Comparison of various methods of vessel ligation: what is the safest method?

Chang-Sup Lim1, Jin-Young Jang, Seung Eun Lee, You Jeong Lee, Mee Joo Kang, Sun-Whe Kim.   

Abstract

BACKGROUND: Recently, various hemostatic devices, such as metal or plastic clips, ultrasonic coagulating shears, and electrothermal bipolar vessel sealer, have been widely applied during an operation. The purpose of this study was to analyze the mechanical, histological, and biochemical differences of vessels sealed with various hemostatic devices.
METHODS: Thirty New Zealand White rabbits were randomly allocated into five groups, and short gastric vessels were ligated with metal clip, plastic clip, Harmonic Ace, LigaSure, or tie ligation. These vessels were harvested 3 days after operation and histologically analyzed at the site of ligation (proximal) and 5 mm from it (distal). Perivascular fibrosis was assessed on a score of mild to severe according to the severity. Inducible and endothelial nitric oxide synthase (iNOS and eNOS) mRNA expressions were measured quantitatively by real-time PCR at the sites of ligation. Abdominal aorta and inferior vena cava also were harvested and divided with each tool, and bursting pressures were measured.
RESULTS: Bursting pressures were measured in 164 arteries and 141 veins. LigaSure showed significantly lower bursting pressures in arteries (p < 0.001), and plastic clip showed significantly lower bursting pressures in veins (p < 0.001). There was a negative relationship between outer diameters and bursting pressures in both the arteries (rho = -0.505, p < 0.001) and the veins (rho = -0.106, p = 0.24). The degree of perivascular fibrosis was not statistically different in either the proximal (p = 0.447) or distal (p = 0.381) sites. The expressions of iNOS and eNOS were significantly lower in the LS group (p < 0.001, and p < 0.001, respectively).
CONCLUSIONS: There might be no clinical limitations when applying various hemostatic devices to small vessels under physiologic blood pressures. There were no acute histological differences between the hemostatic devices. However, LS showed the lowest iNOS and eNOS expressions, which might be due to thermal injuries of the whole vessel wall.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23468332     DOI: 10.1007/s00464-013-2866-y

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  17 in total

1.  Preservation of endothelial integrity and function in experimental vascular anastomosis with non-penetrating clips.

Authors:  C Zeebregts; J van den Dungen; H Buikema; J van der Want; R van Schilfgaarde
Journal:  Br J Surg       Date:  2001-09       Impact factor: 6.939

2.  Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model.

Authors:  Jaime Landman; Kurt Kerbl; Jamil Rehman; Cassio Andreoni; Peter A Humphrey; William Collyer; Ephrem Olweny; Chandru Sundaram; Ralph V Clayman
Journal:  J Urol       Date:  2003-02       Impact factor: 7.450

3.  Comparison of the Ti-knot device and Hem-o-lok clips with other devices commonly used for laparoscopic renal-artery ligation.

Authors:  Jean Joseph; Yuk-Yuen Leung; Louis Eichel; Karl Scheidweiler; Erdal Erturk; Ronald Wood
Journal:  J Endourol       Date:  2004-03       Impact factor: 2.942

4.  Prospective randomized trial of ligasure versus harmonic hemostasis technique in thyroidectomy.

Authors:  Reza Rahbari; Aarti Mathur; Mio Kitano; Marlon Guerrero; Wen T Shen; Quan-Yang Duh; Orlo H Clark; Electron Kebebew
Journal:  Ann Surg Oncol       Date:  2010-11-12       Impact factor: 5.344

5.  Bursting strength with various methods of renal artery ligation and potential mechanisms of failure.

Authors:  Sean P Elliott; Andrew B Joel; Maxwell V Meng; Marshall L Stoller
Journal:  J Endourol       Date:  2005-04       Impact factor: 2.942

6.  Reported failures of the polymer self-locking (Hem-o-lok) clip: review of data from the Food and Drug Administration.

Authors:  Maxwell V Meng
Journal:  J Endourol       Date:  2006-12       Impact factor: 2.942

Review 7.  Inducible nitric oxide synthase and vascular injury.

Authors:  M Kibbe; T Billiar; E Tzeng
Journal:  Cardiovasc Res       Date:  1999-08-15       Impact factor: 10.787

8.  Vascular clips have no significant effect on the cellular proliferation, intimal changes, or peak systolic velocity at anastomoses in rabbit vein grafts.

Authors:  J M Caiati; J D Madigan; G Bhagat; A I Benvenisty; R Nowygrod; G J Todd
Journal:  J Surg Res       Date:  2000-07       Impact factor: 2.192

9.  In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries.

Authors:  Ronald H Clements; Rajendra Palepu
Journal:  Surg Endosc       Date:  2007-05-04       Impact factor: 4.584

10.  Expression of endothelial nitric oxide synthase in human and rabbit gastrointestinal smooth muscle cells.

Authors:  B Teng; K S Murthy; J F Kuemmerle; J R Grider; K Sase; T Michel; G M Makhlouf
Journal:  Am J Physiol       Date:  1998-08
View more
  2 in total

1.  Validation of a Laparoscopic Ferromagnetic Technology-based Vessel Sealing Device and Comparative Study to Ultrasonic and Bipolar Laparoscopic Devices.

Authors:  Jennwood Chen; Curtis R Jensen; Preston K Manwaring; Robert E Glasgow
Journal:  Surg Laparosc Endosc Percutan Tech       Date:  2017-04       Impact factor: 1.719

2.  Pull-off characteristics of double-shanked compared to single-shanked ligation clips: an animal study.

Authors:  Martin Schenk; Gottfried Müller; Tim Oliver Greiner; Christine Fahrner; Alfred Königsrainer; Christian Thiel
Journal:  Innov Surg Sci       Date:  2016-06-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.