PURPOSE: To compare two wavefront-sensing devices based on different principles.
METHODS: Thirty-eight healthy eyes of 19 patients were measured five times in the reproducibility study. Twenty eyes of 10 patients were measured in the comparison study. The Tracey Visual Function Analyzer (VFA), based on the ray-tracing principle and the Nidek optical pathway difference (OPD)-Scan, based on the dynamic skiascopy principle were compared. Standard deviation (SD) of root mean square (RMS) errors was compared to verify the reproducibility. We evaluated RMS errors, Zernike terms and conventional refractive indexes (Sph, Cyl, Ax, and spherical equivalent).
RESULTS: In RMS errors reading, both devices showed similar ratios of SD to the mean measurement value (VFA: 57.5+/-11.7%, OPD-Scan: 53.9+/-10.9%). Comparison on the same eye showed that almost all terms were significantly greater using the VFA than using the OPD-Scan. However, certain high spatial frequency aberrations (tetrafoil, pentafoil, and hexafoil) were consistently measured near zero with the OPD-Scan.
CONCLUSION: Both devices showed similar level of reproducibility; however, there was considerable difference in the wavefront reading between machines when measuring the same eye. Differences in the number of sample points, centration, and measurement algorithms between the two instruments may explain our results.

Entities:  

Mesh:

Year:  2007        PMID: 17571088      PMCID: PMC2666015          DOI: 10.1038/sj.eye.6702901

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  24 in total

1.  Wave-front measurement errors from restricted concentric subdomains.

Authors:  K A Goldberg; K Geary
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2001-09       Impact factor: 2.129

2.  Comparison of a ray-tracing refractometer, autorefractor, and computerized videokeratography in measuring pseudophakic eyes.

Authors:  Li Wang; Manjula Misra; Ioannis G Pallikaris; Douglas D Koch
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

3.  Repeatability of ocular wavefront measurement.

Authors:  Nigel Davies; Luis Diaz-Santana; David Lara-Saucedo
Journal:  Optom Vis Sci       Date:  2003-02       Impact factor: 1.973

4.  Evaluation of refractive error measurements of the Wavescan Wavefront system and the Tracey Wavefront aberrometer.

Authors:  Li Wang; Nan Wang; Douglas D Koch
Journal:  J Cataract Refract Surg       Date:  2003-05       Impact factor: 3.351

5.  Matrix method to find a new set of Zernike coefficients from an original set when the aperture radius is changed.

Authors:  Charles E Campbell
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2003-02       Impact factor: 2.129

6.  Scaling Zernike expansion coefficients to different pupil sizes.

Authors:  Jim Schwiegerling
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2002-10       Impact factor: 2.129

7.  Higher order wavefront aberrations of cornea and magnitude of refractive correction in laser in situ keratomileusis.

Authors:  Tetsuro Oshika; Kazunori Miyata; Tadatoshi Tokunaga; Tomokazu Samejima; Shiro Amano; Sumiyoshi Tanaka; Yoko Hirohara; Toshifumi Mihashi; Naoyuki Maeda; Takashi Fujikado
Journal:  Ophthalmology       Date:  2002-06       Impact factor: 12.079

8.  Clinical experience with the Tscherning aberrometer.

Authors:  M Kaemmerer; M Mrochen; P Mierdel; H E Krinke; T Seiler
Journal:  J Refract Surg       Date:  2000 Sep-Oct       Impact factor: 3.573

9.  Clinical applications of the OPD-Scan wavefront aberrometer/corneal topographer.

Authors:  Phil Buscemi
Journal:  J Refract Surg       Date:  2002 May-Jun       Impact factor: 3.573

10.  Measurement repeatability of corneal aberrations.

Authors:  Marine Gobbe; Michel Guillon; Cecile Maissa
Journal:  J Refract Surg       Date:  2002 Sep-Oct       Impact factor: 3.573

View more
  4 in total

1.  Internal spherical aberration by ray tracing-type aberrometry in multifocal pseudophakic eyes.

Authors:  I Jun; Y J Choi; E K Kim; K Y Seo; T-I Kim
Journal:  Eye (Lond)       Date:  2012-06-29       Impact factor: 3.775

2.  Long-term outcomes of limbal relaxing incisions during cataract surgery: aberrometric analysis.

Authors:  Gaspare Monaco; Antonio Scialdone
Journal:  Clin Ophthalmol       Date:  2015-08-31

3.  Wavefront aberrations and retinal image quality in different lenticular opacity types and densities.

Authors:  Cheng-Zhe Wu; Hua Jin; Zhen-Nv Shen; Ying-Jun Li; Xun Cui
Journal:  Sci Rep       Date:  2017-11-10       Impact factor: 4.379

4.  The efficiency of aspheric intraocular lens according to biometric measurements.

Authors:  Woong-Joo Whang; Junjie Piao; Young-Sik Yoo; Choun-Ki Joo; Geunyoung Yoon
Journal:  PLoS One       Date:  2017-10-16       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.