Cardiac output derived from left ventricular pressure during conductance catheter evaluations: an extended Modelflow method.

Sergio Valsecchi1, Giovanni B Perego, Jan J Schreuder, Federica Censi, Jos R C Jansen.   

Abstract

OBJECTIVE: The Modelflow method computes cardiac output (CO) from arterial pressure (CO-MFao) by simulating a non-linear three-element Windkessel model of aortic input impedance. We present a novel technique to apply the Modelflow method to the left ventricular pressure (Plv) signal, to obtain an estimation of CO (CO-MFlv).
METHODS: We extended the model by simulating the aortic valve as a resistance placed in series to the characteristic impedance. In our model the valve resistance is infinite in diastole, while it has a patient related value during systole. Twenty one patients with heart failure were studied with a combined pressure-conductance catheter positioned in the left ventricle and a micromanometer in the aorta. CO changes were induced during temporary sequential atrio-biventricular pacing with different atrio-ventricular and inter-ventricular intervals. After a single calibration we compared CO-MFlv with CO-MFao (516 measurement series) and with CO estimated by the conductance catheter volume technique (CO-cond) (267 series).
RESULTS: CO ranged from 2.42 to 7.59 l/min with an overall mean of 4.36 (1.38) l/min. Comparing the two Modelflow methods, the bias was -0.04 (0.36) l/min, with limits of agreement of -0.77 and 0.70 l/min, and a coefficient of variation of 8.4%. Of the 516 changes in CO from baseline values, 112 were greater than 0.5 l/min and 110 of these (98%) were scored in the same direction by both methods. For comparisons between CO-MFlv and CO-cond the bias was -0.10 (0.49), with limits of agreement of -1.12 and 0.90 l/min, and a coefficient of variation of 12.5%.
CONCLUSIONS: Cardiac output estimates by the modelflow method from aortic pressure and left ventricular pressure are interchangeable in patients without mitral and aortic abnormalities. After an initial calibration, CO-MFlv presents near zero bias and an adequate precision.

Entities:  

Mesh:

Year:  2007        PMID: 17541715     DOI: 10.1007/s10877-007-9078-8

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   1.977


  21 in total

1.  A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.

Authors:  L A Critchley; J A Critchley
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

2.  Simultaneous conductance catheter and dimension assessment of left ventricle volume in the intact animal.

Authors:  R J Applegate; C P Cheng; W C Little
Journal:  Circulation       Date:  1990-02       Impact factor: 29.690

3.  Left intraventricular pressure-impedance diagrams (DPZ) to assess cardiac function. Part I: morphology and potential sources of artifacts.

Authors:  O E Clavin; J C Spinelli; H Alonso; P Solarz; M E Valentinuzzi; R H Pichel
Journal:  Med Prog Technol       Date:  1986

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model.

Authors:  G J Langewouters; K H Wesseling; W J Goedhard
Journal:  J Biomech       Date:  1984       Impact factor: 2.712

6.  Automatic intraaortic balloon pump timing using an intrabeat dicrotic notch prediction algorithm.

Authors:  Jan J Schreuder; Alessandro Castiglioni; Andrea Donelli; Francesco Maisano; Jos R C Jansen; Ramzi Hanania; Pat Hanlon; Jan Bovelander; Ottavio Alfieri
Journal:  Ann Thorac Surg       Date:  2005-03       Impact factor: 4.330

7.  A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients.

Authors:  J R Jansen; J J Schreuder; J P Mulier; N T Smith; J J Settels; K H Wesseling
Journal:  Br J Anaesth       Date:  2001-08       Impact factor: 9.166

8.  The volume-dependency of parallel conductance throughout the cardiac cycle and its consequence for volume estimation of the left ventricle in patients.

Authors:  L Kornet; J J Schreuder; E T van der Velde; J R Jansen
Journal:  Cardiovasc Res       Date:  2001-09       Impact factor: 10.787

9.  Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter.

Authors:  J Baan; T T Jong; P L Kerkhof; R J Moene; A D van Dijk; E T van der Velde; J Koops
Journal:  Cardiovasc Res       Date:  1981-06       Impact factor: 10.787

10.  Hemodynamic effects of long-term cardiac resynchronization therapy: analysis by pressure-volume loops.

Authors:  Paul Steendijk; Sven A Tulner; Jeroen J Bax; Pranobe V Oemrawsingh; Gabe B Bleeker; Lieselot van Erven; Hein Putter; Harriette F Verwey; Ernst E van der Wall; Martin J Schalij
Journal:  Circulation       Date:  2006-03-06       Impact factor: 29.690

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.