INTRODUCTION: The aim of this study was to carry out an independent evaluation of the efficacy and security of a number of vessel ligation devices and ligatures.
MATERIALS AND METHODS: A vascular ligation model was devised using fresh, ex vivo porcine internal carotid arteries of varying external diameters. Coloured normal saline was infused via a pressure/monitor device through the artery. The end lumen was occluded by five different techniques: (i) braided suture in a surgeon's knot; (ii) a monofilament suture in a granny knot; (iii) a metallic clip (Ligaclip, Johnson and Johnson); (iv) a bipolar diathermy system (Ligasure, ValleyLab); and (v) an ultrasonically activated scalpel (Harmonic Scalpel, Johnson and Johnson). The vessels were subjected to supraphysiological pressures. Loss of haemostasis was evident by leakage of coloured perfusion fluid.
RESULTS: Secure haemostasis was obtained with all the techniques in all vessels below 5 mm in diameter. In vessels over 5 mm, secure haemostasis was obtained with all modalities except harmonic scalpel. With the harmonic scalpel, leaks occurred in 3/27 (11%) vessels between 5-6 mm and 3/5 (60%) vessels over 6 mm, confirming the manufacturer's instructions.
CONCLUSIONS: In this first, independent, randomised study comparing vessel ligation devices and ligatures, the manufacturer's claims for each of the haemostatic methods were accurate. We find that all the modalities tested perform as well as the traditional surgeon's knot in vessels of 5 mm and below.

Entities:  

Mesh:

Year:  2007        PMID: 17535611      PMCID: PMC1963577          DOI: 10.1308/003588407X183418

Source DB:  PubMed          Journal:  Ann R Coll Surg Engl        ISSN: 0035-8843            Impact factor:   1.891


  7 in total

1.  Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model.

Authors:  Jaime Landman; Kurt Kerbl; Jamil Rehman; Cassio Andreoni; Peter A Humphrey; William Collyer; Ephrem Olweny; Chandru Sundaram; Ralph V Clayman
Journal:  J Urol       Date:  2003-02       Impact factor: 7.450

2.  Comparison of the Ti-knot device and Hem-o-lok clips with other devices commonly used for laparoscopic renal-artery ligation.

Authors:  Jean Joseph; Yuk-Yuen Leung; Louis Eichel; Karl Scheidweiler; Erdal Erturk; Ronald Wood
Journal:  J Endourol       Date:  2004-03       Impact factor: 2.942

3.  Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries.

Authors:  K L Harold; H Pollinger; B D Matthews; K W Kercher; R F Sing; B T Heniford
Journal:  Surg Endosc       Date:  2003-06-13       Impact factor: 4.584

4.  Safety evaluation of laparoscopically applied clips.

Authors:  T Papaioannou; L Daykhovsky; W S Grundfest
Journal:  J Laparoendosc Surg       Date:  1996-04

5.  High-burst-strength, feedback-controlled bipolar vessel sealing.

Authors:  J S Kennedy; P L Stranahan; K D Taylor; J G Chandler
Journal:  Surg Endosc       Date:  1998-06       Impact factor: 4.584

6.  Ligasure versus sutures in bladder replacement with Montie ileal neobladder after radical cystectomy.

Authors:  Pasqualino Manasia; Antonio Alcaraz; Juan Alcover
Journal:  Arch Ital Urol Androl       Date:  2003-12

7.  Extravasal occlusion of large vessels with titanic clips: efficiency, indications, and contraindications.

Authors:  Yu V Vasilenko; A I Kim; S A Kotov
Journal:  Bull Exp Biol Med       Date:  2002-11       Impact factor: 0.804

  7 in total
  4 in total

1.  Multicenter randomized comparison of LigaSure versus conventional surgery for gastrointestinal carcinoma.

Authors:  Nobuhiro Takiguchi; Matsuo Nagata; Hiroaki Soda; Yukihiro Nomura; Wataru Takayama; Jun Yasutomi; Youichi Tohyama; Munemasa Ryu
Journal:  Surg Today       Date:  2010-11-03       Impact factor: 2.549

2.  Thyroidectomy with LigaSure.

Authors:  Mile Ignjatović; Zoran Kostić
Journal:  Surg Today       Date:  2011-05-28       Impact factor: 2.549

3.  Comparison of various methods of vessel ligation: what is the safest method?

Authors:  Chang-Sup Lim; Jin-Young Jang; Seung Eun Lee; You Jeong Lee; Mee Joo Kang; Sun-Whe Kim
Journal:  Surg Endosc       Date:  2013-03-07       Impact factor: 4.584

4.  Vessel sealing comparison: old school is still hip.

Authors:  Sasha J Tharakan; Dennis Hiller; Rachel M Shapiro; Sourav K Bose; Thane A Blinman
Journal:  Surg Endosc       Date:  2016-02-19       Impact factor: 4.584

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.