Comparison study of MOSFET detectors and diodes for entrance in vivo dosimetry in 18 MV x-ray beams.

N Jornet1, P Carrasco, D Jurado, A Ruiz, T Eudaldo, M Ribas.   

Abstract

The feasibility of dual bias dual metal oxide semiconductor field effect transistors (MOSFETs) for entrance in vivo dose measurements in high energy x-rays beams (18 MV) was investigated. A comparison with commercially available diodes for in vivo dosimetry for the same energy range was performed. As MOSFETs are sold without an integrated build-up cap, different caps were tested: 3 cm bolus, 2 cm bolus, 2 cm hemispherical cap of a water equivalent material (Plastic Water) and a metallic hemispherical cap. This metallic build-up cap is the same as the one that is mounted on the in vivo diode used in this study. Intrinsic precision and response linearity with dose were determined for MOSFETs and diodes. They were then calibrated for entrance in vivo dosimetry in an 18 MV x-ray beam. Calibration included determination of the calibration factor in standard reference conditions and of the correction factors (CF) when irradiation conditions differed from those of reference. Correction factors for field size, source surface distance, wedge, and temperature were determined. Sensitivity variation with accumulated dose and the lifetime of both types of detectors were also studied. Finally, the uncertainties of entrance in vivo measurements using MOSFET and diodes were discussed. Intrinsic precision for MOSFETs for the high sensitivity mode was 0.7% (1 s.d.) as compared to the 0.05% (1 s.d.) for the studied diodes. The linearity of the response with dose was excellent (R2 = 1.000) for both in vivo dosimetry systems. The absolute values of the studied correction factors for the MOSFETs when covered by the different build-up caps were of the same order of those determined for the diodes. However, the uncertainties of the correction factors for MOSFETs were significantly higher than for diodes. Although the intrinsic precision and the uncertainty on the CF was higher for MOSFET detectors than for the studied diodes, the total uncertainty in entrance dose determination, once they were calibrated, was of 2.9% (1 s.d.) while for diodes it was 2.0% (1 s.d.). MOSFETs showed no sensitivity variation with accumulated dose or temperature. When used in the high sensitivity mode, after approximately 50 Gy of accumulated dose MOSFETs could no longer be used as radiation dosimeters. In conclusion, MOSFETs can be used for entrance in vivo dosimetry in high energy x-rays beams if covered by an appropriate build-up cap. Metallic build-up caps, such as those used for in vivo diodes, have the advantage of greater patient comfort and less perturbation of the treatment field than the other build-up caps tested, while keeping the correction factors of the same order.

Entities:  

Mesh:

Year:  2004        PMID: 15487735     DOI: 10.1118/1.1785452

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Verification of Entrance Dose Measurements with Thermoluminescent Dosimeters in Conventional Radiotherapy Procedures Delivered with Co-60 Teletherapy Machine.

Authors:  O B Evwierhurhoma; Z A Ibitoye; C A Ojieh; Jtk Duncan
Journal:  Ann Med Health Sci Res       Date:  2015 Nov-Dec

2.  Radioluminescence characterization of in situ x-ray nanodosimeters: Potential real-time monitors and modulators of external beam radiation therapy.

Authors:  Jeffrey S Souris; Shih-Hsun Cheng; Charles Pelizzari; Nai-Tzu Chen; Patrick La Riviere; Chin-Tu Chen; Leu-Wei Lo
Journal:  Appl Phys Lett       Date:  2014-11-20       Impact factor: 3.791

3.  Real-time in vivo rectal wall dosimetry using plastic scintillation detectors for patients with prostate cancer.

Authors:  Landon Wootton; Rajat Kudchadker; Andrew Lee; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-01-17       Impact factor: 3.609

4.  A novel nanometric DNA thin film as a sensor for alpha radiation.

Authors:  Atul Kulkarni; Byeonghoon Kim; Sreekantha Reddy Dugasani; Pranav Joshirao; Jang Ah Kim; Chirag Vyas; Vijay Manchanda; Taesung Kim; Sung Ha Park
Journal:  Sci Rep       Date:  2013       Impact factor: 4.379

5.  Measurement and comparison of skin dose using OneDose MOSFET and Mobile MOSFET for patients with acute lymphoblastic leukemia.

Authors:  Essam H Mattar; Lina F Hammad; Huda I Al-Mohammed
Journal:  Med Sci Monit       Date:  2011-07

6.  Sensitivity study of an automated system for daily patient QA using EPID exit dose images.

Authors:  Audrey H Zhuang; Arthur J Olch
Journal:  J Appl Clin Med Phys       Date:  2018-03-06       Impact factor: 2.102

7.  MOSFET dosimeter characterization in MR-guided radiation therapy (MRgRT) Linac.

Authors:  Poonam Yadav; Abdelbasset Hallil; Dinesh Tewatia; David A P Dunkerley; Bhudatt Paliwal
Journal:  J Appl Clin Med Phys       Date:  2019-12-18       Impact factor: 2.102

8.  Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams.

Authors:  A Sathish Kumar; S D Sharma; B Paul Ravindran
Journal:  J Med Phys       Date:  2014-07

9.  In vivo prostate IMRT dosimetry with MOSFET detectors using brass buildup caps.

Authors:  Raj Varadhan; John Miller; Brenden Garrity; Michael Weber
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

10.  Patient dosimetry for total body irradiation using single-use MOSFET detectors.

Authors:  Tina Marie Briere; Ramesh Tailor; Naresh Tolani; Karl Prado; Richard Lane; Shiao Woo; Chul Ha; Michael T Gillin; A Sam Beddar
Journal:  J Appl Clin Med Phys       Date:  2008-11-03       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.