Horizontal-plane localization of noise and speech signals by postlingually deafened adults fitted with bilateral cochlear implants.

D Wesley Grantham1, Daniel H Ashmead, Todd A Ricketts, Robert F Labadie, David S Haynes.   

Abstract

OBJECTIVES: The main purpose of the study was to assess the ability of adults with bilateral cochlear implants to localize noise and speech signals in the horizontal plane. A second objective was to measure the change in localization performance in these adults between approximately 5 and 15 mo after activation. A third objective was to evaluate the relative roles of interaural level difference (ILD) and interaural temporal difference (ITD) cues in localization by these subjects.
DESIGN: Twenty-two adults, all postlingually deafened and all bilaterally fitted with MED-EL COMBI 40+ cochlear implants, were tested in a modified source identification task. Subjects were tested individually in an anechoic chamber, which contained an array of 43 numbered loudspeakers extending from -90 degrees to +90 degrees azimuth. On each trial, a 200-msec signal (either a noise burst or a speech sample) was presented from one of 17 active loudspeakers (span: +/-80 degrees ), and the subject had to identify which source from the 43 loudspeakers in the array produced the signal. Subjects were tested in three conditions: left device only active, right device only active, and both devices active. Twelve of the 22 subjects were retested approximately 10 mo after their first test. In Experiment 2, the spectral content and rise-decay time of the noise stimulus were manipulated.
RESULTS: The relationship between source azimuth and response azimuth was characterized in terms of the adjusted constant error (ĉ). (1) With both devices active, ĉ for the noise stimulus varied from 8.1 degrees to 43.4 degrees (mean: 24.1 degrees ). By comparison, ĉ for a group of listeners with normal hearing ranged from 3.5 degrees to 7.8 degrees (mean: 5.6 degrees ). When subjects listened in unilateral mode (with one device turned off), ĉ was at or near chance (50.5 degrees ) in all cases. However, when considering unilateral performance on each subject's better side, average ĉ for the speech stimulus was 47.9 degrees , which was significantly (but only slightly) better than chance. (2) When listening bilaterally, error score was significantly lower for the speech stimulus (mean ĉ = 21.5 degrees ) than for the noise stimulus (mean ĉ = 24.1 degrees ). (3) As a group, the 12 subjects who were retested 10 mo after their first visit showed no significant improvement in localization performance during the intervening time. However, two subjects who performed very poorly during their first visit showed dramatic improvement (error scores were halved) over the intervening time. In Experiment 2, removing the high-frequency content of noise signals resulted in significantly poorer performance, but removing the low-frequency content or increasing the rise-decay time did not have an effect.
CONCLUSIONS: In agreement with previously reported data, subjects with bilateral cochlear implants localized sounds in the horizontal plane remarkably well when using both of their devices, but they generally could not localize sounds when either device was deactivated. They could localize the speech signal with slightly, but significantly better accuracy than the noise, possibly due to spectral differences in the signals, to the availability of envelope ITD cues with the speech but not the noise signal, or to more central factors related to the social salience of speech signals. For most subjects the remarkable ability to localize sounds has stabilized by 5 mo after activation. However, for some subjects who perform poorly initially, there can be substantial improvement past 5 mo. Results from Experiment 2 suggest that ILD cues underlie localization ability for noise signals, and that ITD cues do not contribute.

Entities:  

Mesh:

Year:  2007        PMID: 17609614     DOI: 10.1097/AUD.0b013e31806dc21a

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  86 in total

1.  Current research with cochlear implants at Arizona State University.

Authors:  Michael F Dorman; Anthony Spahr; Rene H Gifford; Sarah Cook; Ting Zhang; Louise Loiselle; William Yost; Lara Cardy; JoAnne Whittingham; David Schramm
Journal:  J Am Acad Audiol       Date:  2012-06       Impact factor: 1.664

2.  Current and planned cochlear implant research at New York University Laboratory for Translational Auditory Research.

Authors:  Mario A Svirsky; Matthew B Fitzgerald; Arlene Neuman; Elad Sagi; Chin-Tuan Tan; Darlene Ketten; Brett Martin
Journal:  J Am Acad Audiol       Date:  2012-06       Impact factor: 1.664

3.  Multi-microphone adaptive noise reduction strategies for coordinated stimulation in bilateral cochlear implant devices.

Authors:  Kostas Kokkinakis; Philipos C Loizou
Journal:  J Acoust Soc Am       Date:  2010-05       Impact factor: 1.840

4.  Channel Interaction and Current Level Affect Across-Electrode Integration of Interaural Time Differences in Bilateral Cochlear-Implant Listeners.

Authors:  Katharina Egger; Piotr Majdak; Bernhard Laback
Journal:  J Assoc Res Otolaryngol       Date:  2015-09-16

5.  Effect of multi-electrode configuration on sensitivity to interaural timing differences in bilateral cochlear-implant users.

Authors:  Alan Kan; Heath G Jones; Ruth Y Litovsky
Journal:  J Acoust Soc Am       Date:  2015-12       Impact factor: 1.840

6.  A longitudinal study in adults with sequential bilateral cochlear implants: time course for individual ear and bilateral performance.

Authors:  Ruth M Reeder; Jill B Firszt; Laura K Holden; Michael J Strube
Journal:  J Speech Lang Hear Res       Date:  2014-06-01       Impact factor: 2.297

7.  Sound source localization by hearing preservation patients with and without symmetrical low-frequency acoustic hearing.

Authors:  Louise H Loiselle; Michael F Dorman; William A Yost; René H Gifford
Journal:  Audiol Neurootol       Date:  2015-04-01       Impact factor: 1.854

8.  Restoration of spatial hearing in adult cochlear implant users with single-sided deafness.

Authors:  Ruth Y Litovsky; Keng Moua; Shelly Godar; Alan Kan; Sara M Misurelli; Daniel J Lee
Journal:  Hear Res       Date:  2018-04-14       Impact factor: 3.208

9.  Lateralization of Interaural Level Differences with Multiple Electrode Stimulation in Bilateral Cochlear-Implant Listeners.

Authors:  Olga A Stakhovskaya; Matthew J Goupell
Journal:  Ear Hear       Date:  2017 Jan/Feb       Impact factor: 3.570

10.  Benefits of localization and speech perception with multiple noise sources in listeners with a short-electrode cochlear implant.

Authors:  Camille C Dunn; Ann Perreau; Bruce Gantz; Richard S Tyler
Journal:  J Am Acad Audiol       Date:  2010-01       Impact factor: 1.664

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.