for young patients. First-generation proximally coated titanium fiber-metal femoral components had a high rate of failure. We prospectively followed 56 patients (70 hips) whom we believed were at high risk for loosening of a cemented femoral component; all patients were implanted with one second-generation proximally coated titanium femoral component. Ten patients (11 hips) died and three patients (four hips) were lost to minimum followup. Three models of titanium fiber-metal acetabular component were implanted with screws. Patients were clinically evaluated using the Harris hip score. We used standardized radiographs to assess fixation status and osteolysis. Forty-four patients (55 hips) with a mean age of 45 years were followed for a minimum of 7 years (mean, 10.5 years; range, 7-15 years). No femoral component had been revised for any reason. Bone ingrowth (spot welding) was seen in 41 hips (75%) and proximal femoral osteolysis in only four hips. Reoperations for polyethylene wear were performed in seven patients (12.7%) and two patients had liner exchange for recurrent dislocation. Even with an articulating surface with considerable polyethylene wear debris, these second-generation proximally coated titanium fiber-metal femoral components had a survival rate of 100% at a mean 10.5-year followup.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17621229 DOI: 10.1097/BLO.0b013e318137a167
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176