Literature DB >> 12361159

Measurement repeatability of corneal aberrations.

Marine Gobbe1, Michel Guillon, Cecile Maissa.   

Abstract

PURPOSE: Customized procedures that aim to improve visual performance beyond traditional sole correction of sphere and cylinder require the reliable determination of higher order corneal and ocular aberrations for individual eyes. The aim of the current investigation was to determine the repeatability of corneal aberration measurement.
METHODS: Ten subjects were measured on 10 different days using a videokeratoscope. The Zernike coefficients and the wavefront error were calculated from the height data files using CTView. Repeatability was calculated for apertures of 3, 4.5, and 6 mm at the 95% confidence level.
RESULTS: The wavefront error absolute repeatability was similar for the 4th and 10th order analysis. The absolute repeatability for individual Zernike coefficients was coefficient dependent; the poorest repeatability was recorded for coma-like aberrations (uncertainty 3 mm = 0.060 microm; 4.5 mm = 0.083 microm; 6 mm = 0.109 microm). The relative repeatability compared to the population means was good for wavefront error (5% to 20%) but poor for some individual Zernike coefficients.
CONCLUSION: Videokeratoscope clinical repeatability was sufficient to reliably determine wavefront error at a single session, but may require several examinations and/or repeated independent measurements at the same examination to determine reliably some Zernike coefficients. This has significant practical implications for accurate customized correction of aberrations.

Entities:  

Mesh:

Year:  2002        PMID: 12361159     DOI: 10.3928/1081-597X-20020901-14

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  5 in total

1.  Comparison of laser ray-tracing and skiascopic ocular wavefront-sensing devices.

Authors:  D-U G Bartsch; K Bessho; L Gomez; W R Freeman
Journal:  Eye (Lond)       Date:  2007-06-15       Impact factor: 3.775

2.  Contact lenses and special back surface design after penetrating keratoplasty to improve contact lens fit and visual outcome.

Authors:  C Gruenauer-Kloevekorn; U Kloevekorn-Fischer; G I W Duncker
Journal:  Br J Ophthalmol       Date:  2005-12       Impact factor: 4.638

3.  Videokeratoscopic indices in relation to epidemiological exposure to keratoconus.

Authors:  Jose Luis Mato; Isabel Lema; Elío Díez-Feijoo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-07       Impact factor: 3.117

4.  Measurement accuracy and uncertainty in plant biomechanics.

Authors:  Nathanael Nelson; Christopher J Stubbs; Ryan Larson; Douglas D Cook
Journal:  J Exp Bot       Date:  2019-07-23       Impact factor: 6.992

5.  Intra- and Intersession Repeatability of an Optical Quality and Intraocular Scattering Measurement System in Children.

Authors:  Mi Tian; Huamao Miao; Yang Shen; Jian Gao; Xiaofen Mo; Xingtao Zhou
Journal:  PLoS One       Date:  2015-11-04       Impact factor: 3.240

  5 in total