| Literature DB >> 17589735 |
Silvia Nápole Fichino1, Doris Ruthy Lewis, Mariana Lopes Fávero.
Abstract
UNLABELLED: The differential diagnosis of hearing loss with air and bone Auditory Brainstem Response in small children has not been enough studied in Brazil. AIM: To compare air and bone Auditory Brainstem Response results in children under 2 months of age with normal hearing. STUDYEntities:
Mesh:
Year: 2007 PMID: 17589735 PMCID: PMC9450619 DOI: 10.1016/s1808-8694(15)31074-0
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
Parameters used in BAEP recordings by AC and BC; adapted from Hood.4
| Parameter | Air conduction | Bone conduction |
|---|---|---|
| Stimulus | 100msec click | 100msec click |
| Polarity | Alternated | Alternated |
| Intensities | 80,60,40,30 dB | 40,30 dB |
| Stimulus frequency | 27.7/Sec | 27.7/Sec |
| Window | 25msec | 25msec |
| Filters | 100-3000 Hz | 100-3000Hz |
| Number of stimulus | At least 2000 | At least 2000 |
| Reproductions | 2 records | 2 records |
Joint frequencies and percentage distribution of AC and BC response occurrence, right ear, intensity of 40 dBHL.
| VA LE 30 - V | Total | |||
|---|---|---|---|---|
| VO - 30 - V | No | Yes | ||
| No | Frequency | 1 | 1 | 2 |
| % | 8% | 8% | 17% | |
| Yes | Frequency | 1 | 9 | 10 |
| % | 8% | 75% | 83% | |
| Total | Frequency | 2 | 10 | 12 |
| % | 17% | 83% | 100% |
Joint frequencies and percentage distribution of AC and BC response occurrence, right ear, intensity of 30 dBHL.
| VA RE 30 - V | Total | |||
|---|---|---|---|---|
| VO - 30 - V | No | Yes | ||
| No | Frequency | 2 | 2 | |
| % | 17% | 17% | ||
| Yes | Frequency | 3 | 7 | 10 |
| % | 25% | 58% | 83% | |
| Total | Frequency | 3 | 9 | 12 |
| % | 25% | 75% | 100% |
Joint frequencies and percentage distribution of AC and BC response occurrence, left ear, intensity of 30 dBHL.
| VA LE 30 - V | Total | |||
|---|---|---|---|---|
| VO - 30 - V | No | Yes | ||
| No | Frequency | 1 | 1 | 2 |
| % | 8% | 8% | 17% | |
| Yes | Frequency | 1 | 9 | 10 |
| % | 8% | 75% | 83% | |
| Total | Frequency | 2 | 10 | 12 |
| % | 17% | 83% | 100% |
Response estimate likelihood at the intensity of 40 dBHL
| Situation | N | p | Confidence interval |
|---|---|---|---|
| AC RE 40 V | 12 | 0,92 | (0,62; 1,0) |
| AC LE 40 V | 12 | 1 | (0,78; 1,0) |
| BC VO 40 V | 12 | 1 | (0,78; 1,0) |
Response estimate likelihood at the intensity of 30 dBHL
| Situation | N | p | Confidence interval |
|---|---|---|---|
| AC RE 30 V | 12 | 0,75 | (0,43; 0,95) |
| AC LE 30 V | 12 | 0,83 | (0,52; 0,98) |
| BC 30 V | 12 | 0,83 | (0,52; 0,98) |
Figure 1Graph, showing the mean latencies (ms) for wave V at each intensity level (dBHL) for Air and Bone conduction – -o- AC -<- BC.